Article

Real-Time Depth of Anaesthesia Assessment Based on Hybrid Statistical Features of EEG

1
School of Engineering, University of Southern Queensland, Toowoomba 4350, Australia
2
School of Mathematics, Physics and Computing, University of Southern Queensland, Toowoomba 4350, Australia
*
Author to whom correspondence should be addressed.
Sensors2022,22(16), 6099;https://doi.org/10.3390/s22166099
Submission received: 7 July 2022 /Revised: 8 August 2022 /Accepted: 12 August 2022 /Published: 15 August 2022
(This article belongs to the SectionBiomedical Sensors)

Abstract

:
This paper proposed a new depth of anaesthesia (DoA) index for the real-time assessment of DoA using electroencephalography (EEG). In the proposed new DoA index, a wavelet transform threshold was applied to denoise raw EEG signals, and five features were extracted to construct classification models. Then, the Gaussian process regression model was employed for real-time assessment of anaesthesia states. The proposed real-time DoA index was implemented using a sliding window technique and validated using clinical EEG data recorded with the most popular commercial DoA product Bispectral Index monitor (BIS). The results are evaluated using the correlation coefficients and Bland–Altman methods. The outcomes show that the highest and the average correlation coefficients are 0.840 and 0.814, respectively, in the testing dataset. Meanwhile, the scatter plot of Bland–Altman shows that the agreement between BIS and the proposed index is 94.91%. In contrast, the proposed index is free from the electromyography (EMG) effect and surpasses the BIS performance when the signal quality indicator (SQI) is lower than 15, as the proposed index can display high correlation and reliable assessment results compared with clinic observations.

    1. Introduction

    Monitoring the depth of anaesthesia (DoA) during surgery is challenging [1,2]. When anaesthetic agents are applied, the response of the central nervous system could be reflected in the electroencephalography (EEG) [3]. The EEG signals exhibit low-voltage high-frequency and high-voltage low-frequency characteristics in the awake state and deep anaesthesia state, respectively [4]. Thus, the neural activities in the central nervous system could be revealed by EEG. During surgery, the main problem is prescribing precision doses of appropriate anesthetic agents for each patient. An excessively high dosage of anaesthetic agents may lead to coma and postoperative complications, and on the other hand, if the dosage of anaesthetic agents is insufficient, the patient may suffer from pain and waking up [5]. Thus, the real-time DoA assessment can help both patients and anaesthetists.
    In the past two decades, the Bispectral Index (BIS) has been widely used for assessing and monitoring the DoA. BIS was developed by Aspect Medical Systems in 1992 [6,7], and is one of the most popular commercial products in the DoA monitoring markets. The BIS index was developed from several complex parameters in the time domain and frequency domain, which are integrated as a one dimension index ranging from 0 (unconscious state) to 100 (awake state) [8,9,10]. Besides, there are some other electrophysiological monitors, such as the Narcotrend index [11], M-Entropy [12], and patient state index [13], for tracing the DoA. However, these DoA monitors still have some limitations, such as inconsistent states between clinical observation and device monitoring index, insensitivity at switch points between consciousness and unconsciousness, and not being accurate across patients [14,15]. Moreover, these real-time monitor systems in clinical use have poor performance when the signal quality index (SQI) is low.
    A wide range of features, including frequency domain and time domain features, have been proposed for monitoring DoA over the years. Nguyen-Ky et al. [14] proposed to measure the DoA using a wavelet transform method. In this method, the EEG signal was decomposed into different levels to extract the desired features, and an eigenvector of wavelet coefficients was calculated to develop a new index. Wavelet-weighted median frequency and wavelet coefficient energy entropy methods were used to develop new indexes with high rates of agreement with BIS [16]. Several studies commonly use permutation and sample entropy to create a correlated index with BIS [17,18]. Sarkela et al. [19] used spectral characteristics of the EEG burst suppression as features to propose an automatic method for burst suppression detection and segmentation. Lashkari and Boostani [20] introduced an improved instantaneous frequency (IF) by applying a Kalman filter to develop a noticeable index correlated with the BIS index. In addition, several different types of regression models and classifiers were used to monitor the trending of anaesthetic states, such as an artificial neural network [21,22], an adaptive neuro-fuzzy system [18], a genetic algorithm with a support vector machine (GA-SVM) [23], random forest [24], and convolutional neural network [25]. Neural networks have the advantage of being more flexible in classifying EEG signals. However, these classifiers require a large amount of data to train a robust model, most of the datasets in the research field of anaesthesia are not open to the public, and it is even more challenging to find the public DoA datasets with observation comments and labels from the attending anaesthetists [25].
    This study used the threshold value for the wavelet denoising method to pre-process the raw EEG signal data. After feature extraction and selection, five features were evaluated as the input of the Gaussian process regression model to classify different states of anaesthesia. All of the experiments in this study were carried out using MATLAB 2018b and a workstation with an Inter® Core™ i9-10900K CPU @ 3.7 GHz processor.
    The first section of the paper provided a brief introduction to the work.Section 2 described the details of the datasets and data acquisition. The pre-processing, features extraction, states of anaesthesia classification, and real-time application are also introduced in this section.Section 3 presents the experiments and results obtained. The discussion on the EMG effect and signal quality were evaluated inSection 4.Section 5 concluded the paper.

