Movatterモバイル変換


[0]ホーム

URL:


Jump to content
WikipediaThe Free Encyclopedia
Search

Microscopy

From Wikipedia, the free encyclopedia
Viewing of objects which are too small to be seen with the naked eye
Not to be confused withMicroscopic orMicroscope.
Scanning electron microscope image ofpollen (false colors)
Microscopic examination in a biochemical laboratory

Microscopy is the technical field of usingmicroscopes to view subjects too small to be seen with thenaked eye (objects that are not within the resolution range of the normal eye).[1] There are three well-known branches of microscopy:optical,electron, andscanning probe microscopy, along with the emerging field ofX-ray microscopy.[citation needed]

Optical microscopy and electron microscopy involve thediffraction,reflection, orrefraction ofelectromagnetic radiation/electron beams interacting with thespecimen, and the collection of the scattered radiation or another signal in order to create an image. This process may be carried out by wide-field irradiation of the sample (for example standard light microscopy andtransmission electron microscopy) or by scanning a fine beam over the sample (for exampleconfocal laser scanning microscopy andscanning electron microscopy).Scanning probe microscopy involves the interaction of a scanning probe with the surface of the object of interest. The development of microscopy revolutionizedbiology, gave rise to the field ofhistology and so remains an essential technique in thelife andphysical sciences. X-ray microscopy is three-dimensional and non-destructive, allowing for repeated imaging of the same sample forin situ or 4D studies, and providing the ability to "see inside" the sample being studied before sacrificing it to higher resolution techniques. A 3D X-ray microscope uses the technique of computed tomography (microCT), rotating the sample 360 degrees and reconstructing the images. CT is typically carried out with a flat panel display. A 3D X-ray microscope employs a range of objectives, e.g., from 4X to 40X, and can also include a flat panel.

History

[edit]
Antonie van Leeuwenhoek (1632–1723)

The field of microscopy (optical microscopy) dates back to at least the 17th-century. Earlier microscopes, singlelensmagnifying glasses with limited magnification, date at least as far back as the wide spread use of lenses ineyeglasses in the 13th century[2] but more advancedcompound microscopes first appeared in Europe around 1620[3][4] The earliest practitioners of microscopy includeGalileo Galilei, who found in 1610 that he could close focus his telescope to view small objects close up[5][6] andCornelis Drebbel, who may have invented the compound microscope around 1620.[7][8]Antonie van Leeuwenhoek developed a very high magnification simple microscope in the 1670s and is often considered to be the first acknowledgedmicroscopist andmicrobiologist.[9][10]

Optical microscopy

[edit]
See also:Optical microscope
Stereo microscope

Optical or light microscopy involves passingvisible light transmitted through or reflected from the sample through a single lens or multiplelenses to allow a magnified view of the sample.[11] The resulting image can be detected directly by the eye, imaged on aphotographic plate, orcaptured digitally. The single lens with its attachments, or the system of lenses and imaging equipment, along with the appropriate lighting equipment, sample stage, and support, makes up the basic light microscope. The most recent development is thedigital microscope, which uses aCCD camera to focus on the exhibit of interest. The image is shown on a computer screen, so eye-pieces are unnecessary.[12]

Limitations

[edit]

Limitations of standard optical microscopy (bright field microscopy) lie in three areas;

  • The technique can only image dark or strongly refracting objects effectively.
  • There is adiffraction-limited resolution depending on incident wavelength; in visible range, the resolution of optical microscopy is limited to approximately 0.2  micrometres (see:microscope) and the practical magnification limit to ~1500x.[13]
  • Out-of-focus light from points outside the focal plane reduces image clarity.[14]

Live cells in particular generally lack sufficient contrast to be studied successfully, since the internal structures of the cell are colorless and transparent. The most common way to increase contrast is tostain the structures with selective dyes, but this often involves killing andfixing the sample.[15] Staining may also introduceartifacts, which are apparent structural details that are caused by the processing of the specimen and are thus not features of the specimen. In general, these techniques make use of differences in the refractive index of cell structures. Bright-field microscopy is comparable to looking through a glass window: one sees not the glass but merely the dirt on the glass. There is a difference, as glass is a denser material, and this creates a difference in phase of the light passing through. The human eye is not sensitive to this difference in phase, but clever optical solutions have been devised to change this difference in phase into a difference in amplitude (light intensity).[citation needed]

Techniques

[edit]
Main article:Optical microscope

To improve specimencontrast or highlight structures in a sample, special techniques must be used. A huge selection of microscopy techniques are available to increase contrast or label a sample.

Bright field

[edit]
Main article:Bright field microscopy

Bright field microscopy is the simplest of all the light microscopy techniques. Sample illumination is via transmitted white light, i.e. illuminated from below and observed from above. Limitations include low contrast of most biological samples and low apparent resolution due to the blur of out-of-focus material. The simplicity of the technique and the minimal sample preparation required are significant advantages.[citation needed]

Oblique illumination

[edit]

The use of oblique (from the side) illumination gives the image a three-dimensional appearance and can highlight otherwise invisible features. A more recent technique based on this method isHoffmann's modulation contrast, a system found on inverted microscopes for use in cell culture. Oblique illumination enhances contrast even in clear specimens; however, because light enters off-axis, the position of an object will appear to shift as the focus is changed. This limitation makes techniques like optical sectioning or accurate measurement on the z-axis impossible.

Dark field

[edit]
Main article:Dark field microscopy

Dark field microscopy is a technique for improving the contrast of unstained, transparent specimens.[16] Dark field illumination uses a carefully aligned light source to minimize the quantity of directly transmitted (unscattered) light entering the image plane, collecting only the light scattered by the sample. Dark field can dramatically improve image contrast – especially of transparent objects – while requiring little equipment setup or sample preparation. However, the technique suffers from low light intensity in the final image of many biological samples and continues to be affected by low apparent resolution.

Adiatom under Rheinberg illumination

Rheinberg illumination is a variant of dark field illumination in which transparent, colored filters are inserted just before thecondenser so that light rays at high aperture are differently colored than those at low aperture (i.e., the background to the specimen may be blue while the object appears self-luminous red). Other color combinations are possible, but their effectiveness is quite variable.[17]

Dispersion staining

[edit]
Main article:Dispersion staining

Dispersion staining is an optical technique that results in a colored image of a colorless object. This is an optical staining technique and requires no stains or dyes to produce a color effect. There are five different microscope configurations used in the broader technique of dispersion staining. They include brightfield Becke line, oblique, darkfield, phase contrast, and objective stop dispersion staining.

