Medical ultrasound includesdiagnostic techniques (mainlyimaging) usingultrasound, as well astherapeutic applications of ultrasound. In diagnosis, it is used to create an image of internal body structures such astendons,muscles, joints, blood vessels, and internal organs, to measure some characteristics (e.g., distances and velocities) or to generate an informative audible sound. The usage of ultrasound to produce visual images for medicine is calledmedical ultrasonography or simplysonography. Sonography using ultrasound reflection is calledechography. There are also transmission methods, such asultrasound transmission tomography. The practice of examiningpregnant women using ultrasound is calledobstetric ultrasonography, and was an early development of clinical ultrasonography. The machine used is called anultrasound machine, asonograph or anechograph. The visual image formed using this technique is called anultrasonogram, asonogram or anechogram.
Ultrasound is composed ofsound waves withfrequencies greater than 20,000 Hz, which is the approximate upper threshold of human hearing.[1] Ultrasonic images, also known as sonograms, are created by sending pulses of ultrasound intotissue using aprobe. The ultrasoundpulses echo off tissues with different reflection properties and are returned to the probe which records and displays them as an image.
A general-purposeultrasonic transducer may be used for most imaging purposes but some situations may require the use of a specialized transducer. Most ultrasound examination is done using a transducer on the surface of the body, but improved visualization is often possible if a transducer can be placed inside the body. For this purpose, special-use transducers, includingtransvaginal, endorectal, andtransesophageal transducers are commonly employed. At the extreme, very small transducers can be mounted on small diameter catheters and placed within blood vessels to image the walls and disease of those vessels.
The imaging mode refers to probe and machine settings that result in specific dimensions of the ultrasound image.[2]Several modes of ultrasound are used in medical imaging:[3][4]
A-mode: Amplitude mode refers to the mode in which the amplitude of thetransducer voltage is recorded as a function of two-way travel time of an ultrasound pulse. A single pulse is transmitted through the body and scatters back to the same transducer element. The voltage amplitudes recorded correlate linearly to acoustic pressure amplitudes. A-mode is one-dimensional.
B-mode: In brightness mode, an array of transducer elements scans a plane through the body resulting in a two-dimensional image. Each pixel value of the image correlates to voltage amplitude registered from the backscattered signal. The dimensions of B-mode images are voltage as a function of angle and two-way time.
M-mode: In motion mode, A-mode pulses are emitted in succession. The backscattered signal is converted to lines of bright pixels, whose brightness linearly correlates to backscattered voltage amplitudes. Each next line is plotted adjacent to the previous, resulting in an image that looks like a B-mode image. The M-mode image dimensions are however voltage as a function of two-way time and recording time. This mode is an ultrasound analogy to streakvideo recording in high-speed photography. As moving tissue transitions produce backscattering, this can be used to determine the displacement of specific organ structures, most commonly the heart.
Most machines convert two-way time to imaging depth using as assumed speed of sound of 1540 m/s. As the actual speed of sound varies greatly in different tissue types, anultrasound image is therefore not a true tomographic representation of the body.[5]
Three-dimensional imaging is done by combining B-mode images, using dedicated rotating or stationary probes. This has been referred to asC-mode.[4]
A hybrid rotating ultrasound/photoacoustic system produces 3D images, including blood and tissue without the use of magnets or ionizing radiation. It uses an arc of detectors to create a volumetric image. A laser pulse causes hemoglobin molecules to vibrate and generate acoustic signal also absorbed by the detectors.[6]
An imaging technique is the method of signal generation and processing that supports in a specific application. Most techniques use B-mode.
Doppler sonography: This imaging technique makes use of theDoppler effect in detection and measuring moving targets, typically blood.
