Tympanic, infrared skin, and temporal artery scan thermometers compared with rectal measurement in children: a real-life assessment
- PMID:25067984
- PMCID: PMC4008772
- DOI: 10.1016/j.curtheres.2013.11.005
Tympanic, infrared skin, and temporal artery scan thermometers compared with rectal measurement in children: a real-life assessment
Abstract
Introduction: Body temperature measurement in children is of clinical relevance. Although rectal measurement is the gold standard, less invasive tools have become available. We aimed to describe the accuracy of tympanic, infrared skin, or temporal artery scan thermometers compared with rectal measurement to reflect core temperature.
Methods: Rectal (Filac 3000; Covidien, Mechelen, Belgium), tympanic (AccuSystem Genius2 Typmanic Infrared Ear Thermometer, Covidien, Mechelen, Belgium), temporal artery scan (Exergen, Exergen Corp, Watertown, Massachusetts), and infrared (ThermoFlash Contactless Medical Electronic Thermometer, Visiomedlab, Paris, France) body temperature measurements were randomly performed and readings were collected once. Temperature readings were described as median and range, and observations were compared with rectal temperature readings (using Wilcoxon, Bland-Altman, sensitivity, and specificity tests). The child's comfort was assessed by the child, parent, and nurse (using Likert scales) and ease of use was assessed by nurses (using visual analog scale).
Results: Based on observations in 294 (median age = 3.2 years, range = 0.02-17 years) children, the mean difference was 0.49°C (tympanic scan; P < 0.0001), 0.34°C (infrared skin scan; P < 0.0001), and 0°C (temporal artery scan; P = 0.9288), respectively, when compared with rectal temperature readings. Based on visual inspection of Bland-Altman plots, all tools overestimated the temperature at lower body temperature and underestimated the temperature at higher body temperature, resulting in a sensitivity of 22% to 41% and a specificity of 98% to 100% for rectal temperatures above 38°C. The Likert scale scores and the visual analog scale scores for rectal measurement were only slightly higher when compared with the other methods.
Conclusions: All noninvasive techniques underperformed compared with rectal measurement. The temporal artery scan deviations were smallest, but all noninvasive techniques overestimate lower temperatures and underestimate higher temperatures compared with rectal measurement. In our hands, temporal artery scan measurement seems to be second best, but not yet ideal.
Keywords: Bland-Altman; body temperature; infrared skin scan; rectal; temporal artery scan; tympanic scan.
Figures



Similar articles
- Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes.Mogensen CB, Wittenhoff L, Fruerhøj G, Hansen S.Mogensen CB, et al.BMC Pediatr. 2018 Jan 26;18(1):15. doi: 10.1186/s12887-018-0994-1.BMC Pediatr. 2018.PMID:29373961Free PMC article.
- Non-contact infrared thermometers compared with current approaches in primary care for children aged 5 years and under: a method comparison study.Van den Bruel A, Verbakel J, Wang K, Fleming S, Holtman G, Glogowska M, Morris E, Edwards G, Abakar Ismail F, Curtis K, Goetz J, Barnes G, Slivkova R, Nesbitt C, Aslam S, Swift E, Williams H, Hayward G.Van den Bruel A, et al.Health Technol Assess. 2020 Oct;24(53):1-28. doi: 10.3310/hta24530.Health Technol Assess. 2020.PMID:33111663Free PMC article.
- [Infrared temperature measurement in the ear canal with the DIATEK 9000 Instatemp and the DIATEK 9000 Thermoguide. Comparison with methods of temperature measurement in other body parts].Fritz U, Rohrberg M, Lange C, Weyland W, Bräuer A, Braun U.Fritz U, et al.Anaesthesist. 1996 Nov;45(11):1059-66. doi: 10.1007/s001010050340.Anaesthesist. 1996.PMID:9012301Clinical Trial.German.
- A systematic review of the accuracy of peripheral thermometry in estimating core temperatures among febrile critically ill patients.Jefferies S, Weatherall M, Young P, Beasley R.Jefferies S, et al.Crit Care Resusc. 2011 Sep;13(3):194-9.Crit Care Resusc. 2011.PMID:21880009Review.
- Infrared ear thermometry compared with rectal thermometry in children: a systematic review.Craig JV, Lancaster GA, Taylor S, Williamson PR, Smyth RL.Craig JV, et al.Lancet. 2002 Aug 24;360(9333):603-9. doi: 10.1016/S0140-6736(02)09783-0.Lancet. 2002.PMID:12241932Review.
Cited by
- Fever Incidence Is Much Lower in the Morning than the Evening: Boston and US National Triage Data.Harding C, Pompei F, Bordonaro SF, McGillicuddy DC, Burmistrov D, Sanchez LD.Harding C, et al.West J Emerg Med. 2020 Jun 24;21(4):909-917. doi: 10.5811/westjem.2020.3.45215.West J Emerg Med. 2020.PMID:32726264Free PMC article.
- Clinical accuracy of infrared temperature measurement devices: a comparison against non-invasive core-body temperature.Holder T, Hooper FSW, Yates D, Tse Z, Patil S, Moussa A, Batten L, Radhakrishnan V, Allison M, Hewitt C, Keding A, Forshaw G, Jayagopal V.Holder T, et al.Clin Med (Lond). 2023 Mar;23(2):157-163. doi: 10.7861/clinmed.2022-0252.Clin Med (Lond). 2023.PMID:36958833Free PMC article.
- An investigation into the effects of acupuncture on radial pressure pulse waves in patients with low back pain: A protocol for a quasi-experimental study.Ng HP, Huang CM, Ho WC, Lee YC.Ng HP, et al.Contemp Clin Trials Commun. 2019 May 23;15:100384. doi: 10.1016/j.conctc.2019.100384. eCollection 2019 Sep.Contemp Clin Trials Commun. 2019.PMID:31198880Free PMC article.
- Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes.Mogensen CB, Wittenhoff L, Fruerhøj G, Hansen S.Mogensen CB, et al.BMC Pediatr. 2018 Jan 26;18(1):15. doi: 10.1186/s12887-018-0994-1.BMC Pediatr. 2018.PMID:29373961Free PMC article.
- Regression model for predicting core body temperature in infrared thermal mass screening.Limpabandhu C, Hooper FSW, Li R, Tse Z.Limpabandhu C, et al.IPEM Transl. 2022 Nov-Dec;3:100006. doi: 10.1016/j.ipemt.2022.100006. Epub 2022 Jul 15.IPEM Transl. 2022.PMID:35854880Free PMC article.
References
- Davis T. NICE guideline: feverish illness in children—assessment and initial management in children younger than 5 years. Arch Dis Child Educ Pract Ed. 2013;98:232–235. - PubMed
- Walson P.D., Jones J., Chesney R. Antipyretic efficacy and tolerability of a single intravenous dose of the acetaminophen prodrug propacetamol in children: a randomized, double-blind, placebo-controlled trial. Clin Ther. 2006;28:762–769. - PubMed
- Allegaert K., van den Anker J. Pharmacokinetics and pharmacodynamics of intravenous acetaminophen in neonates. Expert Rev Clin Pharmacol. 2011;4:713–718. - PubMed
- Craig J.V., Lancaster G.A., Taylor S. Infrared ear thermometry compared with rectal thermometry in children: a systematic review. Lancet. 2002;360:603–609. - PubMed
Related information
LinkOut - more resources
Full Text Sources
Other Literature Sources