Phase contrast

[edit]
Phase-contrastlight micrograph of undecalcifiedhyaline cartilage showingchondrocytes andorganelles,lacunae andextracellular matrix
Main article:Phase contrast microscopy
Inelectron microscopy:Phase-contrast imaging

More sophisticated techniques will show proportional differences in optical density.Phase contrast is a widely used technique that shows differences inrefractive index as difference in contrast. It was developed by the Dutch physicistFrits Zernike in the 1930s (for which he was awarded the Nobel Prize in 1953). The nucleus in a cell for example will show up darkly against the surrounding cytoplasm. Contrast is excellent; however it is not for use with thick objects. Frequently, a halo is formed even around small objects, which obscures detail. The system consists of a circular annulus in the condenser, which produces a cone of light. This cone is superimposed on a similar sized ring within the phase-objective. Every objective has a different size ring, so for every objective another condenser setting has to be chosen. The ring in the objective has special optical properties: it, first of all, reduces the direct light in intensity, but more importantly, it creates an artificial phase difference of about a quarter wavelength. As the physical properties of this direct light have changed, interference with the diffracted light occurs, resulting in the phase contrast image. One disadvantage of phase-contrast microscopy is halo formation (halo-light ring).

Differential interference contrast

[edit]
Main article:Differential interference contrast microscopy

Superior and much more expensive is the use ofinterference contrast. Differences in optical density will show up as differences in relief. A nucleus within a cell will actually show up as a globule in the most often useddifferential interference contrast system according toGeorges Nomarski. However, it has to be kept in mind that this is anoptical effect, and the relief does not necessarily resemble the true shape. Contrast is very good and the condenser aperture can be used fully open, thereby reducing the depth of field and maximizing resolution.

The system consists of a special prism (Nomarski prism,Wollaston prism) in the condenser that splits light in an ordinary and an extraordinary beam. The spatial difference between the two beams is minimal (less than the maximum resolution of the objective). After passage through the specimen, the beams are reunited by a similar prism in the objective.

In a homogeneous specimen, there is no difference between the two beams, and no contrast is being generated. However, near a refractive boundary (say a nucleus within the cytoplasm), the difference between the ordinary and the extraordinary beam will generate a relief in the image. Differential interference contrast requires apolarized light source to function; two polarizing filters have to be fitted in the light path, one below the condenser (the polarizer), and the other above the objective (the analyzer).

Note: In cases where the optical design of a microscope produces an appreciable lateral separation of the two beams we have the case ofclassical interference microscopy, which does not result in relief images, but can nevertheless be used for the quantitative determination of mass-thicknesses of microscopic objects.

Interference reflection

[edit]
Main article:Interference reflection microscopy

An additional technique using interference isinterference reflection microscopy (also known as reflected interference contrast, or RIC). It relies on cell adhesion to the slide to produce an interference signal. If there is no cell attached to the glass, there will be no interference.

Interference reflection microscopy can be obtained by using the same elements used by DIC, but without the prisms. Also, the light that is being detected is reflected and not transmitted as it is when DIC is employed.

Fluorescence

[edit]
Images may also containartifacts. This is aconfocal laser scanningfluorescencemicrograph ofthale cress anther (part ofstamen). The picture shows among other things a nice red flowing collar-like structure just below the anther. However, an intact thale cress stamen does not have such collar, this is a fixation artifact: the stamen has been cut below the picture frame, andepidermis (upper layer of cells) of stamen stalk has peeled off, forming a non-characteristic structure. Photo: Heiti Paves fromTallinn University of Technology.
Main article:Fluorescence microscopy

When certain compounds are illuminated with high energy light, they emit light of a lower frequency. This effect is known asfluorescence. Often specimens show their characteristicautofluorescence image, based on their chemical makeup.

This method is of critical importance in the modern life sciences, as it can be extremely sensitive, allowing the detection of single molecules. Many fluorescentdyes can be used to stain structures or chemical compounds. One powerful method is the combination ofantibodies coupled to a fluorophore as inimmunostaining. Examples of commonly used fluorophores arefluorescein orrhodamine.

The antibodies can be tailor-made for a chemical compound. For example, one strategy often in use is the artificial production of proteins, based on the genetic code (DNA). These proteins can then be used to immunize rabbits, forming antibodies which bind to the protein. The antibodies are then coupled chemically to a fluorophore and used to trace the proteins in the cells under study.

Highly efficient fluorescentproteins such as thegreen fluorescent protein (GFP) have been developed using themolecular biology technique ofgene fusion, a process that links theexpression of the fluorescent compound to that of the target protein. This combined fluorescent protein is, in general, non-toxic to the organism and rarely interferes with the function of the protein under study. Genetically modified cells or organisms directly express the fluorescently tagged proteins, which enables the study of the function of the original proteinin vivo.

Growth ofprotein crystals results in both protein and salt crystals. Both are colorless and microscopic. Recovery of the protein crystals requires imaging which can be done by the intrinsic fluorescence of the protein or by using transmission microscopy. Both methods require an ultraviolet microscope as proteins absorbs light at 280 nm. Protein will also fluorescence at approximately 353 nm when excited with 280 nm light.[18]

Sincefluorescence emission differs inwavelength (color) from the excitation light, an ideal fluorescent image shows only the structure of interest that was labeled with the fluorescent dye. This high specificity led to the widespread use of fluorescence light microscopy in biomedical research. Different fluorescent dyes can be used to stain different biological structures, which can then be detected simultaneously, while still being specific due to the individual color of the dye.

To block the excitation light from reaching the observer or the detector,filter sets of high quality are needed. These typically consist of anexcitation filter selecting the range of excitationwavelengths, adichroic mirror, and anemission filter blocking the excitation light. Most fluorescencemicroscopes are operated in the Epi-illumination mode (illumination and detection from one side of the sample) to further decrease the amount of excitation light entering the detector.

See also:total internal reflection fluorescence microscopeNeuroscience

Confocal

[edit]
Main article:Confocal microscopy

Confocal laser scanning microscopy uses a focusedlaser beam (e.g. 488 nm) that is scanned across the sample to excitefluorescence in a point-by-point fashion. The emitted light is directed through a pinhole to prevent out-of-focus light from reaching the detector, typically aphotomultiplier tube. The image is constructed in a computer, plotting the measured fluorescence intensities according to the position of the excitation laser. Compared to full sample illumination, confocal microscopy gives slightly higher lateral resolution and significantly improvesoptical sectioning (axial resolution). Confocal microscopy is, therefore, commonly used where 3D structure is important.