    2. Materials and Methods

    The proposed threshold wavelet denoising method started with applying a sliding window technique. The sliding window size was 10 s, having a 9 s overlap with the previous window. Then, the statistical features of the denoised EEG signal were extracted. The regression model was trained by the machine learning method to classify the depth of anaesthesia. Finally, the new index,NDoA, was proposed to monitor the DoA in real-time.Figure 1 describes the workflow of the proposed methodology in this study.

    2.1. Subjects and EEG Recording

    The EEG and BIS datasets are directly from our collaborating hospitals, including Toowoomba Hospital, Brisbane Prince Charles Hospital, Queensland, Australia, and our industry partners, such as Shenzhen Delica Medical Equipment Co., Ltd., Shenzhen, China. The EEG and BIS data were obtained using a BISVISTATM monitoring system, version 3.22. All intravenous dosing and intraoperative events were recorded on data log files by the attending anaesthetist. The data log files also contain the BIS index, raw EEG data, signal quality indicator (SQI), impedance, electromyography (EMG), and monitor log of errors. Among the two channels of raw EEG, this study used the one with more consistent EEG quality. In addition, the sampling frequency of the recording data was 128 Hz with a 16-bit signed integer. In this study, a total of 73 subjects with corresponding data were used and the summary of patient demographics is shown inTable 1.

    2.2. Signal Denoising and Pre-Processing

    It is difficult to accurately discriminate the anaesthetic response signal from the raw EEG signal because the noises, such as eye movements, muscle activities, and other artefacts, may corrupt the recorded signals, especially in the awake state. Therefore, all raw EEG signals were pre-processed to reduce the noise. Firstly, the mean and standard-deviation-based threshold methods are used to remove outliers [26]. Then, a wavelet threshold method based on entropy is used to remove the low-amplitude noise and spike noise interference.
    Two threshold methods are used to conduct the denoising process, whereTh is the threshold value. Donoho [27,28] proposed a threshold wavelet method to remove the noise using a threshold value:
    Th=σ2logN
    σ=MAD/0.6745
    whereσ is the noise standard deviation,N is the size of the wavelet coefficient arrays, andMAD is the media of absolute detail coefficient of each level in the wavelet transform.
    The function of soft-thresholdSj has a sign function related to the detail coefficientsCj.
    Sj={sign(Cj)(|Cj|Th),    if |Cj|>Th0,                                       if |Cj|Th 
    The function of hard-thresholdHj has a value equal to 0 if the detail coefficient is smaller than the thresholdT. Otherwise, it is equal toCj.
    Hj={Cj,      if |Cj|>Th0,       if |Cj|Th
    An improved method was proposed using Stein’s unbiased risk estimate [28]. An adaptive wavelet energy entropy method was used to filter noise in [29]. In this study, the threshedThnew used a permutation entropy method based on Stein’s unbiased risk estimate [28].
    Hpe(m)=j=1m!pjlnpj
    wherem is the dimension of time series signals andpj is the possibility of each permutation.
    The total energy of permutation entropy is
    PEtotal=k=1Kj=1m!pjlnpj=k=1KHpe(m)
    The relative permutation entropy energy is
    rpe=Hpe(m)PEtotal
    The new adaptive thresholdThnew was proposed as
    Thnew=(logrpelog(nlog(n))2a)×b
    wheren is the window length of an EEG segment anda andb are two constants (9 and 6, respectively) that are calculated offline empirically. In this study, a sliding window technique with a step of one second (s) and a fixed length (L = 10 s) was used to read and process the raw EEG signals. Thus, there are 9 s signals that are overlapped between two sequential sliding windows.

