Movatterモバイル変換


[0]ホーム

URL:


Jump to content
WikipediaThe Free Encyclopedia
Search

Cryogenic electron microscopy

From Wikipedia, the free encyclopedia
(Redirected fromCryo-EM)
Electron microscopy technique
This articlemay requirecopy editing for grammar and sentence structure. You can assist byediting it.(November 2025) (Learn how and when to remove this message)
Titan Krios at theUniversity of Leeds

Cryogenic electron microscopy (cryo-EM) is atransmission electron microscopy technique applied to samples cooled tocryogenic temperatures. For biological specimens, the structure is preserved by embedding in an environment ofvitreous ice. An aqueous sample solution is applied to a grid-mesh and plunge-frozen in liquidethane or a mixture of liquid ethane andpropane.[1] While development of the technique began in the 1970s, recent advances in detector technology and software algorithms have allowed for the determination of biomolecular structures at near-atomic resolution.[2] This has attracted wide attention to the approach as an alternative toX-ray crystallography orNMR spectroscopy in thestructural biology field.[3]

In 2017, theNobel Prize in Chemistry was awarded toJacques Dubochet,Joachim Frank, andRichard Henderson "for developing cryo-electron microscopy for the high-resolution structure determination of biomolecules in solution."[4]Nature Methods also named cryo-EM as the "Method of the Year" in 2015.[5]

History

[edit]

Early development

[edit]

In the 1960s, the use oftransmission electron microscopy for structure determination of biological samples was limited because of the radiation damage due to high energy electron beams. Scientists hypothesized that examining specimens at low temperatures would reduce beam-induced radiation damage.[6] Bothliquid helium (−269 °C or 4 K or −452.2 °F) andliquid nitrogen (−195.79 °C or 77 K or −320 °F) were considered as cryogens,[7] however high stability was never achieved. In 1980,Erwin Knapek andJacques Dubochet published comments on beam damage at cryogenic temperatures sharing observations that:

Thin crystals mounted on carbon film were found to be from 30 to 300 times more beam-resistant at 4 K than at room temperature... Most of our results can be explained by assuming that cryoprotection in the region of 4 K is strongly dependent on the temperature.[8]

However, these results were not reproducible and just two years later amendments were published,[9] along with a commentary inNature,[10] indicating that the beam resistance was less significant than initially anticipated. The protection gained at 4 K was closer to "tenfold for standard samples of L-valine", than what was previously stated.[10] While cryo-EM samples are routinely collected at liquid nitrogen temperatures,[11] work has continued to understand sample behavior and collect at liquid helium temperatures.[12][13][11]

In 1981,Alasdair McDowall and Jacques Dubochet, scientists at theEuropean Molecular Biology Laboratory, reported the first successful implementation of cryo-EM.[14] McDowall and Dubochetvitrified pure water in a thin film by spraying it onto a hydrophilic carbon film that was rapidly plunged intocryogen (liquidpropane or liquidethane cooled to 77 K). The thin layer ofamorphous ice was less than 1 μm thick and anelectron diffraction pattern confirmed the presence of amorphous/vitreous ice. In 1984, Dubochet's group demonstrated the power of cryo-EM instructural biology with analysis ofvitrifiedadenovirus type 2,T4 bacteriophage,Semliki Forest virus,Bacteriophage CbK, andVesicular-Stomatitis-Virus.[15] This paper is generally considered to mark the origin of Cryo-EM, and the technique has been developed to the point of becoming routine at numerous laboratories throughout the world.

The energy of the electrons used for imaging (80–300 kV) can lead to the breaking ofcovalent bonds in organic and biological samples.[16] When imaging such specimens it is necessary to limit the electron exposure used to acquire the image. These low exposures require that the images of thousands or even millions of identical frozen molecules be selected, aligned, and averaged to obtain high-resolution maps, using specialized software. A significant improvement in structural features was achieved in 2012 by the introduction ofdirect electron detectors and better computational algorithms.[17]

Recent advancements

[edit]

Advances inelectron detector technology, particularly Direct Electron Detectors, as well as more powerful software imaging algorithms have allowed for the determination of macromolecular structures at near-atomic resolution.[18] Imaged macromolecules includeviruses,ribosomes,mitochondria,ion channels, andenzyme complexes. Starting in 2018, cryo-EM could be applied to structures as small ashemoglobin (64kDa)[19] and with resolutions up to 1.8Å.[20] In 2019, cryo-EM structures represented 2.5% of structures deposited in theProtein Data Bank,[21] and this number continues to grow.[22] An application of cryo-EM iscryo-electron tomography (cryo-ET), where a 3D reconstruction of the sample is created from tilted 2D images.