A subclass of confocal microscopes arespinning disc microscopes which are able to scan multiple points simultaneously across the sample. A corresponding disc with pinholes rejects out-of-focus light. The light detector in a spinning disc microscope is a digital camera, typicallyEM-CCD orsCMOS.

Two-photon microscopy

[edit]

A two-photon microscope is also a laser-scanning microscope, but instead of UV, blue or green laser light, apulsed infrared laser is used for excitation. Only in the tiny focus of the laser is the intensity high enough to generate fluorescence bytwo-photon excitation, which means that no out-of-focus fluorescence is generated, and no pinhole is necessary to clean up the image.[19] This allows imaging deep in scattering tissue, where a confocal microscope would not be able to collect photons efficiently.[20] Two-photon microscopes with wide-field detection are frequently used for functional imaging, e.g.calcium imaging, in brain tissue.[21] They are marketed asMultiphoton microscopes by several companies, although the gains of using 3-photon instead of 2-photon excitation are marginal.

Single plane illumination microscopy and light sheet fluorescence microscopy

[edit]
Main article:Light sheet fluorescence microscopy

Using a plane of light formed by focusing light through a cylindrical lens at a narrow angle or by scanning a line of light in a plane perpendicular to the axis of objective, high resolution optical sections can be taken.[22][23][24] Single plane illumination, or light sheet illumination, is also accomplished usingbeam shaping techniques incorporating multiple-prism beam expanders.[25][26] The images are captured by CCDs. These variants allow very fast and high signal to noise ratio image capture.

Wide-field multiphoton microscopy

[edit]
Main article:Wide-field multiphoton microscopy

Wide-field multiphoton microscopy[27][28][29][30] refers to anoptical non-linear imaging technique in which a large area of the object is illuminated and imaged without the need for scanning. High intensities are required to induce non-linear optical processes such astwo-photon fluorescence orsecond harmonic generation. Inscanning multiphoton microscopes the high intensities are achieved by tightly focusing the light, and the image is obtained by beam scanning. Inwide-field multiphoton microscopy the high intensities are best achieved using anoptically amplified pulsed laser source to attain a large field of view (~100 μm).[27][28][29] The image in this case is obtained as a single frame with a CCD camera without the need of scanning, making the technique particularly useful to visualize dynamic processes simultaneously across the object of interest. With wide-field multiphoton microscopy the frame rate can be increased up to a 1000-fold compared tomultiphoton scanning microscopy.[28] In scattering tissue, however, image quality rapidly degrades with increasing depth.

Deconvolution

[edit]

Fluorescence microscopy is a powerful technique to show specifically labeled structures within a complex environment and to provide three-dimensional information of biological structures. However, this information is blurred by the fact that, upon illumination, all fluorescently labeled structures emit light, irrespective of whether they are in focus or not. So an image of a certain structure is always blurred by the contribution of light from structures that are out of focus. This phenomenon results in a loss of contrast especially when using objectives with a high resolving power, typically oil immersion objectives with a high numerical aperture.

Mathematically modeled Point Spread Function of a pulsed THz laser imaging system.[31]

However, blurring is not caused by random processes, such as light scattering, but can be well defined by the optical properties of the image formation in the microscope imaging system. If one considers a small fluorescent light source (essentially a bright spot), light coming from this spot spreads out further from our perspective as the spot becomes more out of focus. Under ideal conditions, this produces an "hourglass" shape of thispoint source in the third (axial) dimension. This shape is called thepoint spread function (PSF) of the microscope imaging system. Since any fluorescence image is made up of a large number of such small fluorescent light sources, the image is said to be "convolved by the point spread function". The mathematically modeled PSF of a terahertz laser pulsed imaging system is shown on the right.

The output of an imaging system can be described using the equation:

s(x,y)=PSF(x,y)o(x,y)+n{\displaystyle s(x,y)=PSF(x,y)*o(x,y)+n}

Wheren is the additive noise.[32] Knowing this point spread function[33] means that it is possible to reverse this process to a certain extent by computer-based methods commonly known asdeconvolution microscopy.[34] There are various algorithms available for 2D or 3D deconvolution. They can be roughly classified innonrestorative andrestorative methods. While the nonrestorative methods can improve contrast by removing out-of-focus light from focal planes, only the restorative methods can actually reassign light to its proper place of origin. Processing fluorescent images in this manner can be an advantage over directly acquiring images without out-of-focus light, such as images fromconfocal microscopy, because light signals otherwise eliminated become useful information. For 3D deconvolution, one typically provides a series of images taken from different focal planes (called a Z-stack) plus the knowledge of the PSF, which can be derived either experimentally or theoretically from knowing all contributing parameters of the microscope.

Sub-diffraction techniques

[edit]
Main article:Super-Resolution microscopy
Example of super-resolution microscopy. Image ofHer3 andHer2, target of thebreast cancer drugTrastuzumab, within a cancer cell.

A multitude of super-resolution microscopy techniques have been developed in recent times which circumvent thediffraction limit.

This is mostly achieved by imaging a sufficiently static sample multiple times and either modifying the excitation light or observing stochastic changes in the image. The deconvolution methods described in the previous section, which removes the PSF induced blur and assigns a mathematically 'correct' origin of light, are used, albeit with slightly different understanding of what the value of a pixel mean. Assumingmost of the time, one single fluorophore contributes to one single blob on one single taken image, the blobs in the images can be replaced with their calculated position, vastly improving resolution to well below the diffraction limit.

To realize such assumption, Knowledge of and chemical control over fluorophore photophysics is at the core of these techniques, by which resolutions of ~20 nanometers are obtained.[35][36]

Serial time-encoded amplified microscopy

[edit]
Main article:Serial time-encoded amplified microscopy

Serial time encoded amplified microscopy (STEAM) is an imaging method that provides ultrafast shutter speed and frame rate, by using optical image amplification to circumvent the fundamental trade-off between sensitivity and speed, and a single-pixelphotodetector to eliminate the need for a detector array and readout time limitations[37] The method is at least 1000 times faster than the state-of-the-artCCD andCMOS cameras. Consequently, it is potentially useful for scientific, industrial, and biomedical applications that require high image acquisition rates, including real-time diagnosis and evaluation of shockwaves,microfluidics,MEMS, andlaser surgery.[38]

Extensions

[edit]

Most modern instruments provide simple solutions for micro-photography and image recording electronically. However such capabilities are not always present and the more experienced microscopist may prefer a hand drawn image to a photograph. This is because a microscopist with knowledge of the subject can accurately convert a three-dimensional image into a precise two-dimensional drawing. In a photograph or other image capture system however, only one thin plane is ever in good focus.[citation needed]

The creation of accurate micrographs requires a microscopical technique using a monocular eyepiece. It is essential that both eyes are open and that the eye that is not observing down the microscope is instead concentrated on a sheet of paper on the bench besides the microscope. With practice, and without moving the head or eyes, it is possible to accurately trace the observed shapes by simultaneously "seeing" the pencil point in the microscopical image.[citation needed]

It is always less tiring to observe with the microscope focused so that the image is seen at infinity and with both eyes open at all times.[citation needed]

Other enhancements

[edit]
Main article:stereomicroscope

Microspectroscopy:spectroscopy with a microscope

X-ray

[edit]
Main article:X-ray microscopy

As resolution depends on thewavelength of the light. Electron microscopy has been developed since the 1930s that use electron beams instead of light. Because of the much smaller wavelength of the electron beam, resolution is far higher.