Harmonic imaging: backscattered signal from tissue is filtered to comprise only frequency content of at least twice the centre frequency of the transmitted ultrasound. Harmonic imaging used for perfusion detection when using ultrasoundcontrast agents and for the detection of tissue harmonics. Common pulse schemes for the creation of harmonic response without the need of real-timeFourier analysis are pulse inversion and power modulation.[7]
B-flow is an imaging technique that digitally highlights moving reflectors (mainlyred blood cells) while suppressing the signals from the surrounding stationary tissue. It aims to visualize flowing blood and surrounding stationary tissues simultaneously.[8] It is thus an alternative or complement toDoppler ultrasonography in visualizing blood flow.[9]
Therapeutic ultrasound aimed at a specific tumor or calculus is not an imaging mode. However, for positioning a treatment probe to focus on a specific region of interest, A-mode and B-mode are typically used, often during treatment.[10]
Compared to other medical imaging modalities, ultrasound has several advantages. It provides images in real-time,[11] isportable, and can consequently be brought to the bedside. It is substantially lower in cost than other imaging strategies. Drawbacks include various limits on its field of view, the need for patient cooperation, dependence on patient physique, difficulty imaging structures obscured bybone, air or gases,[note 1] and the necessity of a skilled operator, usually with professional training.
Urinary bladder (black butterfly-like shape) and hyperplasticprostate (BPH) visualized by medical ultrasound
Sonography (ultrasonography) is widely used inmedicine. It is possible to perform bothdiagnosis andtherapeutic procedures, usingultrasound to guide interventional procedures such asbiopsies or to drain collections of fluid, which can be both diagnostic and therapeutic.Sonographers are medical professionals who perform scans which are traditionally interpreted by radiologists, physicians who specialize in the application and interpretation of medical imaging modalities, or by cardiologists in the case of cardiac ultrasonography (echocardiography). Sonography is effective for imaging soft tissues of the body.[12] Superficial structures such asmuscle,tendon,testis,breast,thyroid and parathyroid glands, and theneonatal brain are imaged at higherfrequencies (7–18 MHz), which provide better linear (axial) and horizontal (lateral)resolution. Deeper structures such as liver and kidney are imaged at lower frequencies (1–6 MHz) with lower axial and lateral resolution as a price of deeper tissue penetration.[citation needed]
Inanesthesiology, ultrasound is commonly used to guide the placement of needles when injecting local anesthetic solutions in the proximity ofnerves identified within the ultrasound image (nerve block). It is also used for vascular access such ascannulation of large central veins and for difficultarterial cannulation.Transcranial Doppler is frequently used by neuro-anesthesiologists for obtaining information about flow-velocity in the basalcerebral vessels.[citation needed]
Intravascular ultrasound image of a coronary artery (left), with color-coding on the right, delineating the lumen (yellow), external elastic membrane (blue) and the atherosclerotic plaque burden (green)
Ultrasonography of liver tumors allows for both detection and characterization.[15]Ultrasound imaging studies are often obtained during the evaluation process ofFatty liver disease. Ultrasonography reveals a "bright" liver with increased echogenicity. Pocket-sized ultrasound devices might be used as point-of-care screening tools to diagnose liver steatosis.[16][17]
Obstetrical sonography was originally developed in the late 1950s and 1960s by SirIan Donald[21][22] and is commonly used duringpregnancy to check the development and presentation of thefetus. It can be used to identify many conditions that could be potentially harmful to the mother and/or baby possibly remaining undiagnosed or with delayed diagnosis in the absence of sonography. It is currently believed that the risk of delayed diagnosis is greater than the small risk, if any, associated with undergoing an ultrasound scan. However, its use for non-medical purposes such as fetal "keepsake" videos and photos is discouraged.[23]
According to the European Committee of Medical Ultrasound Safety (ECMUS)[24]
Ultrasonic examinations should only be performed by competent personnel who are trained and updated in safety matters. Ultrasound produces heating, pressure changes and mechanical disturbances in tissue. Diagnostic levels of ultrasound can produce temperature rises that are hazardous to sensitive organs and the embryo/fetus. Biological effects of non-thermal origin have been reported in animals but, to date, no such effects have been demonstrated in humans, except when a micro-bubblecontrast agent is present.
Nonetheless, care should be taken to use low power settings and avoid pulsed wave scanning of the fetal brain unless specifically indicated in high risk pregnancies.[citation needed]
Figures released for the period 2005–2006 by the UK Government (Department of Health) show that non-obstetric ultrasound examinations constituted more than 65% of the total number of ultrasound scans conducted.