The 2010s were marked with drastic advancements of electron cameras. Notably, the improvements made todirect electron detectors have led to a "resolution revolution"[23] pushing the resolution barrier beneath the crucial ~2-3 Å limit to resolve amino acid position and orientation.[24]

Henderson (MRC Laboratory of Molecular Biology, Cambridge, UK) formed a consortium with engineers at theRutherford Appleton Laboratory and scientists at theMax Planck Society to fund and develop a first prototype. The consortium then joined forces with the electron microscope manufacturerFEI to roll out and market the new design. At about the same time, Gatan Inc. of Pleasanton, California came out with a similar detector designed by Peter Denes (Lawrence Berkeley National Laboratory) andDavid Agard (University of California, San Francisco). A third type of camera was developed byNguyen-Huu Xuong at the Direct Electron company (San Diego, California).[23]

More recently, advancements in the use of protein-based imaging scaffolds are helping to solve the problems of sample orientation bias and size limit. Though the minimum size for Cryo-EM remains undetermined,proteins smaller than ~50kDa generally have too low asignal-to-noise ratio (SNR) to be able to resolve protein particles in the image, making 3D reconstruction difficult or impossible.[25][26] Multiple techniques have been reported to improve SNR when determining the structures of small proteins.[27][28] Based on high-affinityDARPins,nanobodies,antibody fragments,[29] these methods rigidly bind the target protein and thereby increase the effective particle size and introduce symmetry to improve SNR for Cryo-EM map reconstruction.

2017 Nobel Prize in Chemistry

[edit]

In recognition of the impact cryo-EM has had on biochemistry, three scientists,Jacques Dubochet,Joachim Frank andRichard Henderson, were awarded theNobel Prize in Chemistry "for developing cryo-electron microscopy for the high-resolution structure determination of biomolecules in solution."[4]

Comparisons to X-ray crystallography

[edit]
Main article:X-ray crystallography

Traditionally, X-ray crystallography has been the most popular technique for determining the 3D structures of biological molecules.[30] However, the aforementioned improvements in cryo-EM have increased its popularity as a tool for examining the details of biological molecules. Since 2010, yearly cryo-EM structure deposits have outpaced X-ray crystallography.[31] Though X-ray crystallography has drastically more total deposits due to a decades-longer history, total deposits of the two methods are projected to eclipse around 2035.[31]

The resolution of X-ray crystallography is limited by crystal homogeneity,[32] and coaxing biological molecules with unknown ideal crystallization conditions into a crystalline state can be very time-consuming, in extreme cases taking months or even years.[33] To contrast, sample preparation in cryo-EM may require several rounds of screening and optimization to overcome issues such as protein aggregation and preferred orientations,[34][35] but it does not require the sample to form a crystal, rather samples for cryo-EM are flash-frozen and examined in their near-native states.[36]

According toProteopedia, the median resolution achieved by X-ray crystallography (as of May 19, 2019) on theProtein Data Bank is 2.05Å,[32] and the highest resolution achieved on record (as of September 30, 2022) is 0.48 Å.[37] As of 2020, the majority of the protein structures determined by cryo-EM are at a lower resolution of 3–4 Å.[38] However, as of 2020, the best cryo-EM resolution has been recorded at 1.22 Å,[35] making it a competitor in resolution in some cases.

Biological specimens

[edit]
icon
This sectionneeds additional citations forverification. Please helpimprove this article byadding citations to reliable sources in this section. Unsourced material may be challenged and removed.(October 2025) (Learn how and when to remove this message)

Thin film

[edit]

The biological material is spread on an electron microscopy grid and is preserved in afrozen-hydrated state by rapid freezing, usually in liquidethane nearliquid nitrogen temperature. By maintaining specimens at liquid nitrogen temperature or colder, they can be introduced into the high-vacuum of theelectron microscope column. Most biological specimens are extremelyradiosensitive, so they must be imaged with low-dose techniques (usefully, the low temperature of transmission electron cryomicroscopy provides an additional protective factor against radiation damage).