Though less common,X-ray microscopy has also been developed since the late 1940s. The resolution of X-ray microscopy lies between that of light microscopy and electron microscopy.

Electron microscopy

[edit]
Main article:Electron microscope

Until the invention of sub-diffraction microscopy, the wavelength of the light limited the resolution of traditional microscopy to around 0.2 micrometers. In order to gain higher resolution, the use of an electron beam with a far smaller wavelength is used in electron microscopes.

  • Transmission electron microscopy (TEM) is quite similar to the compound light microscope, by sending an electron beam through a very thin slice of the specimen. The resolution limit in 2005 was around 0.05[dubiousdiscuss] nanometer and has not increased appreciably since that time.
  • Scanning electron microscopy (SEM) visualizes details on the surfaces of specimens and gives a very nice 3D view. It gives results much like those of the stereo light microscope. The best resolution for SEM in 2011 was 0.4 nanometer.

Electron microscopes equipped forX-ray spectroscopy can provide qualitative and quantitative elemental analysis. This type of electron microscope, also known as analytical electron microscope, can be a very powerful tool for investigation ofnanomaterials.[39]

Scanning probe microscopy

[edit]
Main article:Scanning probe microscopy

This is a sub-diffraction technique. Examples of scanning probe microscopes are theatomic force microscope (AFM), thescanning tunneling microscope, thephotonic force microscope and therecurrence tracking microscope. All such methods use the physical contact of a solid probe tip to scan the surface of an object, which is supposed to be almost flat.

Ultrasonic force

[edit]

Ultrasonic force microscopy (UFM) has been developed in order to improve the details and image contrast on "flat" areas of interest where AFM images are limited in contrast. The combination of AFM-UFM allows a near field acoustic microscopic image to be generated. The AFM tip is used to detect the ultrasonic waves and overcomes the limitation of wavelength that occurs in acoustic microscopy. By using the elastic changes under the AFM tip, an image of much greater detail than the AFM topography can be generated.

Ultrasonic force microscopy allows the local mapping ofelasticity in atomic force microscopy by the application of ultrasonic vibration to the cantilever or sample. To analyze the results of ultrasonic force microscopy in a quantitative fashion, a force-distance curve measurement is done with ultrasonic vibration applied to the cantilever base, and the results are compared with a model of the cantilever dynamics and tip-sample interaction based on the finite-difference technique.

Ultraviolet microscopy

[edit]

Ultraviolet microscopes have two main purposes. The first is to use the shorter wavelength of ultraviolet electromagnetic energy to improve the image resolution beyond that of the diffraction limit of standard optical microscopes. This technique is used for non-destructive inspection of devices with very small features such as those found in modern semiconductors. The second application for UV microscopes is contrast enhancement where the response of individual samples is enhanced, relative to their surrounding, due to the interaction of light with the molecules within the sample itself. One example is in the growth ofprotein crystals. Protein crystals are formed in salt solutions. As salt and protein crystals are both formed in the growth process, and both are commonly transparent to the human eye, they cannot be differentiated with a standard optical microscope. As thetryptophan ofprotein absorbs light at 280 nm, imaging with a UV microscope with 280 nm bandpass filters makes it simple to differentiate between the two types of crystals. The protein crystals appear dark while the salt crystals are transparent.

Infrared microscopy

[edit]

The terminfrared microscopy refers to microscopy performed atinfrared wavelengths. In the typical instrument configuration, aFourier Transform Infrared Spectrometer (FTIR) is combined with an optical microscope and aninfrared detector. The infrared detector can be a single point detector, a linear array or a 2D focal plane array. FTIR provides the ability to perform chemical analysis viainfrared spectroscopy and the microscope and point or array detector enable this chemical analysis to be spatially resolved, i.e. performed at different regions of the sample. As such, the technique is also called infrared microspectroscopy[40][41]An alternative architecture calledLaser Direct Infrared (LDIR) Imaging involves the combination of a tuneable infrared light source and single point detector on a flying objective. This technique is frequently used for infraredchemical imaging, where the image contrast is determined by the response of individual sample regions to particular IR wavelengths selected by the user, usually specific IR absorption bands and associatedmolecular resonances. A key limitation of conventional infrared microspectroscopy is that the spatial resolution isdiffraction-limited. Specifically the spatial resolution is limited to a figure related to the wavelength of the light. For practical IR microscopes, the spatial resolution is limited to 1–3x the wavelength, depending on the specific technique and instrument used. For mid-IR wavelengths, this sets a practical spatial resolution limit of ~3-30 μm.

IR versions of sub-diffraction microscopy also exist.[40][41] These include IRNear-field scanning optical microscope (NSOM),[42]photothermal microspectroscopy andatomic force microscope based infrared spectroscopy (AFM-IR), as well as scattering-type Scanning Near-field Optical Microscopy (s-SNOM)[43] &nano-FTIR that provide nanoscale spatial resolution at IR wavelengths.

Digital holographic microscopy

[edit]
Human cells imaged by DHM phase shift (left) andphase contrast microscopy (right)
Main article:Digital holographic microscopy

Indigital holographic microscopy (DHM), interferingwave fronts from acoherent (monochromatic) light-source are recorded on a sensor. The image is digitally reconstructed by a computer from the recordedhologram. Besides the ordinary bright field image, aphase shift image is created.

DHM can operate both in reflection and transmission mode. In reflection mode, the phase shift image provides a relative distance measurement and thus represents atopography map of the reflecting surface. In transmission mode, the phase shift image provides a label-free quantitative measurement of the optical thickness of the specimen. Phase shift images of biological cells are very similar to images of stained cells and have successfully been analyzed byhigh content analysis software.

