.2018 Jan 18:10:13-20.
doi: 10.2147/NSS.S151085. eCollection 2018.

Actigraphy-based sleep estimation in adolescents and adults: a comparison with polysomnography using two scoring algorithms

Affiliations

Actigraphy-based sleep estimation in adolescents and adults: a comparison with polysomnography using two scoring algorithms

Mirja Quante et al. Nat Sci Sleep..

Abstract

Objectives: Actigraphy is widely used to estimate sleep-wake time, despite limited information regarding the comparability of different devices and algorithms. We compared estimates of sleep-wake times determined by two wrist actigraphs (GT3X+ versus Actiwatch Spectrum [AWS]) to in-home polysomnography (PSG), using two algorithms (Sadeh and Cole-Kripke) for the GT3X+ recordings.

Subjects and methods: Participants included a sample of 35 healthy volunteers (13 school children and 22 adults, 46% male) from Boston, MA, USA. Twenty-two adults wore the GT3X+ and AWS simultaneously for at least five consecutive days and nights. In addition, actigraphy and PSG were concurrently measured in 12 of these adults and another 13 children over a single night. We used intraclass correlation coefficients (ICCs), epoch-by-epoch comparisons, pairedt-tests, and Bland-Altman plots to determine the level of agreement between actigraphy and PSG, and differences between devices and algorithms.

Results: Each actigraph showed comparable accuracy (0.81-0.86) for sleep-wake estimation compared to PSG. When analyzing data from the GT3X+, the Cole-Kripke algorithm was more sensitive (0.88-0.96) to detect sleep, but less specific (0.35-0.64) to detect wake than the Sadeh algorithm (sensitivity: 0.82-0.91, specificity: 0.47-0.68). Total sleep time measured using the GT3X+ with both algorithms was similar to that obtained by PSG (ICC=0.64-0.88). In contrast, agreement between the GT3X+ and PSG wake after sleep onset was poor (ICC=0.00-0.10). In adults, the GT3X+ using the Cole-Kripke algorithm provided data comparable to the AWS (mean bias=3.7±19.7 minutes for total sleep time and 8.0±14.2 minutes for wake after sleep onset).

Conclusion: The two actigraphs provided comparable and accurate data compared to PSG, although both poorly identified wake episodes (i.e., had low specificity). Use of actigraphy scoring algorithm influenced the mean bias and level of agreement in sleep-wake times estimates. The GT3X+, when analyzed by the Cole-Kripke, but not the Sadeh algorithm, provided comparable data to the AWS.

Keywords: actigraphy; polysomnography; scoring algorithm; validation.

PubMed Disclaimer

Conflict of interest statement

Disclosure SR has received grant support from Jazz Pharmaceuticals, outside the submitted work. The other authors report no conflicts of interest in this work.

Figures

Figure 1
Figure 1
Bland–Altman plots of GT3X+ and AWS in 22 adults for (A) total sleep time: GT3X+: CK versus AWS,(B) WASO: GT3X+: CK versus AWS,(C) total sleep time: GT3X+: Sadeh versus AWS, and(D) WASO: GT3X+: Sadeh versus AWS. The dotted line indicates the mean of the differences or bias, and the dashed lines indicate the lower and upper 95% limits of agreement.Abbreviations: AWS, Actiwatch Spectrum; CK, Cole–Kripke; WASO, wake after sleep onset.
See this image and copyright information in PMC

Similar articles

See all similar articles

Cited by

See all "Cited by" articles

References

    1. Sadeh A. The role and validity of actigraphy in sleep medicine: an update. Sleep Med Rev. 2011;15(4):259–267. - PubMed
    1. Meltzer LJ, Montgomery-Downs HE, Insana SP, Walsh CM. Use of actigraphy for assessment in pediatric sleep research. Sleep Med Rev. 2012;16(5):463–475. - PMC - PubMed
    1. Amy W, Hawley MD, Rosemary SCH, Sarah NB. Actigraphy and Sleep/Wake Diaries. Oxford; Oxford University Press; 2013.
    1. Sadeh A, Sharkey KM, Carskadon MA. Activity-based sleep-wake identification: an empirical test of methodological issues. Sleep. 1994;17(3):201–207. - PubMed
    1. Cole RJ, Kripke DF, Gruen W, Mullaney DJ, Gillin JC. Automatic sleep/wake identification from wrist activity. Sleep. 1992;15(5):461–469. - PubMed

Grants and funding

LinkOut - more resources

Full text links
Dove Medical Press full text link Dove Medical Press Free PMC article
Cite
Send To

NCBI Literature Resources

MeSHPMCBookshelfDisclaimer

The PubMed wordmark and PubMed logo are registered trademarks of the U.S. Department of Health and Human Services (HHS). Unauthorized use of these marks is strictly prohibited.


[8]ページ先頭

©2009-2025 Movatter.jp