Consequently, the images are extremelynoisy. For some biological systems it is possible to average images to increase the signal-to-noise ratio and retrieve high-resolution information about the specimen using the technique known assingle particle analysis. This approach in general requires that the things being averaged are identical, although some limited conformational heterogeneity can now be studied (e.g.ribosome). Three-dimensional reconstructions from CryoTEM images of protein complexes andviruses have been solved to sub-nanometer or near-atomic resolution, allowing new insights into the structure and biology of these large assemblies.

Analysis of ordered arrays of protein, such as 2-Dcrystals oftransmembrane proteins orhelical arrays of proteins, also allows a kind of averaging which can provide high-resolution information about the specimen. This technique is calledelectron crystallography.

Vitreous sections

[edit]

The thin film method is limited to thin specimens (typically < 500 nm) because the electrons cannot cross thicker samples without multiple scattering events. Thicker specimens can be vitrified by plunge freezing (cryofixation) in ethane (up to tens of μm in thickness) or more commonly byhigh pressure freezing (up to hundreds of μm). They can then be cut in thin sections (40 to 200 nm thick) with a diamond knife in acryoultramicrotome at temperatures lower than −135 °C (devitrification temperature). The sections are collected on an electron microscope grid and are imaged in the same manner as specimen vitrified in thin film. This technique is called transmission electron cryomicroscopy of vitreous sections (CEMOVIS) or transmission electron cryomicroscopy of frozen-hydrated sections.

Material specimens

[edit]

In addition to allowing vitrified biological samples to be imaged, CryoTEM can also be used to image material specimens that are too volatile in vacuum to image using standard, room temperature electron microscopy. For example, vitrified sections of liquid-solid interfaces can be extracted for analysis by CryoTEM,[39] and sulfur, which is prone to sublimation in the vacuum of electron microscopes, can be stabilized and imaged in CryoTEM.[40]

Image processing in cryo-TEM

[edit]

Even though in the majority of approaches in electron microscopy one tries to get the best resolution image of the material, it is not always the case in cryo-TEM. Besides all the benefits of high resolution images, the signal to noise ratio remains the main hurdle that prevents assigning orientation to each particle. For example, in macromolecule complexes, there are several different structures that are being projected from 3D to 2D during imaging and if they are not distinguished the result of image processing will be a blur. That is why the probabilistic approaches become more powerful in this type of investigation.[41] There are two popular approaches that are widely used nowadays in cryo-EM image processing, the maximum likelihood approach that was discovered in 1998[42] and relatively recently adapted Bayesian approach.[43]

Themaximum likelihood estimation approach comes to this field from the statistics. Here, all the possible orientations of particles are summed up to get the resulting probability distribution. We can compare this to a typicalleast square estimation where particles get exact orientations per image.[44] This way, the particles in the sample get "fuzzy" orientations after calculations, weighted by corresponding probabilities. The whole process is iterative and with each next iteration the model gets better. The good conditions for making the model that closely represent the real structure is when the data does not have too much noise and the particles do not have any preferential direction. The main downside of maximum likelihood approach is that the result depends on the initial guess and model optimization can sometimes get stuck at local minimum.[45]

TheBayesian approach that is now being used in cryo-TEM is empirical by nature. This means that the distribution of particles is based on the original dataset. Similarly, in the usualBayesian method there is a fixedprior probability that is changed after the data is observed. The main difference from the maximum likelihood estimation lies in special reconstruction term that helps smoothing the resulting maps while also decreasing the noise during reconstruction.[44] The smoothing of the maps occurs through assuming prior probability to be a Gaussian distribution and analyzing the data in the Fourier space. Since the connection between the prior knowledge and the dataset is established, there is less chance for human factor errors which potentially increases the objectivity of image reconstruction.[43]

With emerging new methods of cryo-TEM imaging and image reconstruction the new software solutions appear that help to automate the process. After the empirical Bayesian approach have been implemented in the open source computer program RELION (REgularized LIkelihood OptimizatioN) for 3D reconstruction,[46][47] the program became widespread in the cryo-TEM field. It offers a range of corrections that improve the resolution of reconstructed images, allows implementing versatile scripts usingpython language and executes the usual tasks of 2D/3D model classifications or creatingde novo models.[48][49]

Techniques

[edit]