A unique feature of DHM is the ability to adjust focus after the image is recorded, since all focus planes are recorded simultaneously by the hologram. This feature makes it possible to image moving particles in a volume or to rapidly scan a surface. Another attractive feature is The ability of DHM to use low cost optics by correcting optical aberrations by software.

Digital pathology (virtual microscopy)

[edit]
Main article:Digital pathology

Digital pathology is an image-based information environment enabled by computer technology that allows for the management of information generated from a digital slide. Digital pathology is enabled in part byvirtual microscopy, which is the practice of converting glass slides into digital slides that can be viewed, managed, and analyzed.

Laser microscopy

[edit]

Laser microscopy is a rapidly growing field that uses laser illumination sources in various forms of microscopy.[44] For instance, laser microscopy focused on biological applications usesultrashort pulse lasers, in a number of techniques labeled as nonlinear microscopy, saturation microscopy, andtwo-photon excitation microscopy.[45]

High-intensity, short-pulse laboratoryx-ray lasers have been under development for several years. When this technology comes to fruition, it will be possible to obtain magnified three-dimensional images of elementary biological structures in the living state at a precisely defined instant. For optimum contrast between water and protein and for best sensitivity and resolution, the laser should be tuned near the nitrogen line at about 0.3 nanometers. Resolution will be limited mainly by the hydrodynamic expansion that occurs while the necessary number of photons is being registered.[46] Thus, while the specimen is destroyed by the exposure, its configuration can be captured before it explodes.[47][48][49][50][51][52][excessive citations]

Scientists have been working on practical designs and prototypes for x-ray holographic microscopes, despite the prolonged development of the appropriate laser.[53][54][55][56][57][58][59][60][excessive citations]

Photoacoustic microscopy

[edit]
Main article:Photoacoustic microscopy
Photoacoustic micrograph of human red blood cells.

A microscopy technique relying on thephotoacoustic effect,[61] i.e. the generation of (ultra)sound caused by light absorption.A focused and intensity modulated laser beam is raster scanned over a sample. The generated (ultra)sound is detected via an ultrasound transducer. Commonly piezoelectric ultrasound transducers are employed.[62]

The image contrast is related to the sample's absorption coefficientα{\displaystyle \alpha }. This is in contrast to bright or dark field microscopy, where the image contrast is due to transmittance or scattering. In principle, the contrast of fluorescence microscopy is proportional to the sample's absorption too. However, in fluorescence microscopy the fluorescence quantum yieldηfl{\displaystyle \eta _{fl}} needs to be unequal to zero in order that a signal can be detected. In photoacoustic microscopy, however, every absorbing substance gives a photoacoustic signalPA{\displaystyle PA} which is proportional to

PAαΓ((1ηfl)Eg+(ElaserEg)){\displaystyle PA\propto \alpha *\Gamma *((1-\eta _{fl})*E_{g}+(E_{laser}-E_{g}))}

HereΓ{\displaystyle \Gamma } is the Grüneisen coefficient,Elaser{\displaystyle E_{laser}} is the laser's photon energy andEg{\displaystyle E_{g}} is the sample's band gap energy. Therefore, photoacoustic microscopy seems well suited as a complementary technique to fluorescence microscopy, as a high fluorescence quantum yield leads to high fluorescence signals and a low fluorescence quantum yield leads to high photoacoustic signals.

Neglecting non-linear effects, the lateral resolutiondx is limited by theAbbe diffraction limit:

dx=λ/(2NA){\displaystyle dx=\lambda /(2*NA)}

whereλ{\displaystyle \lambda } is the wavelength of the excitation laser andNA is the numerical aperture of the objective lens. The Abbe diffraction limit holds if the incoming wave front is parallel. In reality, however, the laser beam profile is Gaussian. Therefore, in order to the calculate the achievable resolution, formulas for truncated Gaussian beams have to be used.[63]

Amateur microscopy

[edit]

Amateur Microscopy is the investigation and observation ofbiological and non-biological specimens for recreational purposes. Collectors ofminerals,insects,seashells, andplants may usemicroscopes as tools to uncover features that help themclassify their collected items. Other amateurs may be interested in observing the life found in pond water and of other samples. Microscopes may also prove useful for the water quality assessment for people that keep ahome aquarium. Photographic documentation and drawing of the microscopic images are additional pleasures. There are competitions forphotomicrograph art. Participants of this pastime may use commercially prepared microscopic slides or prepare their own slides.

While microscopy is a central tool in the documentation of biological specimens, it is often insufficient to justify the description of a new species based on microscopic investigations alone. Often genetic and biochemical tests are necessary to confirm the discovery of a new species. Alaboratory and access to academic literature is a necessity. There is, however, one advantage that amateurs have above professionals: time to explore their surroundings. Often, advanced amateurs team up with professionals to validate their findings and possibly describe new species.

In the late 1800s, amateur microscopy became a popular hobby in the United States and Europe. Several 'professional amateurs' were being paid for their sampling trips and microscopic explorations by philanthropists, to keep them amused on the Sunday afternoon (e.g., the diatom specialist A. Grunow, being paid by (among others) a Belgian industrialist). ProfessorJohn Phin published "Practical Hints on the Selection and Use of the Microscope (Second Edition, 1878)," and was also the editor of the "American Journal of Microscopy."

Examples of amateur microscopy images:

  • "house bee" Mouth 100X
    "house bee" Mouth 100X
  • Rice Stem cs 400X
    Rice Stem cs 400X
  • Rabbit Testis 100X
    Rabbit Testis 100X
  • Fern Prothallium 400X
    Fern Prothallium 400X

Application in forensic science

[edit]

Microscopy has applications in the forensic sciences.[64] The microscope can detect, resolve and image the smallest items of evidence, often without any alteration or destruction. The microscope is used to identify and compare fibers, hairs, soils, and dust...etc.