Most structures of the neck, including thethyroid andparathyroid glands,[26]lymph nodes, andsalivary glands, are well-visualized by high-frequency ultrasound with exceptional anatomic detail. Ultrasound is the preferred imaging modality for thyroid tumors and lesions, and its use is important in the evaluation, preoperative planning, and postoperative surveillance of patients withthyroid cancer. Many other benign and malignant conditions in the head and neck can be differentiated, evaluated, and managed with the help of diagnostic ultrasound and ultrasound-guided procedures.[citation needed]
Inneonatology,transcranial Doppler can be used for basic assessment of intracerebral structural abnormalities, suspected hemorrhage,ventriculomegaly orhydrocephalus and anoxic insults (periventricular leukomalacia). It can be performed through the soft spots in the skull of a newborn infant (Fontanelle) until these completely close at about 1 year of age by which time they have formed a virtually impenetrable acoustic barrier to ultrasound.[27] The most common site forcranial ultrasound is the anterior fontanelle. The smaller the fontanelle, the more the image is compromised.[citation needed]
Lung ultrasound has been found to be useful in diagnosing common neonatal respiratory diseases such as transient tachypnea of the newborn, respiratory distress syndrome, congenital pneumonia, meconium aspiration syndrome, and pneumothorax.[28] A neonatal lung ultrasound score, first described by Brat et al., has been found to highly correlate with oxygenation in the newborn.[29][30]
B-scan ultrasonography, orB-scan-Brightness scan, is aB-mode scan that produces a cross-sectional view of theeye and theorbit. It is an essential tool in ophthalmology for diagnosing and managing a wide array of conditions affecting the posterior segment of the eye.It is non invasive and uses frequency 10–15 MHz. It is often used in conjunction with other imaging techniques (like OCT or fluorescein angiography) for a more comprehensive evaluation of ocular conditions.
Ultrasound is used to assess thelungs in a variety of settings including critical care, emergency medicine, trauma surgery, as well as general medicine and nursing. This imaging modality is used at the bedside or examination table to evaluate a number of different lung abnormalities as to guiderespiratory therapy andmechanical ventilation, and procedures such asthoracentesis, (drainage of pleural fluid (effusion)), needle aspiration biopsy, andcatheter placement.[31] Although air present in the lungs does not allow good penetration of ultrasound waves, interpretation of specific artifacts created on the lung surface can be used to detect abnormalities.[32]
Lung, or pulmonary ultrasound does not visualise the lung itself, but the tissue-air-interface at the pleural line. The fact that artefact interpretation is essential sets lung ultrasound apart from modalities aimed at imaging an actual organ. This distinction has mechanical implications that are important when trying to interpret artifactual patterns. Where modern sonography is aided by software filters and acoustic harmonics to enhance visualisation of organs such as the heart or liver, they corrupt the specific patterns sought in lung ultrasound. Therefore, a software preset void of imaging filters, harmonic imaging, and beam compounding, is preferred.[33]
The Normal Lung Surface: The lung surface is composed of visceral and parietalpleura. These two surfaces are typically pushed together and make up the pleural line, which is the basis of lung (or pleural) ultrasound. This line is visible less than a centimeter below the rib line in most adults. On ultrasound, it is visualized as ahyperechoic (bright white) horizontal line if the ultrasound probe is applied perpendicularly to the skin.