A variety of techniques can be used in CryoTEM.[50] Popular techniques include:

  1. Single particle analysis (SPA)
    1. Time-resolved CryoTEM[51][52][53]
  2. Electron cryotomography (cryoET)
  3. Electron crystallography
    1. Analysis of two-dimensional crystals
    2. Analysis of helical filaments or tubes
    3. Microcrystal Electron Diffraction (MicroED)[54][55][56][57]

Correlative light cryo-TEM and cryo-ET

[edit]
Main article:Correlative light-electron microscopy

In 2019,correlative light cryo-TEM and cryo-ET were used to observetunnelling nanotubes (TNTs) in neuronal cells.[58]

Scanning electron cryomicroscopy

[edit]
Main article:Scanning electron cryomicroscopy

Scanning electron cryomicroscopy (cryoSEM) is ascanning electron microscopy technique with a scanning electron microscope's cold stage in a cryogenic chamber.

Cryogenic transmission electron microscopy

[edit]
Main article:Transmission electron cryomicroscopy

Cryogenic transmission electron microscopy (cryo-TEM) is atransmission electron microscopy technique that is used instructural biology andmaterials science. Colloquially, the term "cryogenic electron microscopy" or its shortening "cryo-EM" refers to cryogenic transmission electron microscopy by default, as the vast majority of cryo-EM is done in transmission electron microscopes, rather than scanning electron microscopes.

Single particle analysis workflow

Advanced methods

[edit]

See also

[edit]
Wikibooks has a book on the topic of:Software Tools For Molecular Microscopy