In ink markings, blood stains or bullets, no specimen treatment is required and the evidence shows directly from microscopical examination. For traces of particular matter, the sample preparation must be done before microscopical examination occurs.[clarification needed]

Light microscopes are the most use in forensics, using photons to form images, microscopes which are most applicable for examining forensic specimens are as follows:[65]

1. The compound microscope

2. The comparison microscope

3. The stereoscopic microscope

4. The polarizing microscope

5. The micro spectrophotometer

This diversity of the types of microscopes in forensic applications comes mainly from their magnification ranges, which are (1- 1200X), (50 -30,000X) and (500- 250,000X) for the optical microscopy, SEM and TEM respectively.[65]

See also

[edit]

References

[edit]
  1. ^The University of Edinburgh (March 6, 2018)."What is Microscopy?".The University of Edinburgh.Archived from the original on April 10, 2018. RetrievedApril 9, 2018.
  2. ^Atti Della Fondazione Giorgio Ronchi E Contributi Dell'Istituto Nazionale Di Ottica, Volume 30, La Fondazione-1975, page 554
  3. ^Albert Van Helden; Sven Dupré; Rob van Gent (2010).The Origins of the Telescope. Amsterdam University Press. p. 24.ISBN 978-90-6984-615-6.Archived from the original on 15 February 2017.
  4. ^William Rosenthal, Spectacles and Other Vision Aids: A History and Guide to Collecting, Norman Publishing, 1996, page 391 - 392
  5. ^Robert D. Huerta, Giants of Delft: Johannes Vermeer and the Natural Philosophers : the Parallel Search for Knowledge During the Age of Discovery, Bucknell University Press - 2003, page 126
  6. ^A. Mark Smith, From Sight to Light: The Passage from Ancient to Modern Optics, University of Chicago Press - 2014, page 387
  7. ^Raymond J. Seeger, Men of Physics: Galileo Galilei, His Life and His Works, Elsevier - 2016, page 24
  8. ^J. William Rosenthal, Spectacles and Other Vision Aids: A History and Guide to Collecting, Norman Publishing, 1996, page 391
  9. ^Ford, Brian J. (1992)."From Dilettante to Diligent Experimenter: a Reappraisal of Leeuwenhoek as microscopist and investigator".Biology History.5 (3).Archived from the original on 2021-04-19. Retrieved2021-04-16.
  10. ^Lane, Nick (6 March 2015). "The Unseen World: Reflections on Leeuwenhoek (1677) 'Concerning Little Animal'."Philos Trans R Soc Lond B Biol Sci. 2015 Apr; 370 (1666): 20140344. [doi:10.1098/rstb.2014.0344]
  11. ^Abramowitz M, Davidson MW (2007)."Introduction to Microscopy".Molecular Expressions.Archived from the original on 2021-04-27. Retrieved2007-08-22.
  12. ^"Microscope - Compound, Numerical Aperture, and Immersion Microscopy | Britannica".www.britannica.com.Archived from the original on 2023-06-22. Retrieved2023-06-22.
  13. ^Abbe, E. (1874) A contribution to the theory of the microscope and the nature of microscopic vision. Proc. Bristol Nat. Soc. 1, 200–261.
  14. ^Pawley JB (editor) (2006). Handbook of Biological Confocal Microscopy (3rd ed.). Berlin: Springer.ISBN 0-387-25921-X.
  15. ^Lakowicz, Joseph R. (1999) Principles Of Fluorescence Spectroscopy. New York: Kluwer Academic/Plenum.
  16. ^Abramowitz M, Davidson MW (2003-08-01)."Darkfield Illumination".Archived from the original on 2015-11-23. Retrieved2008-10-21.
  17. ^Abramowitz M, Davidson MW (2003-08-01)."Rheinberg Illumination".Archived from the original on 2008-09-29. Retrieved2008-10-21.
  18. ^Gill, Harindarpal (January 2010)."Evaluating the efficacy of tryptophan fluorescence and absorbance as a selection tool for identifying protein crystals".Acta Crystallographica.F66 (Pt 3):364–372.Bibcode:2010AcCrF..66..364G.doi:10.1107/S1744309110002022.PMC 2833058.PMID 20208182.
  19. ^Denk, Winfried; Svoboda, Karel (March 1997)."Photon Upmanship: Why Multiphoton Imaging Is More than a Gimmick".Neuron.18 (3):351–357.doi:10.1016/S0896-6273(00)81237-4.PMID 9115730.S2CID 2414593.
  20. ^Denk, W.; Delaney, K.R.; Gelperin, A.; Kleinfeld, D.; Strowbridge, B.W.; Tank, D.W.; Yuste, R. (1994). "Anatomical and functional imaging of neurons using 2-photon laser scanning microscopy".Journal of Neuroscience Methods.54 (2):151–162.doi:10.1016/0165-0270(94)90189-9.PMID 7869748.S2CID 3772937.
  21. ^Svoboda, Karel; Denk, Winfried; Kleinfeld, David; Tank, David W. (1997)."In vivo dendritic calcium dynamics in neocortical pyramidal neurons".Nature.385 (6612):161–165.Bibcode:1997Natur.385..161S.doi:10.1038/385161a0.ISSN 0028-0836.PMID 8990119.S2CID 4251386.Archived from the original on 2021-05-25. Retrieved2020-05-20.
  22. ^Voie, A.H. (1993). "Imaging the intact guinea pig tympanic bulla by orthogonal-plane fluorescence optical sectioning microscopy".Hearing Research.71 (1–2):119–128.doi:10.1016/S0378-5955(02)00493-8.ISSN 0378-5955.PMID 12204356.S2CID 12775304.
  23. ^Greger, K.; J. Swoger; E. H. K. Stelzer (2007)."Basic building units and properties of a fluorescence single plane illumination microscope".Review of Scientific Instruments.78 (2): 023705–023705–7.Bibcode:2007RScI...78b3705G.doi:10.1063/1.2428277.ISSN 0034-6748.PMID 17578115.
  24. ^Buytaert, J.A.N.; E. Descamps; D. Adriaens; J.J.J. Dirckx (2012)."The OPFOS Microscopy Family: High-Resolution Optical Sectioning of Biomedical Specimens".Anatomy Research International.2012: 206238.arXiv:1106.3162.doi:10.1155/2012/206238.ISSN 2090-2743.PMC 3335623.PMID 22567307.
