CT Image Reconstruction Basics



CT Image Reconstruction Basics


Joachim Hornegger

Andreas Maier

Markus Kowarschik



Computed tomography perfusion (CTP) imaging requires the reconstruction of a series of time-dependent volumetric datasets. The sequence of CT volumes measures the dynamics of contrast agent both in the vasculature and in the parenchyma.

CT image reconstruction refers to the computational process of determining tomographic images from x-ray projection images of the irradiated patient. Image reconstruction is a computationally intensive task and one of the most crucial steps in the CT imaging process.

As the basics of x-ray physics were detailed inChapter 1, the reader should assume that the result of the x-ray image formation is an attenuation image. Each individual pixel on the detector, therefore, represents a line integral, that is, the accumulation of all x-ray attenuation coefficients along the projection line. Here, the projection line is the connecting line of the x-ray focal spot with the center of the respective detector pixel.


From Line Integrals to Voxels

Despite the discrete nature of the projection images, most reconstruction theory uses acontinuous framework, in which the reconstruction algorithms are derived mathematically. The problem ofdiscrete sampling is then solved within the final formulation of the reconstruction algorithm. As the discrete sampling is a more technical issue, it is neglected here for the ease of presentation, and the reader should refer to the literature that details these steps.1 Thus, the description will remain in a continuous domain in this section.


Radon Transform

For ease of understanding, this approach is explained in the two-dimensional (2D) (x, y) plane. The source-detector arrangement is rotating around the object (Fig. 2.1). In the following, Dirac’s δ-function is used to select the parts of the object that are traversed by the x-ray. Recall that δ(t) is zero everywhere except att = 0, where δ becomes infinite. Furthermore, the δ-function fulfills for any real-valued functiong(x) the shifting property:


A parallel projectionp(s, θ) at detector elements and at source/detector array rotation angle θ can be written in the following notation:


wheref (x, y) represents the x-ray attenuation coefficient at object location (x, y). Thus, the δ-function selects the linex cos θ +y sin θ =s that connects the detector and the source array ats while both arrays are rotated by θ. Thus,p(s, θ) describes a line integral whose value is observed at the detector.

This process of turning function valuesf (x, y) into line integral valuesp(s, θ) is also referred to as theRadon transform in 2D. The fundamental problem of CT reconstruction is the computation of function valuesf(x, y) from the measured line integral valuesp(s, θ); that is, theinverse Radon transform.

Another important concept in CT reconstruction isback-projection, as described in the following. Backprojection is a kind of conjugate (but not inverse) process to the (forward) projection. For the case of parallel-beam x-ray projections, it assigns to a point (x, y) in object coordinates the integralb(x, y) of the projection values that lie on the x-rays passing through (x, y):



Fourier Slice Theorem

TheFourier slice theorem is fundamental to many CT reconstruction approaches. It states that the 1D Fourier transformP(ω, θ) of a projectionp(s, θ) in parallel-beam geometry for a fixed rotation angle θ is identical to the 1D profile through the origin of the 2D Fourier transformF (ω cos θ, ω sin θ) of the irradiated object (x, y).Figure 2.2 displays this process.






FIGURE 2.1. Parallel-beam geometry and the generation of a parallel-beam projectionp (s, θ).







FIGURE 2.2. Illustration of the Fourier slice theorem.


Filtered Back-Projection for 2D Parallel-Beam Geometry

With the Fourier slice theorem one of the most commonly used reconstruction algorithms—thefiltered back-projection (FBP) method—can be elegantly derived. A derivation is shown in theGeek Box, Fourier slice theorem Proof. The resulting algorithm uses two concepts: (i) filtering and (ii) back-projection.

The filterh(s) is calledramp filter because of the shape of |ω|. This filter corrects the oversampling that occurs at the center of the Fourier space. This is accomplished by enhancing high spatial frequency components while dampening low spatial frequency components. In practice, this filtering operation is often combined with additional filtering to achieve certain image characteristics. Various kernels can be created and embedded into the reconstruction process to obtain smoother or sharper images.



Geek Box: Fourier Slice Theorem Proof

To prove this, we start with the 1D Fourier transformP(ω, θ) ofp(s, θ):


Using the definition of the projectionp(s, θ), we obtain


Rearranging the order of the integrals then yields


which, after elimination of the delta function, reads as


Finally, using the definition of the 2D Fourier transform, we obtain


which results in the proposed statement


By varying θ, we get the complete Fourier transformFpolar(ω, θ) of the unknown functionf(x, y) in polar coordinates (ω, θ).



Geek Box: Derivation of the Filtered Back-Projection Algorithm

To illustrate this derivation, we start with the inverse Fourier transformF(u, v):


and rewrite it to polar coordinatesFpolar(ω, θ):34


According to the Fourier slice theorem, we obtain


which contains a product of the projection in 1D Fourier spaceP(ω, θ) and |ω|. The inverse Fourier transform ofP(ω, θ). |ω| corresponds to a convolution in spatial domain. The inverse Fourier transform of |ω| shall be defined by the filter kernelh(s). Hence, in spatial domain, the previous equation can then be written as


which is the back-projection of the projection datap(s, θ) convolved withh(s).