References

[edit]
  1. ^Tivol WF, Briegel A, Jensen GJ (October 2008)."An improved cryogen for plunge freezing".Microscopy and Microanalysis.14 (5):375–379.Bibcode:2008MiMic..14..375T.doi:10.1017/S1431927608080781.PMC 3058946.PMID 18793481.
  2. ^Cheng Y, Grigorieff N, Penczek PA, Walz T (April 2015)."A primer to single-particle cryo-electron microscopy".Cell.161 (3):438–449.doi:10.1016/j.cell.2015.03.050.PMC 4409659.PMID 25910204.
  3. ^Stoddart C (1 March 2022)."Structural biology: How proteins got their close-up".Knowable Magazine.doi:10.1146/knowable-022822-1.S2CID 247206999. Retrieved25 March 2022.
  4. ^ab"The Nobel Prize in Chemistry 2017".NobelPrize.org. Retrieved2022-09-30.
  5. ^Doerr A (January 2017). "Cryo-electron tomography".Nature Methods.14 (1): 34.doi:10.1038/nmeth.4115.ISSN 1548-7091.S2CID 27162203.
  6. ^Dubochet J, Knapek E (April 2018)."Ups and downs in early electron cryo-microscopy".PLOS Biology.16 (4) e2005550.doi:10.1371/journal.pbio.2005550.PMC 5929567.PMID 29672565.
  7. ^Venables JA (1963). "Liquid helium cooled tilting stage for an electron microscope".Review of Scientific Instruments.34 (5):582–583.
  8. ^Knapek E, Dubochet J (August 1980). "Beam damage to organic material is considerably reduced in cryo-electron microscopy".Journal of Molecular Biology.141 (2):147–161.doi:10.1016/0022-2836(80)90382-4.PMID 7441748.
  9. ^Lepault, J.; Dubochet, J.; Dietrich, I.; Knapek, E.; Zeitler, E. (January 1983). "Amendment to: Electron beam damage to organic specimens at liquid helium temperature".Journal of Molecular Biology.163 (3): 511.doi:10.1016/0022-2836(83)90073-6.
  10. ^abNewmark P (30 September 1982)."Cryo-transmission microscopy Fading hopes".Nature.299 (5882):386–387.Bibcode:1982Natur.299..386N.doi:10.1038/299386c0.
  11. ^abDickerson, JL; Naydenova, K; Peet, MJ; Wilson, H; Nandy, B; McMullan, G; Morrison, R; Russo, CJ (29 April 2025)."Reducing the effects of radiation damage in cryo-EM using liquid helium temperatures".Proceedings of the National Academy of Sciences of the United States of America.122 (17) e2421538122.doi:10.1073/pnas.2421538122.PMC 12054821.PMID 40261934.
  12. ^Pfeil-Gardiner, Olivia; Mills, Deryck J.; Vonck, Janet; Kuehlbrandt, Werner (1 November 2019)."A comparative study of single-particle cryo-EM with liquid-nitrogen and liquid-helium cooling".IUCrJ.6 (6):1099–1105.doi:10.1107/S2052252519011503.PMC 6830223.
  13. ^Dickerson, Joshua L; Russo, Christopher J (25 July 2025). "A Physical Theory For Cryo-EM At Liquid-Helium Temperatures".Microscopy and Microanalysis.31 (Supplement_1).doi:10.1093/mam/ozaf048.514.
  14. ^Dubochet J, McDowall AW (December 1981)."Vitrification of Pure Water for Electron Microscopy".Journal of Microscopy.124 (3):3–4.doi:10.1111/j.1365-2818.1981.tb02483.x.
  15. ^Adrian M, Dubochet J, Lepault J, McDowall AW (March 1984)."Cryo-electron microscopy of viruses".Nature.308 (5954):32–36.Bibcode:1984Natur.308...32A.doi:10.1038/308032a0.PMID 6322001.S2CID 4319199.
  16. ^Garman, Elspeth F.; Owen, Robin Leslie (2006-01-01)."Cryocooling and radiation damage in macromolecular crystallography".Acta Crystallographica Section D Biological Crystallography.62 (1):32–47.doi:10.1107/S0907444905034207.ISSN 0907-4449.
  17. ^Callaway E (September 2015)."The revolution will not be crystallized: a new method sweeps through structural biology".Nature.525 (7568):172–4.Bibcode:2015Natur.525..172C.doi:10.1038/525172a.PMID 26354465.
  18. ^Murata K, Wolf M (Feb 2018)."Cryo-electron microscopy for structural analysis of dynamic biological macromolecules".Biochimica et Biophysica Acta (BBA) - General Subjects.1862 (2):324–334.doi:10.1016/j.bbagen.2017.07.020.PMID 28756276.