  25. ^F. J. Duarte, inHigh Power Dye Lasers (Springer-Verlag, Berlin,1991) Chapter 2.
  26. ^Duarte FJ (1993), Electro-optical interferometric microdensitometer system,US Patent 5255069Archived 2017-10-13 at theWayback Machine.
  27. ^abPeterson, Mark D.; Hayes, Patrick L.; Martinez, Imee Su; Cass, Laura C.; Achtyl, Jennifer L.; Weiss, Emily A.; Geiger, Franz M. (2011-05-01). "Second harmonic generation imaging with a kHz amplifier [Invited]".Optical Materials Express.1 (1): 57.Bibcode:2011OMExp...1...57P.doi:10.1364/ome.1.000057.
  28. ^abcMacias-Romero, Carlos; Didier, Marie E. P.; Jourdain, Pascal; Marquet, Pierre; Magistretti, Pierre; Tarun, Orly B.; Zubkovs, Vitalijs;Radenovic, Aleksandra; Roke, Sylvie (2014-12-15)."High throughput second harmonic imaging for label-free biological applications".Optics Express.22 (25):31102–12.Bibcode:2014OExpr..2231102M.doi:10.1364/oe.22.031102.hdl:10754/563269.PMID 25607059.Archived from the original on 2020-06-20. Retrieved2019-12-11.
  29. ^abCheng, Li-Chung; Chang, Chia-Yuan; Lin, Chun-Yu; Cho, Keng-Chi; Yen, Wei-Chung; Chang, Nan-Shan; Xu, Chris; Dong, Chen Yuan; Chen, Shean-Jen (2012-04-09)."Spatiotemporal focusing-based widefield multiphoton microscopy for fast optical sectioning".Optics Express.20 (8):8939–48.Bibcode:2012OExpr..20.8939C.doi:10.1364/oe.20.008939.PMID 22513605.
  30. ^Oron, Dan; Tal, Eran; Silberberg, Yaron (2005-03-07)."Scanningless depth-resolved microscopy".Optics Express.13 (5):1468–76.Bibcode:2005OExpr..13.1468O.doi:10.1364/opex.13.001468.PMID 19495022.
  31. ^Ahi, Kiarash; Anwar, Mehdi (2016-05-26)."Modeling of terahertz images based on x-ray images: a novel approach for verification of terahertz images and identification of objects with fine details beyond terahertz resolution". In Anwar, Mehdi F; Crowe, Thomas W; Manzur, Tariq (eds.).Terahertz Physics, Devices, and Systems X: Advanced Applications in Industry and Defense. Proceedings of SPIE. Vol. 9856. p. 985610.Bibcode:2016SPIE.9856E..10A.doi:10.1117/12.2228685.S2CID 124315172.
  32. ^Solomon, Chris (2010).Fundamentals of Digital Image Processing. John Wiley & Sons, Ltd.ISBN 978-0-470-84473-1.
  33. ^Nasse M. J.; Woehl J. C. (2010). "Realistic modeling of the illumination point spread function in confocal scanning optical microscopy".J. Opt. Soc. Am. A.27 (2):295–302.Bibcode:2010JOSAA..27..295N.doi:10.1364/JOSAA.27.000295.PMID 20126241.
  34. ^Wallace W, Schaefer LH, Swedlow JR (2001)."A workingperson's guide to deconvolution in light microscopy".BioTechniques.31 (5):1076–8, 1080, 1082 passim.doi:10.2144/01315bi01.PMID 11730015.
  35. ^Kaufmann Rainer; Müller Patrick; Hildenbrand Georg; Hausmann Michael; Cremer Christoph (2010). "Analysis of Her2/neu membrane protein clusters in different types of breast cancer cells using localization microscopy".Journal of Microscopy.242 (1):46–54.CiteSeerX 10.1.1.665.3604.doi:10.1111/j.1365-2818.2010.03436.x.PMID 21118230.S2CID 2119158.
  36. ^van de Linde S.; Wolter S.; Sauer S. (2011)."Single-molecule Photoswitching and Localization".Aust. J. Chem.64 (5):503–511.doi:10.1071/CH10284.
  37. ^K. Goda; K. K. Tsia; B. Jalali (2009). "Serial time-encoded amplified imaging for real-time observation of fast dynamic phenomena".Nature.458 (7242):1145–9.Bibcode:2009Natur.458.1145G.doi:10.1038/nature07980.PMID 19407796.S2CID 4415762.
  38. ^Jalali, Bahram; Capewell, Dale; Goda, Keisuke; Tsia, Kevin K. (2010-05-10). "Performance of serial time-encoded amplified microscope".Optics Express.18 (10):10016–10028.Bibcode:2010OExpr..1810016T.doi:10.1364/OE.18.010016.hdl:10722/91333.ISSN 1094-4087.PMID 20588855.S2CID 8077381.
  39. ^Kosasih, Felix Utama;Ducati, Caterina (May 2018)."Characterising degradation of perovskite solar cells through in-situ and operando electron microscopy".Nano Energy.47:243–256.Bibcode:2018NEne...47..243K.doi:10.1016/j.nanoen.2018.02.055.Archived from the original on 2021-01-26. Retrieved2019-06-28.
  40. ^abH M Pollock and S G Kazarian, Microspectroscopy in the Mid-Infrared, in Encyclopedia of Analytical Chemistry (Robert A. Meyers, Ed, 1-26 (2014), John Wiley & Sons Ltd,
  41. ^abPollock Hubert M (2014). "Microspectroscopy in the Mid-Infrared".Encyclopedia of Analytical Chemistry. pp. 1–26.doi:10.1002/9780470027318.a5609.pub2.ISBN 9780470027318.
  42. ^H M Pollock and D A Smith, The use of near-field probes for vibrational spectroscopy and photothermal imaging, in Handbook of vibrational spectroscopy, J.M. Chalmers and P.R. Griffiths (eds), John Wiley & Sons Ltd, Vol. 2, pp. 1472 - 1492 (2002)
  43. ^Keilmann, Fritz; Hillenbrand, Rainer (2004-04-15). Richards, David; Zayats, Anatoly (eds.)."Near-field microscopy by elastic light scattering from a tip".Philosophical Transactions of the Royal Society of London. Series A: Mathematical, Physical and Engineering Sciences.362 (1817):787–805.doi:10.1098/rsta.2003.1347.ISSN 1364-503X.PMID 15306494.S2CID 20211405.Archived from the original on 2020-11-09. Retrieved2020-11-02.
  44. ^Duarte FJ (2016). "Tunable laser microscopy". In Duarte FJ (ed.).Tunable Laser Applications (3rd ed.). Boca Raton:CRC Press. pp. 315–328.ISBN 9781482261066.
  45. ^Thomas JL, Rudolph W (2008). "Biological Microscopy with Ultrashort Laser Pulses". In Duarte FJ (ed.).Tunable Laser Applications (2nd ed.). Boca Raton:CRC Press. pp. 245–80.ISBN 978-1-4200-6009-6.
  46. ^Solem, J. C. (1983). "X-ray imaging on biological specimens".Proceedings of International Conference on Lasers '83:635–640.OSTI 5998195.
