Movatterモバイル変換


[0]ホーム

URL:


Jump to content
WikipediaThe Free Encyclopedia
Search

Energy-dispersive X-ray spectroscopy

From Wikipedia, the free encyclopedia
This articleneeds additional citations forverification. Please helpimprove this article byadding citations to reliable sources. Unsourced material may be challenged and removed.
Find sources: "Energy-dispersive X-ray spectroscopy" – news ·newspapers ·books ·scholar ·JSTOR
(July 2022) (Learn how and when to remove this message)
Analytical technique used for the elemental analysis or chemical characterization of a sample
"EDXA" redirects here. For the German airport, seeAchmer Aerodrome.
EDS spectrum of the mineral crust of the vent shrimpRimicaris exoculata.[1] Most of these peaks are X-rays emitted when electrons return to the K electron shell (K-alpha andK-beta lines). One peak is from the L shell of iron.

Energy-dispersive X-ray spectroscopy (EDS,EDX,EDXS orXEDS), sometimes calledenergy dispersive X-ray analysis (EDXA orEDAX) orenergy dispersive X-ray microanalysis (EDXMA), is an analytical technique used for theelemental analysis orchemical characterization of asample. It relies on an interaction of somesource ofX-ray excitation and a sample. Its characterization capabilities are due in large part to the fundamental principle that each element has a uniqueatomic structure allowing a unique set of peaks on its electromagneticemission spectrum[2] (which is the main principle ofspectroscopy). The peak positions are predicted by theMoseley's law with accuracy much better than experimental resolution of a typical EDX instrument.

To stimulate the emission of characteristic X-rays from a specimen a beam of electrons or X-ray is focused into the sample being studied. At rest, an atom within the sample containsground state (or unexcited) electrons in discrete energy levels orelectron shells bound to the nucleus. The incident beam may excite an electron in an inner shell, ejecting it from the shell while creating anelectron hole where the electron was. An electron from an outer, higher-energy shell then fills the hole, and the difference in energy between the higher-energy shell and the lower energy shell may be released in the form of an X-ray. The number and energy of the X-rays emitted from a specimen can be measured by an energy-dispersive spectrometer. As the energies of the X-rays are characteristic of the difference in energy between the two shells and of the atomic structure of the emitting element, EDS allows the elemental composition of the specimen to be measured.[2]

Equipment

[edit]

Four primary components of the EDS setup are

  1. the excitation source (electron beam or x-ray beam)
  2. theX-ray detector
  3. the pulse processor
  4. the analyzer.[citation needed]

Electron beam excitation is used inelectron microscopes,scanning electron microscopes (SEM) andscanning transmission electron microscopes (STEM). X-ray beam excitation is used inX-ray fluorescence (XRF) spectrometers. A detector is used to convert X-ray energy intovoltage signals; this information is sent to a pulse processor, which measures the signals and passes them onto an analyzer for data display and analysis.[citation needed] The most common detector used to be aSi(Li) detector cooled to cryogenic temperatures withliquid nitrogen. Now, newer systems are often equipped withsilicon drift detectors (SDD) withPeltier cooling systems.

Hazards and Safety

  1. High Voltage: SEM-EDX operates at high voltages (typically several kilovolts), which can pose a risk of electric shock.
  2. X-ray Radiation: While SEM-EDX does not use as high a voltage as some X-ray techniques, it still produces X-rays that can be harmful with prolonged exposure. Proper shielding and safety measures are necessary.
  3. Sample Preparation: Handling and preparation of samples can involve hazardous chemicals or materials. Proper personal protective equipment (PPE) should be used.
  4. Vacuum System: The vacuum system used in SEM-EDX can implode if not properly maintained, leading to potential hazards.
  5. Cryogenic Hazards: Some samples may require cryogenic techniques for analysis, which can pose risks of cold burns or asphyxiation if not appropriately handled.
  6. Mechanical Hazards: If used incorrectly, moving parts in the SEM can cause injury.
  7. Fire and Explosion Risks: Some samples, particularly those involving flammable materials, can pose fire or explosion risks under vacuum conditions.
  8. Ergonomic Risks: Prolonged use of SEM-EDX can lead to ergonomic hazards if the workstation is not correctly set up for the user's comfort and safety.

Technological variants

[edit]
Principle of EDS

The excess energy of the electron that migrates to an inner shell to fill the newly created hole can do more than emit an X-ray.[3] Often, instead of X-ray emission, the excess energy is transferred to a third electron from a further outer shell, prompting its ejection. This ejected species is called anAuger electron, and the method for its analysis is known asAuger electron spectroscopy (AES).[3]

X-ray photoelectron spectroscopy (XPS) is another close relative of EDS, utilizing ejected electrons in a manner similar to that of AES. Information on the quantity andkinetic energy of ejected electrons is used to determine thebinding energy of these now-liberated electrons, which is element-specific and allows chemical characterization of a sample.[citation needed]

EDS is often contrasted with its spectroscopic counterpart,wavelength dispersive X-ray spectroscopy (WDS). WDS differs from EDS in that it uses thediffraction of X-rays on special crystals to separate its raw data into spectral components (wavelengths). WDS has a much finer spectral resolution than EDS. WDS also avoids the problems associated with artifacts in EDS (false peaks, noise from the amplifiers, andmicrophonics).

A high-energy beam of charged particles such aselectrons orprotons can be used to excite a sample rather than X-rays. This is calledparticle-induced X-ray emission or PIXE.