    2.3. Features’ Extraction

    Five features were selected based on the pre-processed data to evaluate their differences in different anaesthetic states. The features are the sample entropy (SE), fuzzy entropy (FE), permutation entropy (PE), Hurst exponent (HE), and power spectral density (PSD), and their calculation methods are provided below.
    Entropy is the parameter to measure the degree of irregularity of a time series signal. In this study, SE, FE, and PE are calculated to produce a new depth of anaesthesia index [17,30].
    SE=lnAm(r)Bm(r)
    whereAm is the number of template pairs havingd[Xm+1(i),Xm+1(j)]<r andBm is the number of template pairs havingd[Xm(i),Xm(j)]<r. Tolerancer was chosen as 0.2 ×std (std is the standard deviation of the time series signals), and the dimension m was chosen as
    FE(X,m,n,r,N)=lnm(n,r)lnm+1(n,r)
    whereN is the length of the time series. Empirically, tolerancer was chosen as 0.15 ×std,n was selected as 2, and dimensionm was chosen as 2 to produce the best performance.
    PE=j=1JPjlnPjln(J)
    where embedding dimensionm and time delayτ are 4 and 1, respectively.
    The Hurst exponent, a numerical method, was used to predict the trend in EEG signals. The time series signalsY with the length ofN were divided into componentsym={y1,n,y2,n,,ym,n,} and each component has the same lengthmϵ{N,N2,N4,}. The following procedure was used to calculate the range response in the Hurst method.
    • Calculating the mean value of each component:
    mean=1mi=1myi,n
    2.
    Creating a mean-adjusted series:
    Xi,n=yi,nmean,      for i=1,2,,m
    3.
    Calculating the cumulative deviate series:
    Zi,n=t=1iXi,n
    4.
    Computing the rangesR:
    Rn=max{Z1,n,Z2,n,,Zm,n}min{Z1,n,Z2,n,,Zm,n}
    5.
    Computing the standard deviationSn:
    Sn=1mi=1m(yi,nmean)2
    6.
    Calculating the rescaled range:
    (R/S)n=1(Nm)n=1[Nm](Rn/Sn)
    7.
    Calculating the rescaled range:
    Rr=minn=1,2,{mean(Rn)}
    Multiple signal classification (MUSIC), as the method of an eigenvector-based frequency estimator, is used. The feature extraction was from the power spectral density (PSD) of wavelet coefficients, and the method is described below based on [14,31]:
    1.
    Aj andDj were used as input signals for the eigenvector method:
    SD:{DjP(Dj)=EDjAjP(Aj)=EAj
    wherej was set as six-level wavelet decomposition, the order of 16 Daubechies wavelet filter was used [32], and the principal eigenvector was 6.
    2.
    Means ofEAj andEDj are as follows:
    Mean:{DjM(EDj)=mean (EDj)AjM(ADj)=mean (ADj)
    3.
    Standard deviation (STD) ofEAj andEDj are as follows:
    STD:{DjS(EDj)=std(EDj)AjS(ADj)=std (ADj)
    4.
    Derived from steps 2 and 3:
    D:{  Mj12{log(mean(M(EDj)))+log(mean(M(ADj)))}Sj12{log(mean(S(EDj)))+log(mean(S(ADj)))}
    5.
    Deriving the feature:
    Fea= k1×Mj+k2× Sjk3
    The constant parametersk1,k2, andk3 were chosen as 28, 90, and 3, respectively.
    These five features were extracted individually after pre-processing the raw EEG signals. Then, they were used as the input data to train a robust model to classify different states of anaesthesia.

    2.4. Regression Models and Evaluation Measures

    In this study, a Gaussian process regression model, as a nonparametric Bayesian approach, was applied. This kind of regression model has its advantage on small datasets and provides reliable uncertain predictions. We also utilized linear regression, SVM, and regression tree models for comparison and evaluation. The dataset with 73 subjects was divided into the training and testing sets. A total of 60 subjects were randomly selected for training, and the aforementioned five features of each subject were calculated for training and evaluating regression models to find the best performance models based on different combinations of features.
    The training models were evaluated using R-square (R2), root-mean-square-error (RMSE), mean-square-error (MSE), and mean absolute error (MAE).
    R2=1i(yifi)2i(yiy¯)2
    whereyi is the data set value (BIS value),y¯ is the mean of the BIS values, andfi is the calculated value from regression results. The range of theR2 values are between 0 and 1. The higher value ofR2 indicates a higher correlation between the BIS value and the extracted feature values, and vice versa.
    MSE=1ni=1n(YiYi¯)2
    MAE= i=1n|YiYi¯|n
    wheren is the number of the data points,Yi is a set of the observed values, andYi¯ is a set of the regression results.
    After the best performance model was selected and evaluated using the testing datasets, the Bland–Altman method was used to assess the degree of agreement between the proposed index,NDoA, and the BIS index. In addition, the Pearson correlation coefficient was also applied to assess the correlation ofNDoA with the BIS index and clinical observations.
    r= (xix¯)(yiy¯) (xix¯)2(yiy¯)2

    2.5. Real-Time DoA Monitor under Sliding Window Framework

    In order to have a real-time response, the sliding window technique was applied and the window size was chosen as 10 s, and the overlap between two adjacent windows was 9 s. In addition, the tuned index,DoAtuned, was calculated using Equation (27), from the mean value between previous and present data.
    DoAtuned=0.8×Indexpre+0.2×Indexcur
    whereIndexpre is the mean value of the DoA index from the last four seconds,Indexcur is the current DoA index, andDoAtuned presents the real-time index second by second. Using this sliding window method, the DoA index could be updated every second. Consequently, the time delay for monitoring the DoA could be reduced. Meanwhile, the starting time of processing the recording raw EEG data was from the fifth second, and there was a four-second delay.

    3. Results

    3.1. Raw EEG Signal Pre-Processing and Features’ Selection

    Figure 2 presents the sample denoising results with the proposed thresholdThnew using the permutation entropy method. The raw EEG signal (from patient ID: L112161431) has low amplitude noise (between 2000 s and 2010 s) and spike noise (between 2980 s and 2990 s), as shown inFigure 2a–c, respectively. Compared with the denoised EEG signal using the thresholdTh, as shown inFigure 2d,e, the denoised EEG signal using the proposed thresholdThnew, as shown inFigure 2f,g, has much less noise. After using both denoising threshold methods, the EEG signal has a similar amplitude. However, the proposed threshold method in this study removed almost all the noise, as shown inFigure 2d–g.
    There are EEG signals from two channels (CH1 and CH2) in the datasets. In this study, CH2 data were selected because their correlation between the features and anaesthetic states was higher than that of CH1. The five features of fuzzy entropy, permutation entropy, sample entropy, power spectral, and Hurst range response were extracted from the denoised CH2 EEG signal. All of the features were measured using coefficients of determination (R2), which reflects how close a feature is to different anaesthetic states. The range ofR2 values is between 0 and 1. A higher value ofR2 indicates a higher correlation between the BIS and the extracted feature, and vice versa. As an example of the patient ID: L01040838, the values ofR2 are 0.7497, 0.6974, 0.7034, and 0.6214 for fuzzy entropy, permutation entropy, sample entropy, and power spectral, respectively. The feature of Hurst range response has a relatively lower correlation with the BIS compared with the other four features. However, the Hurst range response has an obvious state change point from consciousness to unconsciousness at 762 s, as shown inFigure 3.