  47. ^Solem, J. C. (1982)."High-intensity x-ray holography: An approach to high-resolution snapshot imaging of biological specimens".Los Alamos National Laboratory Technical Report LA-9508-MS.83: 23581.Bibcode:1982STIN...8323581S.doi:10.2172/7056325.OSTI 7056325.Archived from the original on 2020-08-04. Retrieved2019-06-28.
  48. ^Solem, J. C.; Baldwin, G. C. (1982). "Microholography of living organisms".Science.218 (4569):229–235.Bibcode:1982Sci...218..229S.doi:10.1126/science.218.4569.229.PMID 17838608.S2CID 32381820.
  49. ^Solem, J. C.; Chapline, G. F. (1984). "X-ray biomicroholography".Optical Engineering.23 (2): 193.Bibcode:1984OptEn..23..193S.doi:10.1117/12.7973410.
  50. ^Solem, J. C. (1985). "Microholography".McGraw-Hill Encyclopedia of Science and Technology.
  51. ^Solem, J. C. (1984). "X-ray holography of biological specimens".Proceedings of Ninth International Congress on Photobiology, Philadelphia, PA, USA, 3 July 1984 (LA-UR-84-3340, CONF-840783-3): 19.OSTI 6314558.
  52. ^Solem, J. C. (1986). "Imaging biological specimens with high-intensity soft X-rays".Journal of the Optical Society of America B.3 (11):1551–1565.Bibcode:1986JOSAB...3.1551S.doi:10.1364/josab.3.001551.
  53. ^Haddad, W. S.; Solem, J. C.; Cullen, D.; Boyer, K.; Rhodes, C. K. (1987). "A description of the theory and apparatus for digital reconstruction of Fourier transform holograms",Proceedings of Electronics Imaging '87, Nov. 2-5, 1987, Boston; Journal of Electronic Imaging, (Institute for Graphic Communication, Inc., Boston), p. 693.
  54. ^Haddad, W. S.; Cullen, D.; Boyer, K.; Rhodes, C. K.; Solem, J. C.; Weinstein, R. S. (1988). "Design for a Fourier-Transform Holographic Microscope".X-Ray Microscopy II. Springer Series in Optical Sciences. Vol. 56. pp. 284–287.doi:10.1007/978-3-540-39246-0_49.ISBN 978-3-662-14490-9.
  55. ^Haddad, W. S.; Solem, J. C.; Cullen, D.; Boyer, K.; Rhodes, C. K. (1988)."Design for a Fourier-transform holographic microscope incorporating digital image reconstruction".Proceedings of CLEO '88 Conference on Lasers and Electro-Optics, Anaheim, CA, April, 1988; Optical Society of America, OSA Technical Digest.7: WS4.Archived from the original on 2016-02-02. Retrieved2016-01-13.
  56. ^Haddad, W. S.; Cullen, D.; Solem, J. C.; Boyer, K.; Rhodes, C. K. (1988)."X-ray Fourier-transform holographic microscope".Proceedings of the OSA Topical Meeting on Short Wavelength Coherent Radiation: Generation and Applications, September 26–29, 1988, Cape Cod, MA., Falcone, R.; Kirz, J.; Eds. (Optical Society of America, Washington, DC):284–289.Archived from the original on 2016-01-27. Retrieved2016-01-13.
  57. ^Solem, J. C.; Boyer, K.; Haddad, W. S.; Rhodes, C. K. (1990). Prosnitz, Donald (ed.). "Prospects for X-ray holography with free-electron lasers".Free Electron Lasers and Applications. Free-Electron Lasers and Applications.1227. Prosnitz, D.; ed. SPIE:105–116.Bibcode:1990SPIE.1227..105S.doi:10.1117/12.18609.S2CID 121807907.
  58. ^Haddad, W. S.; Cullen, D.; Solem, J. C.; Longworth, J. W.; McPherson, L. A.; Boyer, K.; Rhodes, C. K. (1991). "Fourier-transform holographic microscope". In Chang, Winchyi; Milch, James R. (eds.).Camera and Input Scanner Systems. Vol. 1448. pp. 81–88.Bibcode:1991SPIE.1448...81H.doi:10.1117/12.45347.PMID 20733659.S2CID 120679508.{{cite book}}:|journal= ignored (help)
  59. ^Haddad, W. S.; Cullen, D.; Solem, J. C.; Longworth, J.; McPherson, A.; Boyer, K.; Rhodes, C. K. (1992). "Fourier-transform holographic microscope".Applied Optics.31 (24):4973–4978.Bibcode:1992ApOpt..31.4973H.doi:10.1364/ao.31.004973.PMID 20733659.
  60. ^Boyer, K.; Solem, J. C.; Longworth, J.; Borisov, A.; Rhodes, C. K. (1996). "Biomedical three-dimensional holographic microimaging at visible, ultraviolet and X-ray wavelengths".Nature Medicine.2 (8):939–941.doi:10.1038/nm0896-939.PMID 8705867.S2CID 25231033.
  61. ^Bell, A. G. (1880)."On the production and reproduction of sound by light".American Journal of Science. s3-20 (118):305–324.Bibcode:1880AmJS...20..305B.doi:10.2475/ajs.s3-20.118.305.S2CID 130048089.Archived from the original on 2022-11-27. Retrieved2019-06-28.
  62. ^Yao, J.; Wang, L. V. (2013)."Photoacoustic Microscopy".Laser Photonics Rev.7 (5):1–36.Bibcode:2013LPRv....7..758Y.doi:10.1002/lpor.201200060.PMC 3887369.PMID 24416085.
  63. ^Langer, G.; Buchegger, B.; Jacak, J.; Klar, T. A.; Berer, T. (2016)."Frequency domain photoacoustic and fluorescence microscopy".Biomedical Optics Express.7 (7):2692–702.doi:10.1364/BOE.7.002692.PMC 4948622.PMID 27446698.
  64. ^Kotrly M (August 2015)."New Possibilities of Using Micrscopic Techniques in Forensic Field".Proceedings of Microscopy & Microanallysis.21 (S3):1365–1366.Bibcode:2015MiMic..21S1365K.doi:10.1017/S1431927615007618.
  65. ^abBasu S, Millette JR (1986).Electron Microscopy in Forensic Occupational and Environmental Health Sciences. New York: Plenum Press.

Further reading

[edit]

External links

[edit]
Wikimedia Commons has media related toMicroscopy.

General

[edit]

Techniques

[edit]
Illumination and
contrast methods
Fluorescence methods
Sub-diffraction
limit techniques
Instrumentation
Techniques
Sampling
Calibration
Prominentpublications
National
Other
Retrieved from "https://en.wikipedia.org/w/index.php?title=Microscopy&oldid=1296291703"
Categories:
Hidden categories:

[8]ページ先頭

©2009-2025 Movatter.jp