Artifacts: Lung ultrasound relies on artifacts, which would otherwise be considered a hindrance in imaging. Air blocks the ultrasound beam and thus visualizing healthy lung tissue itself with this mode of imaging is not practical. Consequently, physicians and sonographers have learned to recognize patterns that ultrasound beams create when imaging healthy versus diseased lung tissue. Three commonly seen and utilized artifacts in lung ultrasound include lung sliding, A-lines, and B-lines.[34]
§ Lung Sliding: The presence of lung sliding, which indicates the shimmering of the pleural line that occurs with movement of the visceral and parietal pleura against one another with respiration (sometimes described as 'ants marching'), is the most important finding in normal aerated lung.[35] Lung sliding indicates both that the lung is present at the chest wall and that the lung is functioning.[34]
§ A-lines: When the ultrasound beam makes contact with thepleural line, it is reflected back creating a bright white horizontal line. The subsequent reverberation artifacts that appear as equally spaced horizontal lines deep to the pleura are A-lines. Ultimately, A-lines are a reflection of the ultrasound beam from the pleura with the space between A-lines corresponding to the distance between the parietal pleura and the skin surface.[34] A-lines indicate the presence of air, which means that these artifacts can be present in normal healthy lung (and also in patients with pneumothorax).[35]
§ B-lines: B-lines are also reverberation artifacts. They are visualized ashyperechoic vertical lines extending from the pleura to the edge of the ultrasound screen. These lines are sharply defined and laser-like and typically do not fade as they progress down the screen.[34] A few B-lines that move along with the sliding pleura can be seen in normal lung due to acoustic impedance differences between water and air. However, excessive B-lines (three or more) are abnormal and are typically indicative of underlying lung pathology.[35]
Pulmonary edema: Lung ultrasound has been shown to be very sensitive for the detection of pulmonary edema. It allows for improvement in diagnosis and management of critically ill patients, particularly when used in combination with echocardiography. The sonographic feature that is present in pulmonary edema is multiple B-lines. B-lines can occur in a healthy lung; however, the presence of 3 or more in the anterior or lateral lung regions is always abnormal. In pulmonary edema, B-lines indicate an increase in the amount of water contained in the lungs outside of the pulmonary vasculature. B-lines can also be present in a number of other conditions including pneumonia, pulmonary contusion, and lung infarction.[36] Additionally, it is important to note that there are multiple types of interactions between the pleural surface and the ultrasound wave that can generate artifacts with some similarity to B-lines but which do not have pathologic significance.[37]
Atelectasis: Consolidations are, as opposed to artifactual patterns, directly visualised on lung ultrasound. Differentiation of consolidations within the lung can be made with qualitative methods such as the presence of dynamic, or static air bronchograms, being air trapped within smaller airways inside of consolidations. Pathognomic changes in resorptive andcompressive atelectasis, respectively can thus be distinguished.
Pneumothorax: In clinical settings when pneumothorax is suspected, lung ultrasound can aid in diagnosis.[38] In pneumothorax, air is present between the two layers of the pleura and lung sliding on ultrasound is therefore absent. Thenegative predictive value for lung sliding on ultrasound is reported as 99.2–100% – briefly, if lung sliding is present, a pneumothorax is effectively ruled out.[35] The absence of lung sliding, however, is not necessarily specific for pneumothorax as there are other conditions that also cause this finding includingacute respiratory distress syndrome,lung consolidations, pleural adhesions, andpulmonary fibrosis.[35]
Pleural effusion: Lung ultrasound is a cost-effective, safe, and non-invasive imaging method that can aid in the prompt visualization and diagnosis of pleural effusions. Effusions can be diagnosed by a combination of physical exam, percussion, andauscultation of the chest. However, these exam techniques can be complicated by a variety of factors including the presence ofmechanical ventilation, obesity, or patient positioning, all of which reduce the sensitivity of the physical exam. Consequently, lung ultrasound can be an additional tool to augment plainchest Xray andchest CT.[39] Pleural effusions on ultrasound appear as structural images within the thorax rather than an artifact. They will typically have four distinct borders including the pleural line, two rib shadows, and a deep border.[34] In critically ill patients with pleural effusion, ultrasound may guide procedures including needle insertion,thoracentesis, andchest-tube insertion.[39]
Lung cancer staging: Inpulmonology, endobronchial ultrasound (EBUS) probes are applied to standard flexible endoscopic probes and used by pulmonologists to allow for direct visualization of endobronchial lesions and lymph nodes prior to transbronchial needle aspiration. Among its many uses, EBUS aids in lung cancer staging by allowing for lymph node sampling without the need for major surgery.[40]