    3.2. Regression Model Training

    A total of 60 subjects were randomly selected as the training set from 73 subjects. The total EEG recording time for the 60 subjects was 385,116 s, with the corresponding BIS values for each second. The remaining 13 subjects were used to evaluate the training regression models. Fivefold cross-validation, as a resampling procedure, was used to protect against overfitting by partitioning the data sets.
    The MATLAB regression learner app [32] was used for training with four different models, including linear regression, trees, SVM, and Gaussian process models.Table 2 lists the best performance for each type of training model. For example, fine Gaussian SVM has the greatest value ofR2 and smallest value of RMSE among linear SVM, quadratic SVM, cubic SVM, medium SVM, and coarse SVM. The results inTable 2 show that the squared exponential Gaussian process regression model (SEGP-RM) has a better overall performance. Thus, the SEGP-RM was selected for training the robust model.Figure 4 andFigure 5 show the fivefold cross validation results of the SEGP-RM. As a result,R2 is 0.76, while RMSE, MSE, and MAE are 6.18, 38.22, and 5.15, respectively. The response plot is used to explore the correlation between the predicted values and true values, and the correlation coefficient is 0.89. The predicted plot versus actual plot inFigure 4 shows how well the regression model makes predictions for different response values. A perfect regression model has a predicted response equal to the true response, so all of the points (blue points in the below figure) lie on a diagonal line (red line in the below figure). The vertical distance from the line to any point is the prediction error for that point.

    3.3. Performance of the Training Models and Results of the Nex Index

    To further validate the performance of the trained models, the remaining 13 subjects were used for evaluation. The metric, Pearson correlation coefficient, which is defined in Equation (26), is used to evaluate the four trained regression models listed inTable 2.
    Figure 6 indicates the Pearson correlation coefficients of the 13 testing subjects. Patient number 14 represents the average value of all testing data results. The correlation coefficient of the SEGP-RM is higher than the other three regression models, robust linear, fine trees, and fine Gaussian SVM. The highest correlation coefficient is 0.940 for patient No. 8 (subject ID: L01131248), and the lowest one is 0.580 for patient No. 3 (subject ID: L01060841). Meanwhile, the average value of correlation coefficients from these 13 testing subjects is 0.812.
    The difference betweenNDoA and BIS is defined asd=NDoABIS. The mean value of the difference betweenNDoA and BIS isdiff=mean(d) and the standard deviation of the difference issd=std(d). The Bland–Altman method suggests that 95% of thed should lie betweendiff+2sd anddiff2sd (two black lines inFigure 7a) if the datad have a normal distribution. InFigure 7b, the pattern matches well with the distribution ofd. The Bland–Altman plot of the data from 13 testing subjects, calculates the mean differencediff=2.1348 as the ‘bias’ (red dash line). For the 95% agreement limitation, the upper limitation isdiff+2sd=21.46 and the lower one isdiff2sd=17.19. The agreement rate betweenNDoA and BIS is 94.91%, as shown in the scatter plot inFigure 7a.

    4. Results and Discussion

    4.1. Length of Sliding Window

    During surgery and clinical operations, the whole anaesthetic process is controlled by anaesthetists. The surgery’s incisions started when the patient’s anaesthetic states were considered at the middle anaesthesia stage (BIS index between 40 and 60). The real-time technique proposed in this paper only has four seconds’ delay at the beginning of recording data (awake stage), which may not affect clinical data processing and anaesthetic state tracing in clinical applications. The sliding window size affects the real-time results of the anaesthesia monitoring and efficiency. If the sliding window size is too big, the results may cause a delay in the clinical operations, and too small of a window size may affect the feature extractions during the transitions of two anaesthesia states. In this study, a fixed-length sliding window of 10 s was selected.

    4.2. The Effect of Signal Quality

    The signal quality indicator is an index recorded with BIS values in real time, which measures the quality of the raw EEG signals. Artifacts, such as eye movement, muscle activity, head motion, and the electrical knife operations, affect the impedance calculation and result in poor signal quality. We found that the BIS values were not displayed on the monitor screen when SQI was lower than 15. Meanwhile, we reviewed the raw data log files, and the corresponding BIS values are −3276.8 when the signal quality is poor (SQI lower than 15). However, our proposed real-time monitor index,NDoA, can still provide the corresponding anaesthetic states in the case of poor signal quality.
    Figure 8 demonstrates a sample case with poor signal quality (subject ID: L01121339). The BIS values are −3276.8 during the time from 369 to 371 s, from 1708 to 1710 s, and from 1790 to 2017 s. However, the proposed index,NDoA, computed and provided the DoA values clearly during the period.Figure 9 presents another case with poor signal quality (subject ID: L01210841). The subject was a 17-year-old male patient weighing 46 kg. The intraoperative EEG monitoring in surgery started at 08:41:41 and finished at 13:37:27. The detail of anaesthesia medications for this patient is reported inTable 3. In this case, the SQI index was below 55 in the awake stage (from 1 to 610 s). Furthermore, the invalid BIS index (−3276.8) could not display on the screen monitor as poor signal quality (SQI < 15) during these periods from 89 s to 132 s and 389 s to 412 s, while the proposedNDoA was able to estimate the DoA. Compared with the BIS index, the proposed method has a better correlation and agreement with clinical observations during periods of poor signal quality. Moreover, two additional cases with the same phenomenon could be observed inFigure 10 andFigure 11.

