ISFET

From Wikipedia, the free encyclopedia
Type of field-effect transistor
This article is about the transistor. For the concept in Egyptian mythology, seeIsfet (Egyptian mythology).

Anion-sensitive field-effect transistor (ISFET) is afield-effect transistor used for measuring ion concentrations in solution; when the ion concentration (such asH+, seepH scale) changes, the current through thetransistor will change accordingly. Here, the solution is used as the gate electrode. A voltage between substrate andoxide surfaces arises due to anion sheath. It is a special type ofMOSFET (metal–oxide–semiconductor field-effect transistor),[1] and shares the same basic structure, but with themetal gate replaced by an ion-sensitivemembrane,electrolyte solution andreference electrode.[2] Invented in 1970, the ISFET was the firstbiosensor FET (BioFET).

The schematic view of an ISFET. Source and drain are the two electrodes used in a FET system. The electron flow takes place in a channel between the drain and source. The gate potential controls the flow of current between the two electrodes.

The surfacehydrolysis of Si–OH groups of the gate materials varies in aqueous solutions due to pH value. Typical gate materials areSiO2,Si3N4,Al2O3 andTa2O5.

The mechanism responsible for the oxide surface charge can be described by thesite binding model, which describes the equilibrium between the Si–OH surface sites and the H+ ions in the solution. The hydroxyl groups coating an oxide surface such as that of SiO2 can donate or accept a proton and thus behave in an amphoteric way as illustrated by the following acid-base reactions occurring at the oxide-electrolyte interface:

—Si–OH + H2O     ↔   —Si–O     + H3O+
—Si–OH + H3O+   ↔   —Si–OH2+ + H2O

An ISFET's source and drain are constructed as for aMOSFET. The gate electrode is separated from the channel by a barrier which is sensitive tohydrogen ions and a gap to allow the substance under test to come in contact with the sensitive barrier. An ISFET'sthreshold voltage depends on the pH of the substance in contact with its ion-sensitive barrier.

Practical limitations due to the reference electrode

[edit]

An ISFET electrode sensitive to H+ concentration can be used as a conventionalglass electrode to measure thepH of a solution. However, it also requires areference electrode to operate. If the reference electrode used in contact with the solution is of theAgCl orHg2Cl2 classical type, it will suffer the same limitations as conventional pH electrodes (junction potential,KCl leak, andglycerol leak in case of gel electrode). A conventional reference electrode can also be bulky and fragile. A too large volume constrained by a classical reference electrode also precludes the miniaturization of the ISFET electrode, a mandatory feature for some biological orin vivo clinical analyses (disposable mini-catheter pH probe). The breakdown of a conventional reference electrode could also make problem in on-line measurements in the pharmaceutical or food industry if highly valuable products are contaminated by electrode debris or toxic chemical compounds at a late production stage and must be discarded for the sake of safety.

For this reason, since more than 20 years many research efforts have been dedicated to on-chip embedded tiny reference field effect transistors (REFET). Their functioning principle, or operating mode, can vary, depending on the electrode producers and are often proprietary and protected by patents. Semi-conductor modified surfaces required for REFET are also not always in thermodynamical equilibrium with the test solution and can be sensitive to aggressive or interfering dissolved species or not well characterized aging phenomena. This is not a real problem if the electrode can be frequently re-calibrated at regular time interval and is easily maintained during its service life. However, this may be an issue if the electrode has to remain immersed on-line for prolonged period of time, or is inaccessible for particular constrains related to the nature of the measurements itself (geochemical measurements under elevated water pressure in harsh environments or under anoxic or reducing conditions easily disturbed by atmospheric oxygen ingress or pressure changes).

A crucial factor for ISFET electrodes, as for conventional glass electrodes, remains thus the reference electrode. When troubleshooting electrode malfunctions, often, most of the problems have to be searched for from the side of the reference electrode.

Low-frequency noise of ISFET

[edit]

For ISFET-based sensors, low-frequency noise is most detrimental to the overall SNR as it can interfere with biomedical signals which span in the same frequency domain.[3] The noise has mainly three sources. The noise sources outside the ISFET itself are referred to as the external noise, such as environmental interference and instrument noise from terminal read-out circuits. The intrinsic noise refers to that appearing in the solid part of an ISFET, which is mainly caused by the trapping and de-trapping of carriers at the Oxide/Si interface. And the extrinsic noise is generally rooted in the liquid/oxide interface causing by the ion exchange at the liquid/oxide interface. Many methods are invented to suppress the noise of ISFET. For example, to suppress the external noise, we can integrate a bipolar junction transistor with ISFET to realize immediate the internal amplification of drain current.[4] And to suppress the intrinsic noise we can replace the noisy oxide/Si interface by a Schottky junction gate.[5]

Drift of ISFET

[edit]

Drift of ISFET refers to the threshold voltage instability. When the intrinsic response of the pH-ISFET is completed, the output voltage of the ISFET still vary with time gradually and monotonically, and this drift behavior exists during the entire measurement process. It has been one of the serious obstacles in developing commercially viable, ISFET-based biomedical sensors. In particular, the high accuracy desired for continuous monitoring of blood pH imposes stringent requirements on the tolerable drift rate in pH ISFETs.[6]

Proposed explanations for drift include electric field enhanced ion migration within the gate insulator as well as electrochemical nonequilibrium conditions at the insulator solution interface, injection of electrons from the electrolyte at strong anodic polarizations, creating negative space charge inside the insulator films, and slow surface effects.[7]

History

[edit]