  19. ^Khoshouei M, Radjainia M, Baumeister W, Danev R (June 2017)."Cryo-EM structure of haemoglobin at 3.2 Å determined with the Volta phase plate".Nature Communications.8 16099.Bibcode:2017NatCo...816099K.doi:10.1038/ncomms16099.PMC 5497076.PMID 28665412.
  20. ^Merk A, Bartesaghi A, Banerjee S, Falconieri V, Rao P, Davis MI, Pragani R, Boxer MB, Earl LA, Milne JL, Subramaniam S (June 2016)."Breaking Cryo-EM Resolution Barriers to Facilitate Drug Discovery".Cell.165 (7):1698–1707.doi:10.1016/j.cell.2016.05.040.PMC 4931924.PMID 27238019.
  21. ^"PDB Data Distribution by Experimental Method and Molecular Type".www.rcsb.org. Retrieved2019-12-03.
  22. ^"PDB Statistics: Growth of Structures from 3DEM Experiments Released per Year".www.rcsb.org. Retrieved2018-12-22.
  23. ^abKühlbrandt, Werner (2014-03-28)."The Resolution Revolution".Science.343 (6178):1443–1444.Bibcode:2014Sci...343.1443K.doi:10.1126/science.1251652.ISSN 0036-8075.PMID 24675944.S2CID 35524447.
  24. ^Kuster, Daniel J.; Liu, Chengyu; Fang, Zheng; Ponder, Jay W.; Marshall, Garland R. (2015-04-20)."High-Resolution Crystal Structures of Protein Helices Reconciled with Three-Centered Hydrogen Bonds and Multipole Electrostatics".PLOS ONE.10 (4) e0123146.Bibcode:2015PLoSO..1023146K.doi:10.1371/journal.pone.0123146.ISSN 1932-6203.PMC 4403875.PMID 25894612.
  25. ^Henderson, Richard (May 1995). "The potential and limitations of neutrons, electrons and X-rays for atomic resolution microscopy of unstained biological molecules".Quarterly Reviews of Biophysics.28 (2):171–193.doi:10.1017/s003358350000305x.
  26. ^Herzik, Mark A.; Wu, Mengyu; Lander, Gabriel C. (2019-03-04)."High-resolution structure determination of sub-100 kDa complexes using conventional cryo-EM".Nature Communications.10 (1): 1032.Bibcode:2019NatCo..10.1032H.doi:10.1038/s41467-019-08991-8.ISSN 2041-1723.PMC 6399227.PMID 30833564.
  27. ^Wentinck, Koen; Gogou, Christos; Meijer, Dimphna H. (2022)."Putting on molecular weight: Enabling cryo-EM structure determination of sub-100-kDa proteins".Current Research in Structural Biology.4:332–337.doi:10.1016/j.crstbi.2022.09.005.PMC 9562432.
  28. ^Gilman, Morgan S. A.; Skiba, Meredith A. (15 August 2025). "Hidden gems bring proteins into view".Nature Chemical Biology.doi:10.1038/s41589-025-01959-4.
  29. ^
  30. ^Smyth MS, Martin JH (February 2000)."x ray crystallography".Molecular Pathology.53 (1):8–14.doi:10.1136/mp.53.1.8.PMC 1186895.PMID 10884915.
  31. ^abChiu, Wah; Schmid, Michael F.; Pintilie, Grigore D.; Lawson, Catherine L. (January 2021)."Evolution of standardization and dissemination of cryo-EM structures and data jointly by the community, PDB, and EMDB".Journal of Biological Chemistry.296 100560.doi:10.1016/j.jbc.2021.100560.ISSN 0021-9258.PMC 8050867.PMID 33744287.
  32. ^ab"Resolution - Proteopedia, life in 3D".proteopedia.org. Retrieved2020-10-27.
  33. ^Callaway E (February 2020)."Revolutionary cryo-EM is taking over structural biology".Nature.578 (7794): 201.Bibcode:2020Natur.578..201C.doi:10.1038/d41586-020-00341-9.PMID 32047310.
  34. ^Lyumkis, Dmitry (2019-03-29)."Challenges and opportunities in cryo-EM single-particle analysis".Journal of Biological Chemistry.294 (13):5181–5197.doi:10.1074/jbc.rev118.005602.ISSN 0021-9258.PMC 6442032.PMID 30804214.
  35. ^abNakane T, Kotecha A, Sente A, McMullan G, Masiulis S, Brown PM, et al. (November 2020)."Single-particle cryo-EM at atomic resolution".Nature.587 (7832):152–156.Bibcode:2020Natur.587..152N.doi:10.1038/s41586-020-2829-0.PMC 7611073.PMID 33087931.
  36. ^Wang HW, Wang JW (January 2017)."How cryo-electron microscopy and X-ray crystallography complement each other".Protein Science.26 (1):32–39.doi:10.1002/pro.3022.PMC 5192981.PMID 27543495.
  37. ^Schmidt A, Teeter M, Weckert E, Lamzin VS (April 2011)."Crystal structure of small protein crambin at 0.48 Å resolution".Acta Crystallographica. Section F, Structural Biology and Crystallization Communications.67 (Pt 4):424–428.doi:10.1107/S1744309110052607.PMC 3080141.PMID 21505232.