After filtering, the projection data is back-projected into image space, as described in the previous section. Note that the FBP algorithm uses the Fourier slice theorem only implicitly. That is, both filtering and backprojection can be implemented in a way that the actual computation of the Fourier transform is not required. However, as the ramp filter is a global operation, the implementation using fast Fourier methods is often favorable.


Acquisition Geometries

The filtered back-projection algorithm is an efficient and robust method for the computation of tomographic slice images. Using afan-beam geometry based on a single x-ray source that is collimated toward a curved array of detector elements, it is possible to circumvent long acquisition times and bulky hardware that would be required forparallel-beam geometry. Doing so, a single source is sufficient to collect multiple rays at the same time (seeFig. 2.3), but the Fourier slice theorem cannot be applied straightforwardly.







FIGURE 2.3. With fan-beam geometry, one is able to image multiple detector elements at the same time with only one x-ray source. Depending on the type of detector, two different geometries emerge. Using a curved detector, an equiangular geometry described byg (γ, β) is obtained(left), and using a linear detector, an equally spaced geometry denoted byg (t, β) is created(right).

If a full rotation with either parallel-beam or fan-beam acquisition geometries is performed, one can easily observe that both geometries cover identical data. They are merely collected in a different sequence. If β is considered as the rotation angle of source and detector, γ as the angle to the respective detector elements, andD as the source-to-rotation axis distance, the following relations between identical rays in fan-beam and parallel-beam geometry are obtained:



This relation offers two possible solutions for image reconstruction. Either the projection data is reordered into parallel-beam geometry in a so-called “rebinning” step, or the reconstruction algorithm has to be adapted to the acquisition geometry. Both algorithms are used in practice. The interpolation operation in the rebinning step must be handled with caution as it may lead to an unintended loss in image resolution. TheGeek Box, Fan-Beam Reconstruction without Rebinning, sketches an idea how to obtain an algorithm without rebinning.



Geek Box: Fan-Beam Reconstruction without Rebinning

To omit rebinning, one has to reshape the filtered back-projection reconstruction formula by a change of coordinates of the integral variabless and θ according to the ray identities in Eq. (rays). This yields the following fan-beam reconstruction formula for a curved detector:


Note that we introduced the variablesD′ =D′(x, y, β) and γ′ = γ′(x, y, β) that describe distance and angle of the reconstructed point (x, y), as shown inFigure 2. This equation can be interpreted as a convolution of the fan-beam projectiong(γ, β) with a fan-beam ramp filter
. Furthermore, a weighting factor cos γ is applied. This weighting is also referred to as thecosine weight.1

In case of a linear detector with equally spaced detector elements, a slightly different reconstruction formula is obtained, as the integral variables are substituted tot and β, wheret is the index of the detector element:


Again, we introduced several variables: the indext′ =t′(x, y, β) of the projection of the reconstruction point (x, y) on the detector and the depthU′ =U′(x, y, β), as shown inFigure 2.2. In this formulation, we find the parallel-beam ramp filterh(t′ –t) from the previous section. It is applied to a fan-beam projectiong(t, β). Prior to the convolution, this projection was weighted with the factor
, the cosine-weight for the linear detector case. Note that the back-projection is weighted with a distance weight (U′)−2, which is dependent on the image point to be reconstructed.







FIGURE 2.4. Effect of the cone-beam geometry in a circular trajectory with a cone angle of 30 degrees.From left to right: XZ slice of the original Defrise disk phantom,2 the XZ slice of the reconstructed volume using the FDK algorithm, and the projection image for all projection angles.

Modern CT scanners introducedmultirow detector arrays. These arrays introduce a second dimension on the detector and the rays from source to detector form a cone. Thiscone-beam geometry allows even faster data acquisition. However, the acquisition geometry has to be decided with care, as the rays do no longer fall into the same plane. Hence, even if rotation is performed on a full circle, there are rays that are required for reconstruction that are not collected. This missing data causes artifacts in the reconstruction result that are calledcone-beam artifacts (Fig. 2.4).

In a divergent beam scenario, only those line integrals that intersect the path of the x-ray source can be measured as all x-rays originate from there. This path is also referred to as thesource trajectory. To create a theoretically correct reconstruction result, a complete data set must be acquired. According toTuy’s sufficiency condition,2 every plane that intersects the object has to intersect the path of the x-ray source.Figure 2.5 shows examples of incomplete and complete trajectories. Although a circular trajectory is insufficient, as planes that are parallel to the plane of rotation do not intersect with the source trajectory, this problem can be resolved by adding lines to it (seeFig. 2.5, bottom left). A more elegant way is to rotate the source along a helical path as a continuous motion is obtained. Note that a change in the path of the trajectory also often implies a different reconstruction algorithm. For tilted gantries and helical trajectories, for example, the reconstruction method, the size of the field of view, and the length of the pitch have to be adjusted. If the standard reconstruction method is not adjusted in an appropriate manner, it will cause in severe artifacts in the reconstructed image.3

Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Oct 7, 2018 | Posted by inMAGNETIC RESONANCE IMAGING |Comments Off on CT Image Reconstruction Basics

Full access? Get Clinical Tree

Get Clinical Tree app for offline access
Get Clinical Tree app for offline access

[8]ページ先頭

©2009-2025 Movatter.jp