The basis for the ISFET is theMOSFET. Dutch engineerPiet Bergveld, at theUniversity of Twente studied the MOSFET and realized it could be adapted into asensor forelectrochemical andbiological applications.[8][1] This led to Bergveld's invention of the ISFET in 1970.[9][8] He described the ISFET as "a special type of MOSFET with a gate at a certain distance".[1] It was the earliest biosensor FET (BioFET).[10]

ISFET sensors could be implemented inintegrated circuits based onCMOS (complementary MOS) technology. ISFET devices are widely used inbiomedical applications, such as the detection ofDNA hybridization,biomarker detection fromblood,antibody detection,glucose measurement andpH sensing.[2] The ISFET is also the basis for later BioFETs, such as theDNA field-effect transistor (DNAFET),[2][9] used ingenetic technology.[2]

Recently,Graphene field effect transistors (GFETs) operating in liquids have emerged as a promising embodiments of ISFETs. This is because graphene channels offer a more sensitive charge-to-current transducer compared to conventional devices based on classical semiconducting channels. Additionally, the graphene surface is more convenient to functionalize compared to silicon, leading to a wide range of developments in bioFETs.[11]

See also

[edit]

References

[edit]
  1. ^abcBergveld, Piet (October 1985)."The impact of MOSFET-based sensors"(PDF).Sensors and Actuators.8 (2):109–127.Bibcode:1985SeAc....8..109B.doi:10.1016/0250-6874(85)87009-8.ISSN 0250-6874.
  2. ^abcdSchöning, Michael J.; Poghossian, Arshak (10 September 2002)."Recent advances in biologically sensitive field-effect transistors (BioFETs)"(PDF).Analyst.127 (9):1137–1151.Bibcode:2002Ana...127.1137S.doi:10.1039/B204444G.ISSN 1364-5528.PMID 12375833.
  3. ^Bedner, Kristine; Guzenko, Vitaliy A.; Tarasov, Alexey; Wipf, Mathias; Stoop, Ralph L.; Rigante, Sara; Brunner, Jan; Fu, Wangyang; David, Christian; Calame, Michel; Gobrecht, Jens (February 2014)."Investigation of the dominant 1/f noise source in silicon nanowire sensors".Sensors and Actuators B: Chemical.191:270–275.Bibcode:2014SeAcB.191..270B.doi:10.1016/j.snb.2013.09.112.ISSN 0925-4005.
  4. ^Zhang, Da; Gao, Xindong; Chen, Si; Norström, Hans; Smith, Ulf; Solomon, Paul; Zhang, Shi-Li; Zhang, Zhen (2014-08-25)."An ion-gated bipolar amplifier for ion sensing with enhanced signal and improved noise performance".Applied Physics Letters.105 (8): 082102.Bibcode:2014ApPhL.105h2102Z.doi:10.1063/1.4894240.ISSN 0003-6951.
  5. ^Chen, Xi; Chen, Si; Hu, Qitao; Zhang, Shi-Li; Solomon, Paul; Zhang, Zhen (2019-02-22)."Device Noise Reduction for Silicon Nanowire Field-Effect-Transistor Based Sensors by Using a Schottky Junction Gate".ACS Sensors.4 (2):427–433.Bibcode:2019ACSSe...4..427C.doi:10.1021/acssensors.8b01394.ISSN 2379-3694.PMID 30632733.S2CID 58624034.
  6. ^Chou, Jung chuan (2022). "Preparation and study on the drift and hysteresis properties of the tin oxide gate ISFET by the sol–gel method".Sensors and Actuators B: Chemical.86 (1): 58-62.doi:10.1016/S0925-4005(02)00147-8.
  7. ^Jamasb, S.; Collins, S.D.; Smith, R.L. (1997). "A physically-based model for drift in Al2O3-gate pH ISFET's".Proceedings of International Solid State Sensors and Actuators Conference (Transducers '97). Vol. 2. pp. 1379–1382.doi:10.1109/SENSOR.1997.635494.ISBN 0-7803-3829-4.
  8. ^abBergveld, P. (January 1970). "Development of an Ion-Sensitive Solid-State Device for Neurophysiological Measurements".IEEE Transactions on Biomedical Engineering. BME-17 (1):70–71.Bibcode:1970ITBE...17...70B.doi:10.1109/TBME.1970.4502688.PMID 5441220.
  9. ^abChris Toumazou; Pantelis Georgiou (December 2011)."40 years of ISFET technology: From neuronal sensing to DNA sequencing".Electronics Letters.47: S7.doi:10.1049/el.2011.3231. Retrieved13 May 2016.
  10. ^Park, Jeho; Nguyen, Hoang Hiep; Woubit, Abdela; Kim, Moonil (2014)."Applications of Field-Effect Transistor (FET)–Type Biosensors".Applied Science and Convergence Technology.23 (2):61–71.doi:10.5757/ASCT.2014.23.2.61.ISSN 2288-6559.S2CID 55557610.
  11. ^Beraud, A. (2021)."Graphene field-effect transistors as bioanalytical sensors: design, operation and performance".Analyst.146 (2):403–428.Bibcode:2021Ana...146..403B.doi:10.1039/D0AN01661F.PMID 33215184.

Bibliography

[edit]

Further reading

[edit]
Semiconductor
devices
MOS
transistors
Other
transistors
Diodes
Other
devices
Voltage regulators
Vacuum tubes
Vacuum tubes (RF)
Cathode ray tubes
Gas-filled tubes
Adjustable
Passive
Reactive
Other devices
Retrieved from "https://en.wikipedia.org/w/index.php?title=ISFET&oldid=1331241898"
Categories:
Hidden categories:

[8]ページ先頭

©2009-2026 Movatter.jp