Rapid calculation of T1 using variable flip angle gradient refocused imaging
- PMID:3626789
- DOI: 10.1016/0730-725x(87)90021-x
Rapid calculation of T1 using variable flip angle gradient refocused imaging
Abstract
We present a method for rapid measurement of T1 relaxation times using gradient refocused images at limited flip angles and short repetition times. This "variable nutation" techniques was investigated using a T1 phantom. There was a high correlation between measurements obtained with the variable nutation and partial saturation techniques. The ability of this method to create calculated T1 images is also demonstrated. We conclude that the variable nutation method may allow measurement of T1 relaxation times with a significant reduction in acquisition time compared to partial saturation techniques.
Similar articles
- Rapid assessment of longitudinal relaxation time in materials and tissues with extremely fast signal decay using UTE sequences and the variable flip angle method.Springer F, Steidle G, Martirosian P, Syha R, Claussen CD, Schick F.Springer F, et al.Invest Radiol. 2011 Oct;46(10):610-7. doi: 10.1097/RLI.0b013e31821c44cd.Invest Radiol. 2011.PMID:21577126
- Measurement of T1 relaxation times of cardiac phosphate metabolites using BIR-4 adiabatic RF pulses and a variable nutation method.Sakuma H, Nelson SJ, Vigneron DB, Hartiala J, Higgins CB.Sakuma H, et al.Magn Reson Med. 1993 May;29(5):688-91. doi: 10.1002/mrm.1910290517.Magn Reson Med. 1993.PMID:8505906
- Flip angle-optimized fast dynamic T1 mapping with a 3D gradient echo sequence.Dietrich O, Freiermuth M, Willerding L, Reiser MF, Peller M.Dietrich O, et al.Magn Reson Med. 2015 Mar;73(3):1158-63. doi: 10.1002/mrm.25199. Epub 2014 Mar 17.Magn Reson Med. 2015.PMID:24639175
- T1 mapping with the variable flip angle technique: A simple correction for insufficient spoiling of transverse magnetization.Baudrexel S, Nöth U, Schüre JR, Deichmann R.Baudrexel S, et al.Magn Reson Med. 2018 Jun;79(6):3082-3092. doi: 10.1002/mrm.26979. Epub 2017 Oct 20.Magn Reson Med. 2018.PMID:29052267
- WET, a T1- and B1-insensitive water-suppression method for in vivo localized 1H NMR spectroscopy.Ogg RJ, Kingsley PB, Taylor JS.Ogg RJ, et al.J Magn Reson B. 1994 May;104(1):1-10. doi: 10.1006/jmrb.1994.1048.J Magn Reson B. 1994.PMID:8025810
Cited by
- A Single-Scan, Rapid Whole-Brain Protocol for Quantitative Water Content Mapping With Neurobiological Implications.Oros-Peusquens AM, Loução R, Abbas Z, Gras V, Zimmermann M, Shah NJ.Oros-Peusquens AM, et al.Front Neurol. 2019 Dec 20;10:1333. doi: 10.3389/fneur.2019.01333. eCollection 2019.Front Neurol. 2019.PMID:31920951Free PMC article.
- DCE-MRI biomarkers of tumour heterogeneity predict CRC liver metastasis shrinkage following bevacizumab and FOLFOX-6.O'Connor JP, Rose CJ, Jackson A, Watson Y, Cheung S, Maders F, Whitcher BJ, Roberts C, Buonaccorsi GA, Thompson G, Clamp AR, Jayson GC, Parker GJ.O'Connor JP, et al.Br J Cancer. 2011 Jun 28;105(1):139-45. doi: 10.1038/bjc.2011.191. Epub 2011 Jun 14.Br J Cancer. 2011.PMID:21673686Free PMC article.
- Data-driven mapping of hypoxia-related tumor heterogeneity using DCE-MRI and OE-MRI.Featherstone AK, O'Connor JPB, Little RA, Watson Y, Cheung S, Babur M, Williams KJ, Matthews JC, Parker GJM.Featherstone AK, et al.Magn Reson Med. 2018 Apr;79(4):2236-2245. doi: 10.1002/mrm.26860. Epub 2017 Aug 30.Magn Reson Med. 2018.PMID:28856728Free PMC article.
- Deep learning-based T1-enhanced selection of linear attenuation coefficients (DL-TESLA) for PET/MR attenuation correction in dementia neuroimaging.Chen Y, Ying C, Binkley MM, Juttukonda MR, Flores S, Laforest R, Benzinger TLS, An H.Chen Y, et al.Magn Reson Med. 2021 Jul;86(1):499-513. doi: 10.1002/mrm.28689. Epub 2021 Feb 8.Magn Reson Med. 2021.PMID:33559218Free PMC article.
- Functional MRI of the kidneys.Zhang JL, Rusinek H, Chandarana H, Lee VS.Zhang JL, et al.J Magn Reson Imaging. 2013 Feb;37(2):282-93. doi: 10.1002/jmri.23717.J Magn Reson Imaging. 2013.PMID:23355431Free PMC article.Review.
Publication types