    4.3. The Effect of EMG Signals

    EMG signals, caused by muscle activities, were recorded as a spark from forehead muscles areas. These signals can have a negative effect on increasing the value of the BIS index, and the neuromuscular blocking agents can reduce or eliminate this side effect [33]. InFigure 12 there are two spikes with the BIS index higher than 70 in a general anaesthesia state from 2560 s to 2580 s and from 2870 s to 2890 s. The index of EMG signals also reflected this fact. Meanwhile, the SQIs are between 90 and 100 during these two periods, which indicates reliable and accurate signal quality with the BIS index. Therefore, the EMG activities during these periods might result in two spikes for the BIS indexing. However, as observed from the waveform of theNDoA, the trend of the proposed index does not change in the general anaesthesia state.

    5. Conclusions

    This paper studies the real-time monitoring of anaesthesia states using hybrid statistical features and machine learning methods. Firstly, a wavelet threshold method was used to remove the low-amplitude and spike noise interference. Then, hybrid statistical features were extracted from the denoised EEG signals. Next, the regression models were trained using these extracted features by applying the machine learning methods, linear regression, SVM, regression trees, and Gaussian process regression. The squared exponential Gaussian process regression model outperformed the other models. This yields a better result in accuracy and a higher degree of agreement rates. In addition, the proposed index,NDoA, has a higher correlation with anaesthetists’ observations than the BIS index during periods of poor signal quality.
    The results were evaluated by comparing the correlation and agreement with the BIS index (current gold standard) and the clinical observations recorded by the attending anaesthetists. The results ofNDoA are highly correlated with the BIS index. The highest correlation coefficient,R2, is 0.940, and the average is 0.812. Furthermore, the agreement measured by the Bland–Altman method presented a higher degree of agreement between the proposedNDoA and the BIS index. The scatter plot shows an agreement of 94.91%.

    Author Contributions

    Y.H. presented the project idea and completed the modelling, experiments, and the writing of this manuscript, while P.W., B.S., and Y.L., being supervisors, contributed to the design of the study, the completion of the project, and the editing of this manuscript. All authors read and approved the final. All authors have read and agreed to the published version of the manuscript.

    Funding

    This research received no external funding.

    Institutional Review Board Statement

    The study was conducted in accordance with the Declaration of Helsinki, and approved by the Institutional Ethics Committee of University of Southern Queensland (No: H09REA029, 2009/016).

    Informed Consent Statement

    The medical condition and procedure of data collection were explained to all subjects/patients. Consent, including all ethical issues, was obtained from all patients. The study was approved by the University of Southern Queensland Human Research Ethics Committee (No: H09REA029) and the Toowoomba and Darling Downs Health Service District Human Research Ethics Committee (No: TDDHSD HREC 2009/016).

    Data Availability Statement

    The data and materials used in this study are available at the University of Southern Queensland under the research data management policy.

    Conflicts of Interest

    The authors declare no conflict of interest.