Accuracy

[edit]

EDS can be used to determine which chemical elements are present in a sample, and can be used to estimate their relative abundance. EDS also helps to measure multi-layer coating thickness of metallic coatings and analysis of various alloys. The accuracy of this quantitative analysis of sample composition is affected by various factors. Many elements will have overlapping X-ray emission peaks (e.g., Ti Kβ and V Kα, Mn Kβ and Fe Kα). The accuracy of the measured composition is also affected by the nature of the sample. X-rays are generated by any atom in the sample that is sufficiently excited by the incoming beam. These X-rays are emitted in all directions (isotropically), and so they may not all escape the sample. The likelihood of an X-ray escaping the specimen, and thus being available to detect and measure, depends on the energy of the X-ray and the composition, amount, and density of material it has to pass through to reach the detector. Because of this X-ray absorption effect and similar effects, accurate estimation of the sample composition from the measured X-ray emission spectrum requires the application of quantitative correction procedures, which are sometimes referred to as matrix corrections.[2]

Emerging technology

[edit]

There is a trend towards a newer EDS detector, called thesilicon drift detector (SDD). The SDD consists of a high-resistivity silicon chip where electrons are driven to a small collecting anode. The advantage lies in the extremely low capacitance of this anode, thereby utilizing shorter processing times and allowing very high throughput. Benefits of the SDD include:[4]

  1. High count rates and processing,
  2. Better resolution than traditional Si(Li) detectors at high count rates,
  3. Lower dead time (time spent on processing X-ray event),
  4. Faster analytical capabilities and more precise X-ray maps or particle data collected in seconds,
  5. Ability to be stored and operated at relatively high temperatures, eliminating the need forliquid nitrogen cooling.

Because the capacitance of the SDD chip is independent of the active area of the detector, much larger SDD chips can be utilized (40 mm2 or more). This allows for even higher count rate collection. Further benefits of large area chips include:[citation needed]

  1. Minimizing SEM beam current allowing for optimization of imaging under analytical conditions,
  2. Reduced sample damage[5] and
  3. Smaller beam interaction and improved spatial resolution for high speed maps.

Where the X-ray energies of interest are in excess of ~ 30 keV, traditional silicon-based technologies suffer from poor quantum efficiency due to a reduction in the detectorstopping power. Detectors produced from high density semiconductors such ascadmium telluride (CdTe) andcadmium zinc telluride (CdZnTe) have improved efficiency at higher X-ray energies and are capable of room temperature operation. Single element systems, and more recently pixelated imaging detectors such as thehigh energy X-ray imaging technology (HEXITEC) system, are capable of achieving energy resolutions of the order of 1% at 100 keV.

In recent years, a different type of EDS detector, based upon a superconductingmicrocalorimeter, has also become commercially available. This new technology combines the simultaneous detection capabilities of EDS with the high spectral resolution of WDS. The EDS microcalorimeter consists of two components: an absorber, and a superconductingtransition-edge sensor (TES)thermometer. The former absorbs X-rays emitted from the sample and converts this energy into heat; the latter measures the subsequent change in temperature due to the influx of heat. The EDS microcalorimeter has historically suffered from a number of drawbacks, including low count rates and small detector areas. The count rate is hampered by its reliance on thetime constant of the calorimeter's electrical circuit. The detector area must be small in order to keep theheat capacity small and maximize thermal sensitivity (resolution). However, the count rate and detector area have been improved by the implementation of arrays of hundreds of superconducting EDS microcalorimeters, and the importance of this technology is growing.

See also

[edit]

References

[edit]
  1. ^Corbari, L; et al. (2008)."Iron oxide deposits associated with the ectosymbiotic bacteria in the hydrothermal vent shrimp Rimicaris exoculata".Biogeosciences.5 (5):1295–1310.Bibcode:2008BGeo....5.1295C.doi:10.5194/bg-5-1295-2008.
  2. ^abcJoseph Goldstein (2003).Scanning Electron Microscopy and X-Ray Microanalysis. Springer.ISBN 978-0-306-47292-3. Retrieved26 May 2012.
  3. ^abJenkins, R. A.; De Vries, J. L. (1982).Practical X-Ray Spectrometry. Springer.ISBN 978-1-468-46282-1.
  4. ^"Technology".
  5. ^Kosasih, Felix Utama; Cacovich, Stefania; Divitini, Giorgio; Ducati, Caterina (17 November 2020)."Nanometric Chemical Analysis of Beam-Sensitive Materials: A Case Study of STEM-EDX on Perovskite Solar Cells".Small Methods.5 (2): 2000835.doi:10.1002/smtd.202000835.PMID 34927887.

External links

[edit]
Vibrational (IR)
UV–Vis–NIR "Optical"
X-ray and Gamma ray
Electron
Nucleon
Radiowave
Others
Data collection, processing
Measured phenomena
Applications
Basics
Electron interaction
with matter
Instrumentation
Microscopes
EM
SEM
TEM
Techniques
Others
Developers
Manufacturers
Software
Characteristics
Sources and instruments
Interaction with matter
Applications
Imaging
Spectroscopy
Scattering
Others
Retrieved from "https://en.wikipedia.org/w/index.php?title=Energy-dispersive_X-ray_spectroscopy&oldid=1289321432"
Categories:
Hidden categories:

[8]ページ先頭

©2009-2025 Movatter.jp