  38. ^Yip KM, Fischer N, Paknia E, Chari A, Stark H (November 2020). "Atomic-resolution protein structure determination by cryo-EM".Nature.587 (7832):157–161.Bibcode:2020Natur.587..157Y.doi:10.1038/s41586-020-2833-4.PMID 33087927.S2CID 224823207.
  39. ^Zachman MJ, Asenath-Smith E, Estroff LA, Kourkoutis LF (December 2016)."Site-Specific Preparation of Intact Solid-Liquid Interfaces by Label-Free In Situ Localization and Cryo-Focused Ion Beam Lift-Out".Microscopy and Microanalysis.22 (6):1338–1349.Bibcode:2016MiMic..22.1338Z.doi:10.1017/S1431927616011892.PMID 27869059.
  40. ^Levin BD, Zachman MJ, Werner JG, Sahore R, Nguyen KX, Han Y, Xie B, Ma L, Archer LA, Giannelis EP, Wiesner U, Kourkoutis LF, Muller DA (February 2017)."Characterization of Sulfur and Nanostructured Sulfur Battery Cathodes in Electron Microscopy Without Sublimation Artifacts".Microscopy and Microanalysis.23 (1):155–162.Bibcode:2017MiMic..23..155L.doi:10.1017/S1431927617000058.PMID 28228169.S2CID 6801783.
  41. ^Cheng, Yifan (2018-08-31)."Single-particle cryo-EM—How did it get here and where will it go".Science.361 (6405):876–880.Bibcode:2018Sci...361..876C.doi:10.1126/science.aat4346.ISSN 0036-8075.PMC 6460916.PMID 30166484.
  42. ^Sigworth, F.J. (1998)."A Maximum-Likelihood Approach to Single-Particle Image Refinement".Journal of Structural Biology.122 (3):328–339.doi:10.1006/jsbi.1998.4014.PMID 9774537.
  43. ^abScheres, Sjors H.W. (January 2012)."A Bayesian View on Cryo-EM Structure Determination".Journal of Molecular Biology.415 (2):406–418.doi:10.1016/j.jmb.2011.11.010.PMC 3314964.PMID 22100448.
  44. ^abNogales, Eva; Scheres, Sjors H.W. (May 2015)."Cryo-EM: A Unique Tool for the Visualization of Macromolecular Complexity".Molecular Cell.58 (4):677–689.doi:10.1016/j.molcel.2015.02.019.ISSN 1097-2765.PMC 4441764.PMID 26000851.
  45. ^Sigworth, Fred J. (2016-02-01)."Principles of cryo-EM single-particle image processing".Microscopy.65 (1):57–67.doi:10.1093/jmicro/dfv370.ISSN 2050-5698.PMC 4749045.PMID 26705325.
  46. ^Scheres, Sjors H. W. (2012-12-01)."RELION: Implementation of a Bayesian approach to cryo-EM structure determination".Journal of Structural Biology.180 (3):519–530.doi:10.1016/j.jsb.2012.09.006.ISSN 1047-8477.PMC 3690530.PMID 23000701.
  47. ^"RELION: Image-processing software for cryo-electron microscopy".GitHub. 3dem. 27 October 2023. Retrieved27 October 2023.
  48. ^Bai, Xiao-chen; McMullan, Greg; Scheres, Sjors H.W (January 2015). "How cryo-EM is revolutionizing structural biology".Trends in Biochemical Sciences.40 (1):49–57.doi:10.1016/j.tibs.2014.10.005.ISSN 0968-0004.PMID 25544475.S2CID 19727349.
  49. ^Zivanov, Jasenko; Nakane, Takanori; Forsberg, Björn O; Kimanius, Dari; Hagen, Wim JH; Lindahl, Erik; Scheres, Sjors HW (2018-11-09). Egelman, Edward H; Kuriyan, John (eds.)."New tools for automated high-resolution cryo-EM structure determination in RELION-3".eLife.7 e42166.doi:10.7554/eLife.42166.ISSN 2050-084X.PMC 6250425.PMID 30412051.
  50. ^Presentation on Cryoelectron Microscopy | PharmaXChange.info
  51. ^Fu Z, Kaledhonkar S, Borg A, Sun M, Chen B, Grassucci RA, Ehrenberg M, Frank J (December 2016)."Key Intermediates in Ribosome Recycling Visualized by Time-Resolved Cryoelectron Microscopy".Structure.24 (12):2092–2101.doi:10.1016/j.str.2016.09.014.PMC 5143168.PMID 27818103.
  52. ^Feng X, Fu Z, Kaledhonkar S, Jia Y, Shah B, Jin A, Liu Z, Sun M, Chen B, Grassucci RA, Ren Y, Jiang H, Frank J, Lin Q (April 2017)."A Fast and Effective Microfluidic Spraying-Plunging Method for High-Resolution Single-Particle Cryo-EM".Structure.25 (4): 663–670.e3.doi:10.1016/j.str.2017.02.005.PMC 5382802.PMID 28286002.