    References

    1. Bard, J.W. The BIS monitor: A review and technology assessment.AANA J.2001,69, 477–483. [Google Scholar]
    2. Constant, I.; Sabourdin, N. Monitoring depth of anesthesia: From consciousness to nociception. A window on subcortical brain activity.Pediatr. Anesthesia2014,25, 73–82. [Google Scholar] [CrossRef] [PubMed]
    3. Jospin, M.; Caminal, P.; Jensen, E.W.; Litvan, H.; Vallverdu, M.; Struys, M.M.R.F.; Vereecke, H.E.M.; Kaplan, D.T. Detrended Fluctuation Analysis of EEG as a Measure of Depth of Anesthesia.IEEE Trans. Biomed. Eng.2007,54, 840–846. [Google Scholar] [CrossRef] [PubMed]
    4. Musizza, B.; Ribarič, S. Monitoring the Depth of Anaesthesia.Sensors2010,10, 10896–10935. [Google Scholar] [CrossRef]
    5. Pandit, J.J.; Cook, T.M.; Jonker, W.R.; O’Sullivan, E. A national survey of anaesthetists (NAP5 Baseline) to estimate an annual incidence of accidental awareness during general anaesthesia in the UK.Anaesthesia2013,68, 343–353. [Google Scholar] [CrossRef] [PubMed]
    6. Moca, V.V.; Scheller, B.; Mureşan, R.C.; Daunderer, M.; Pipa, G. EEG under anesthesia—Feature extraction with TESPAR.Comput. Methods Programs Biomed.2009,95, 191–202. [Google Scholar] [CrossRef]
    7. Diykh, M.; Miften, F.S.; Abdulla, S.; Saleh, K.; Green, J.H. Robust approach to depth of anaesthesia assessment based on hybrid transform and statistical features.IET Sci. Meas. Technol.2020,14, 128–136. [Google Scholar] [CrossRef]
    8. Ahmadi, B.; Negahbani, E.; Amirfattahi, R.; Zaghari, B.; Mansouri, M. Extraction of BIS index sub-parameters in different anesthetic and sedative levels. In Proceedings of the 2008 9th International Conference on Signal Processing, Beijing, China, 26–29 October 2008; pp. 2665–2668. [Google Scholar] [CrossRef]
    9. Nguyen-Ky, T.; Wen, P.; Li, Y. An Improved Detrended Moving-Average Method for Monitoring the Depth of Anesthesia.IEEE Trans. Biomed. Eng.2010,57, 2369–2378. [Google Scholar] [CrossRef]
    10. Voss, L.; Sleigh, J. Monitoring consciousness: The current status of EEG-based depth of anaesthesia monitors.Best Pract. Res. Clin. Anaesthesiol.2007,21, 313–325. [Google Scholar] [CrossRef]
    11. Kreuer, S.; Bruhn, J.; Larsen, R.; Bialas, P.; Wilhelm, W. Comparability of Narcotrend™ index and bispectral index during propofol anaesthesia.Br. J. Anaesth.2004,93, 235–240. [Google Scholar] [CrossRef]
    12. Bruhn, J.; Myles, P.S.; Sneyd, R.; Struys, M.M.R.F. Depth of anaesthesia monitoring: What’s available, what’s validated and what’s next?Br. J. Anaesth.2006,97, 85–94. [Google Scholar] [CrossRef] [PubMed]
    13. Drover, D.; Ortega, H.R. Patient state index.Best Pract. Res. Clin. Anaesthesiol.2006,20, 121–128. [Google Scholar] [CrossRef]
    14. Nguyen-Ky, T.T.; Wen, P.; Li, Y.; Gray, R. Measuring and Reflecting Depth of Anesthesia Using Wavelet and Power Spectral Density.IEEE Trans. Inf. Technol. Biomed.2011,15, 630–639. [Google Scholar] [CrossRef]
    15. Diykh, M.; Li, Y.; Wen, P.; Li, T. Complex networks approach for depth of anesthesia assessment.Measurement2018,119, 178–189. [Google Scholar] [CrossRef]
    16. Afrasiabi, S.; Boostani, R.; Koochaki, S.; Zand, F. Presenting an effective EEG-based index to monitor the depth of anesthesia. In Proceedings of the 16th CSI International Symposium on Artificial Intelligence and Signal Processing (AISP 2012), Shiraz, Iran, 2–3 May 2012; pp. 557–562. [Google Scholar] [CrossRef]
    17. Liu, Q.; Ma, L.; Fan, S.-Z.; Abbod, M.F.; Shieh, J.-S. Sample entropy analysis for the estimating depth of anaesthesia through human EEG signal at different levels of unconsciousness during surgeries.PeerJ2018,6, e4817. [Google Scholar] [CrossRef] [PubMed]
    18. Shalbaf, A.; Saffar, M.; Sleigh, J.W.; Shalbaf, R. Monitoring the Depth of Anesthesia Using a New Adaptive Neurofuzzy System.IEEE J. Biomed. Heal. Inform.2017,22, 671–677. [Google Scholar] [CrossRef]
    19. Särkelä, M.; Mustola, S.; Seppänen, T.; Koskinen, M.; Lepola, P.; Suominen, K.; Juvonen, T.; Tolvanen-Laakso, H.; Jäntti, V. Automatic Analysis and Monitoring of Burst Suppression in Anesthesia.Int. J. Clin. Monit. Comput.2002,17, 125–134. [Google Scholar] [CrossRef]
    20. Lashkari, A.; Boostani, R. A Kalman-based instantaneous frequency estimation for anesthetic depth measurement. In Proceedings of the 2017 22nd International Conference on Digital Signal Processing (DSP), London, UK, 23–25 August 2017; pp. 1–4. [Google Scholar] [CrossRef]
    21. Gu, Y.; Liang, Z.; Hagihira, S. Use of Multiple EEG Features and Artificial Neural Network to Monitor the Depth of Anesthesia.Sensors2019,19, 2499. [Google Scholar] [CrossRef]
    22. Shalbaf, R.; Behnam, H.; Sleigh, J.W.; Steyn-Ross, A.; Voss, L.J. Monitoring the depth of anesthesia using entropy features and an artificial neural network.J. Neurosci. Methods2013,218, 17–24. [Google Scholar] [CrossRef]
    23. Liang, Z.; Huang, C.; Li, Y.; Hight, D.F.; Voss, L.J.; Sleigh, J.W.; Li, X.; Bai, Y. Emergence EEG pattern classification in sevoflurane anesthesia.Physiol. Meas.2018,39, 045006. [Google Scholar] [CrossRef]
    24. Peker, M.; Arslan, A.; Sen, B.; Celebi, F.V.; But, A. A novel hybrid method for determining the depth of anesthesia level: Combining ReliefF feature selection and random forest algorithm (ReliefF+RF). In Proceedings of the 2015 International Symposium on Innovations in Intelligent SysTems and Applications (INISTA), Madrid, Spain, 2–4 September 2015; pp. 1–8. [Google Scholar] [CrossRef]
    25. Liu, Q.; Cai, J.; Fan, S.-Z.; Abbod, M.F.; Shieh, J.-S.; Kung, Y.; Lin, L. Spectrum Analysis of EEG Signals Using CNN to Model Patient’s Consciousness Level Based on Anesthesiologists’ Experience.IEEE Access2019,7, 53731–53742. [Google Scholar] [CrossRef]
    26. Li, X.; Li, D.; Liang, Z.; Voss, L.J.; Sleigh, J.W. Analysis of depth of anesthesia with Hilbert–Huang spectral entropy.Clin. Neurophysiol.2008,119, 2465–2475. [Google Scholar] [CrossRef] [PubMed]
    27. Donoho, D.L. De-noising by soft-thresholding.IEEE Trans. Inf. Theory1995,41, 613–627. [Google Scholar] [CrossRef]
    28. Donoho, D.L. Nonlinear wavelet methods for recovery of signals, densities, and spectra from indirect and noisy data.Proc. Symp. Appl. Math.1993,47, 173–205. [Google Scholar] [CrossRef]
    29. Sang, Y.-F.; Wang, D.; Wu, J.-C. Entropy-Based Method of Choosing the Decomposition Level in Wavelet Threshold De-noising.Entropy2010,12, 1499–1513. [Google Scholar] [CrossRef]
