Perfusion MRI | |
---|---|
![]() MRI perfusion showing a delayed time-to-maximum flow (Tmax) in thepenumbra in a case of occlusion of the leftmiddle cerebral artery. | |
Purpose | perfusion scanning via MRI |
Perfusion MRI orperfusion-weighted imaging (PWI) isperfusion scanning by the use of a particularMRI sequence[which?]. The acquired data are then post-processed to obtain perfusion maps with different parameters, such as BV (blood volume), BF (blood flow), MTT (mean transit time) and TTP (time to peak).
Incerebral infarction, thepenumbra has decreased perfusion.[1] Another MRI sequence,diffusion weighted MRI, estimates the amount of tissue that is already necrotic, and the combination of those sequences can therefore be used to estimate the amount of brain tissue that is salvageable bythrombolysis and/orthrombectomy.[1]
There are 3 main techniques for perfusion MRI:
It can also be argued that diffusion MRI models, such asintravoxel incoherent motion, also attempt to capture perfusion.
InDynamic susceptibility contrast MR imaging (DSC-MRI, or simply DSC),Gadolinium contrast agent (Gd) is injected (usually intravenously) and a time series of fastT2*-weighted images is acquired. As Gadolinium passes through the tissues, it induces a reduction of T2* in the nearby water protons; the corresponding decrease in signal intensity observed depends on the local Gd concentration, which may be considered a proxy for perfusion. The acquired time series data are then postprocessed to obtain perfusion maps with different parameters, such as BV (blood volume), BF (blood flow), MTT (mean transit time) and TTP (time to peak).
Dynamic contrast-enhanced (DCE) imaging gives information about physiological tissue characteristics such transport from blood to tissue and blood volume. It is typically used to measure how acontrast agent moves from the blood to the tissue. The concentration of the contrast agent is measured as it passes from the blood vessels to theextracellular space of the tissue (it does not pass the membranes ofcells) and as it goes back to the blood vessels.[5][6]
The contrast agents used for DCE-MRI are oftengadolinium based. Interaction with the gadolinium (Gd) contrast agent (commonly a gadolinium ion chelate) causes the relaxation time of water protons to decrease, and therefore images acquired after gadolinium injection display higher signal in T1-weighted images indicating the present of the agent. It is important to note that, unlike some techniques such asPET imaging, the contrast agent is not imaged directly, but by an indirect effect on water protons. The common procedure for a DCE-MRI exam is to acquire a regular T1-weighted MRI scan (with no gadolinium), and then gadolinium is injected (usually as an intravenous bolus at a dose of 0.05–0.1 mmol/kg) before further T1-weighted scanning. DCE-MRI may be acquired with or without a pause for contrast injection and may have varying time resolution depending on preference – faster imaging (less than 10s per imaging volume) allows pharmacokinetic (PK) modelling of contrast agent but can limit possible image resolution. Slower time resolution allows more detailed images, but may limit interpretation to only looking at signal intensity curve shape. In general, persistent increased signal intensity (corresponding to decreased T1 and thus increased Gd interaction) in a DCE-MRI imagevoxel indicates permeable blood vessels characteristic of tumor tissue, where Gd has leaked into the extravascular extracellular space. In tissues with healthy cells or a high cell density, gadolinium re-enters the vessels faster since it cannot pass the cell membranes. In damaged tissues or tissues with a lower cell density, the gadolinium stays in the extracellular space longer.
Pharmacokinetic modelling of gadolinium in DCE-MRI is complex and requires choosing a model. There are a variety of models, which describe tissue structure differently, including size and structure of plasma fraction, extravascular extracellular space, and the resulting parameters relating to permeability, surface area, and transfer constants.[7] DCE-MRI can also provide model-independent parameters, such as T1 (which is not technically part of the contrast scan and can be acquired independently) and (initial) area under the gadolinium curve (IAUGC, often given with number of seconds from injection - i.e., IAUGC60), which may be more reproducible.[8] Accurate measurement of T1 is required for some pharmacokinetic models, which can be estimated from 2 pre-gadolinium images of varying excitation pulse flip angle,[9] though this method is not intrinsically quantitative.[10] Some models require knowledge of the arterial input function, which may be measured on a per patient basis or taken as a population function from literature, and can be an important variable for modelling.[11]
Arterial spin labelling (ASL) has the advantage of not relying on an injected contrast agent, instead inferring perfusion from a drop in signal observed in the imaging slice arising from inflowing spins (outside the imaging slice) having been selectively inverted or saturated. A number of ASL schemes are possible, the simplest being flow alternating inversion recovery (FAIR) which requires two acquisitions of identical parameters with the exception of the out-of-slice inversion; the difference in the two images is theoretically only from inflowing spins, and may be considered a 'perfusion map'.
{{cite journal}}
: CS1 maint: multiple names: authors list (link)