  53. ^Chen B, Kaledhonkar S, Sun M, Shen B, Lu Z, Barnard D, Lu TM, Gonzalez RL, Frank J (June 2015)."Structural dynamics of ribosome subunit association studied by mixing-spraying time-resolved cryogenic electron microscopy".Structure.23 (6):1097–105.doi:10.1016/j.str.2015.04.007.PMC 4456197.PMID 26004440.
  54. ^Shi D, Nannenga BL, Iadanza MG, Gonen T (November 2013)."Three-dimensional electron crystallography of protein microcrystals".eLife.2 e01345.doi:10.7554/eLife.01345.PMC 3831942.PMID 24252878.
  55. ^Nannenga BL, Shi D, Leslie AG, Gonen T (September 2014)."High-resolution structure determination by continuous-rotation data collection in MicroED".Nature Methods.11 (9):927–930.doi:10.1038/nmeth.3043.PMC 4149488.PMID 25086503.
  56. ^Shi D, Nannenga BL, de la Cruz MJ, Liu J, Sawtelle S, Calero G, Reyes FE, Hattne J, Gonen T (May 2016)."The collection of MicroED data for macromolecular crystallography".Nature Protocols.11 (5):895–904.doi:10.1038/nprot.2016.046.PMC 5357465.PMID 27077331.
  57. ^de la Cruz MJ, Hattne J, Shi D, Seidler P, Rodriguez J, Reyes FE, Sawaya MR, Cascio D, Weiss SC, Kim SK, Hinck CS, Hinck AP, Calero G, Eisenberg D, Gonen T (February 2017)."Atomic-resolution structures from fragmented protein crystals with the cryoEM method MicroED".Nature Methods.14 (4):399–402.doi:10.1038/nmeth.4178.PMC 5376236.PMID 28192420.
  58. ^Sartori-Rupp A, Cordero Cervantes D, Pepe A, Gousset K, Delage E, Corroyer-Dulmont S, et al. (January 2019)."Correlative cryo-electron microscopy reveals the structure of TNTs in neuronal cells".Nature Communications.10 (1): 342.Bibcode:2019NatCo..10..342S.doi:10.1038/s41467-018-08178-7.PMC 6341166.PMID 30664666.
  59. ^Bäuerlein, Felix J. B.; Baumeister, Wolfgang (2021-10-01)."Towards Visual Proteomics at High Resolution".Journal of Molecular Biology. From Protein Sequence to Structure at Warp Speed: How Alphafold Impacts Biology.433 (20) 167187.doi:10.1016/j.jmb.2021.167187.ISSN 0022-2836.PMID 34384780.
  60. ^Nannenga BL, Shi D, Leslie AG, Gonen T (September 2014)."High-resolution structure determination by continuous-rotation data collection in MicroED".Nature Methods.11 (9):927–930.doi:10.1038/nmeth.3043.PMC 4149488.PMID 25086503.
  61. ^Jones CG, Martynowycz MW, Hattne J, Fulton TJ, Stoltz BM, Rodriguez JA, et al. (November 2018)."The CryoEM Method MicroED as a Powerful Tool for Small Molecule Structure Determination".ACS Central Science.4 (11):1587–1592.doi:10.1021/acscentsci.8b00760.PMC 6276044.PMID 30555912.
  62. ^de la Cruz MJ, Hattne J, Shi D, Seidler P, Rodriguez J, Reyes FE, et al. (February 2017)."Atomic-resolution structures from fragmented protein crystals with the cryoEM method MicroED".Nature Methods.14 (4):399–402.doi:10.1038/nmeth.4178.PMC 5376236.PMID 28192420.
  63. ^Gruene T, Wennmacher JT, Zaubitzer C, Holstein JJ, Heidler J, Fecteau-Lefebvre A, et al. (December 2018)."Rapid Structure Determination of Microcrystalline Molecular Compounds Using Electron Diffraction".Angewandte Chemie.57 (50):16313–16317.doi:10.1002/anie.201811318.PMC 6468266.PMID 30325568.
  64. ^Cheng Y (August 2018)."Single-particle cryo-EM-How did it get here and where will it go".Science.361 (6405):876–880.Bibcode:2018Sci...361..876C.doi:10.1126/science.aat4346.PMC 6460916.PMID 30166484.
  65. ^Xiao, C., Fischer, M.G., Bolotaulo, D.M., Ulloa-Rondeau, N., Avila, G.A., and Suttle, C.A. (2017) "Cryo-EM reconstruction of the Cafeteria roenbergensis virus capsid suggests novel assembly pathway for giant viruses".Scientific Reports,7: 5484.doi:10.1038/s41598-017-05824-w.
Basics
Electron interaction
with matter
Instrumentation
Microscopes
EM
SEM
TEM
Techniques
Others
Developers
Manufacturers
Software
Retrieved from "https://en.wikipedia.org/w/index.php?title=Cryogenic_electron_microscopy&oldid=1322683670"
Categories:
Hidden categories:

[8]ページ先頭

©2009-2025 Movatter.jp