    30. Liang, Z.; Wang, Y.; Sun, X.; Li, D.; Voss, L.J.; Sleigh, J.W.; Hagihira, S.; Li, X. EEG entropy measures in anesthesia.Front. Comput. Neurosci.2015,9, 16. [Google Scholar] [CrossRef]
    31. Eriksson, A.; Stoica, P.; Soderstrom, T. Markov-based eigenanalysis method for frequency estimation.IEEE Trans. Signal Process.1994,42, 586–594. [Google Scholar] [CrossRef]
    32. MATLAB.9.9.0.1524771 (R2020b); The MathWorks Inc.: Natick, MA, USA, 2020. Available online:https://www.mathworks.com/ (accessed on 15 May 2022).
    33. Whitham, E.M.; Lewis, T.; Pope, K.J.; Fitzgibbon, S.P.; Clark, C.R.; Loveless, S.; DeLosAngeles, D.; Wallace, A.K.; Broberg, M.; Willoughby, J.O. Thinking activates EMG in scalp electrical recordings.Clin. Neurophysiol.2008,119, 1166–1175. [Google Scholar] [CrossRef]
    Sensors 22 06099 g001 550
    Figure 1. The diagram of the new DoA index development.
    Figure 1. The diagram of the new DoA index development.
    Sensors 22 06099 g001
    Sensors 22 06099 g002 550
    Figure 2. Results of denoising raw EEG data, including the low amplitude (time is between the 2000 s and 2010 s) and spike noise (time is between 2980 s and 2990 s). (a) Raw EEG data, patient ID: L12161431. (b) Raw EEG data with low amplitude noise. (c) Raw EEG data having spike noise. (d) EEG signal after denoising low amplitude with thresholdTh. (e) EEG signal after denoising spike with thresholdTh. (f) EEG signal after denoising low amplitude with proposed thresholdThnew. (g) EEG signal after denoising spike with proposed thresholdThnew.
    Figure 2. Results of denoising raw EEG data, including the low amplitude (time is between the 2000 s and 2010 s) and spike noise (time is between 2980 s and 2990 s). (a) Raw EEG data, patient ID: L12161431. (b) Raw EEG data with low amplitude noise. (c) Raw EEG data having spike noise. (d) EEG signal after denoising low amplitude with thresholdTh. (e) EEG signal after denoising spike with thresholdTh. (f) EEG signal after denoising low amplitude with proposed thresholdThnew. (g) EEG signal after denoising spike with proposed thresholdThnew.
    Sensors 22 06099 g002
    Sensors 22 06099 g003 550
    Figure 3. Relationship between the Hurst range response value and the BIS value.
    Figure 3. Relationship between the Hurst range response value and the BIS value.
    Sensors 22 06099 g003
    Sensors 22 06099 g004 550
    Figure 4. Predicted response from the SEGP-RM vs. actual values (BIS).
    Figure 4. Predicted response from the SEGP-RM vs. actual values (BIS).
    Sensors 22 06099 g004
    Sensors 22 06099 g005 550
    Figure 5. Response plot of the SEGR-RM for the 60 training subjects.
    Figure 5. Response plot of the SEGR-RM for the 60 training subjects.
    Sensors 22 06099 g005
    Sensors 22 06099 g006 550
    Figure 6. Pearson correlation coefficients of the testing results and BIS values for 13 subjects (patient 14 represents the average results of the 13 testing subjects).
    Figure 6. Pearson correlation coefficients of the testing results and BIS values for 13 subjects (patient 14 represents the average results of the 13 testing subjects).
    Sensors 22 06099 g006
    Sensors 22 06099 g007 550
    Figure 7. (a) Bland-Altman plot betweenNDoA index and BIS. (b) Distribution plot between new DoA index and BIS.
    Figure 7. (a) Bland-Altman plot betweenNDoA index and BIS. (b) Distribution plot between new DoA index and BIS.
    Sensors 22 06099 g007
    Sensors 22 06099 g008 550
    Figure 8. The proposed index,NDoA, can monitor the anaesthetic states change, but BIS cannot when SQI is lower than 15 (subject ID: L01121339).
    Figure 8. The proposed index,NDoA, can monitor the anaesthetic states change, but BIS cannot when SQI is lower than 15 (subject ID: L01121339).
    Sensors 22 06099 g008
    Sensors 22 06099 g009 550
    Figure 9. (a) Comparison betweenNDoA and BIS in the case of poor signal quality (subject ID: L01210841). (b) Zoom out during the period between 0 and 650 s. (c) Zoom out during the period between 16,940 and 17,748 s.
    Figure 9. (a) Comparison betweenNDoA and BIS in the case of poor signal quality (subject ID: L01210841). (b) Zoom out during the period between 0 and 650 s. (c) Zoom out during the period between 16,940 and 17,748 s.
    Sensors 22 06099 g009
    Sensors 22 06099 g010 550
    Figure 10. Comparison betweenNDoA and BIS in the case of poor signal quality (subject ID: L01131131).
    Figure 10. Comparison betweenNDoA and BIS in the case of poor signal quality (subject ID: L01131131).
    Sensors 22 06099 g010
    Sensors 22 06099 g011 550
    Figure 11. Comparison betweenNDoA and BIS in the case of poor signal quality (subject ID: L01131400).
    Figure 11. Comparison betweenNDoA and BIS in the case of poor signal quality (subject ID: L01131400).
    Sensors 22 06099 g011
    Sensors 22 06099 g012 550
    Figure 12. (a) EMG causes the BIS index to increase at two spikes, and the trend ofNDoA does not change during these periods. Subject ID: L01041002. (b) Zoom out during the period between 2500 and 3000 s.
    Figure 12. (a) EMG causes the BIS index to increase at two spikes, and the trend ofNDoA does not change during these periods. Subject ID: L01041002. (b) Zoom out during the period between 2500 and 3000 s.
    Sensors 22 06099 g012
    Table
    Table 1. Summary of patient demographics.
    Table 1. Summary of patient demographics.
    Age (year)Weight (kg)Height (cm)Gender (M/F)
    2–8311–10190–19439/34
    Table
    Table 2. Training model regression of the anaesthetic states.
    Table 2. Training model regression of the anaesthetic states.
    Training ModelR2RMSEMSEMAE
    Robust Linear0.4311.21125.588.81
    Fine Tress0.69.4188.536.25
    Fine Gaussian SVM0.736.6143.675.12
    Squared Exponential Gaussian Process Regression0.795.6632.044.84
    Table
    Table 3. Details of anaesthesia medication.
    Table 3. Details of anaesthesia medication.
    SubjectEventTimestampTime (s)
    L01210841E1. Add Dexmedetomidine Hydrochloride (30 ug)08:25:00-
    E2. Intraoperative EEG monitoring08:41:410
    E3. Add Dezocine (6 mg)08:55:00799
    E4. Add Tropisetron Hydrochloride (2 mg)09:05:001399
    E5. Add Ketorolac (30 mg)09:16:002059
    E6. Surgery End13:37:2717,746
    Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

    © 2022 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).

    Share and Cite

    MDPI and ACS Style

    Huang, Y.; Wen, P.; Song, B.; Li, Y. Real-Time Depth of Anaesthesia Assessment Based on Hybrid Statistical Features of EEG.Sensors2022,22, 6099. https://doi.org/10.3390/s22166099

    AMA Style

    Huang Y, Wen P, Song B, Li Y. Real-Time Depth of Anaesthesia Assessment Based on Hybrid Statistical Features of EEG.Sensors. 2022; 22(16):6099. https://doi.org/10.3390/s22166099

    Chicago/Turabian Style

    Huang, Yi, Peng Wen, Bo Song, and Yan Li. 2022. "Real-Time Depth of Anaesthesia Assessment Based on Hybrid Statistical Features of EEG"Sensors 22, no. 16: 6099. https://doi.org/10.3390/s22166099

    APA Style

    Huang, Y., Wen, P., Song, B., & Li, Y. (2022). Real-Time Depth of Anaesthesia Assessment Based on Hybrid Statistical Features of EEG.Sensors,22(16), 6099. https://doi.org/10.3390/s22166099

    Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further detailshere.

    Article Metrics

    No
    No

    Article Access Statistics

    For more information on the journal statistics, clickhere.
    Multiple requests from the same IP address are counted as one view.
    Sensors, EISSN 1424-8220, Published by MDPI
    RSSContent Alert

    Further Information

    Article Processing Charges Pay an Invoice Open Access Policy Contact MDPI Jobs at MDPI

    Guidelines

    For Authors For Reviewers For Editors For Librarians For Publishers For Societies For Conference Organizers

    MDPI Initiatives

    Sciforum MDPI Books Preprints.org Scilit SciProfiles Encyclopedia JAMS Proceedings Series

    Follow MDPI

    LinkedIn Facebook X
    MDPI

    Subscribe to receive issue release notifications and newsletters from MDPI journals

    © 1996-2025 MDPI (Basel, Switzerland) unless otherwise stated
    Terms and Conditions Privacy Policy
    We use cookies on our website to ensure you get the best experience.
    Read more about our cookieshere.
    Accept
    Back to TopTop
    [8]ページ先頭

    ©2009-2025 Movatter.jp