COVID-19: Lung ultrasound has proved useful in the diagnosis of COVID-19 especially in cases where other investigations are not available.[41][42][43]
Urinary bladder (black butterfly-like shape) and hyperplasticprostate (BPH) visualized by medical sonographic technique
Ultrasound is routinely used inurology to determine the amount of fluid retained in a patient's bladder. In a pelvic sonogram, images include theuterus andovaries orurinary bladder in females. In males, a sonogram will provide information about the bladder,prostate, ortesticles (for example to urgently distinguishepididymitis fromtesticular torsion). In young males, it is used to distinguish more benign testicular masses (varicocele orhydrocele) fromtesticular cancer, which is curable but must be treated to preserve health and fertility. There are two methods of performing pelvic sonography – externally or internally. The internal pelvic sonogram is performed either transvaginally (in a woman) or transrectally (in a man). Sonographic imaging of the pelvic floor can produce important diagnostic information regarding the precise relationship of abnormal structures with other pelvic organs and it represents a useful hint to treat patients with symptoms related to pelvic prolapse, double incontinence and obstructed defecation.[citation needed] It is also used to diagnose and, at higher frequencies, to treat (break up) kidney stones or kidney crystals (nephrolithiasis).[44]
Ultrasound is an excellent method for the study of thepenis, such as indicated in trauma, priapism, erectile dysfunction or suspectedPeyronie's disease.[46]
Musculoskeletal ultrasound is used to examine tendons, muscles, nerves, ligaments, soft tissue masses, and bone surfaces.[47]It is helpful in diagnosing ligament sprains, muscles strains and joint pathology. It is an alternative or supplement to x-ray imaging in detecting fractures of the wrist, elbow and shoulder for patients up to 12 years[48] (Fracture sonography).
Quantitative ultrasound is an adjunct musculoskeletal test for myopathic disease in children;[49][50] estimates of lean body mass in adults;[51] proxy measures of muscle quality (i.e., tissue composition)[52] in older adults withsarcopenia[53][54]
Innephrology, ultrasonography of the kidneys is essential in the diagnosis and management of kidney-related diseases. The kidneys are easily examined, and most pathological changes are distinguishable with ultrasound. It is an accessible, versatile, relatively economic, and fast aid for decision-making in patients with renal symptoms and for guidance in renal intervention.[55] UsingB-mode imaging, assessment of renal anatomy is easily performed, and US is often used as image guidance for renal interventions. Furthermore, novel applications in renal US have been introduced with contrast-enhanced ultrasound (CEUS), elastography and fusion imaging. However, renal US has certain limitations, and other modalities, such as CT (CECT) and MRI, should be considered for supplementary imaging in assessing renal disease.[55]
Intravenous access, for the collection of blood samples to assist in diagnosis or laboratory investigation including blood culture, or for administration of intravenous fluids for fluid maintenance of replacement or blood transfusion in sicker patients, is a common medical procedure. The need for intravenous access occurs in the outpatient laboratory, in the inpatient hospital units, and most critically in the Emergency Room and Intensive Care Unit. In many situations, intravenous access may be required repeatedly or over a significant time period. In these latter circumstances, a needle with an overlying catheter is introduced into the vein and the catheter is then inserted securely into the vein while the needle is withdrawn.The chosen veins are most frequently selected from the arm, but in challenging situations, a deeper vein from the neck (external jugular vein) or upper arm (subclavian vein) may need to be used.There are many reasons why the selection of a suitable vein may be problematic.These include, but are not limited to, obesity, previous injury to veins from inflammatory reaction to previous 'blood draws', previous injury to veins from recreational drug use.[citation needed]
In these challenging situations, the insertion of a catheter into a vein has been greatly assisted by the use of ultrasound. The ultrasound unit may be 'cart-based' or 'handheld' using a linear transducer with a frequency of 10 to 15megahertz. In most circumstances, choice of vein will be limited by the requirement that the vein is within 1.5 cms. from the skin surface.The transducer may be placed longitudinally or transversely over the chosen vein.Ultrasound training for intravenous cannulation is offered in most ultrasound training programs.[citation needed]
A sound wave is typically produced by apiezoelectric transducer encased in a plastic housing. Strong, short electrical pulses from the ultrasound machine drive thetransducer at the desired frequency. Thefrequencies can vary between 1 and 18MHz, though frequencies up to 50–100 megahertz have been used experimentally in a technique known as biomicroscopy in special regions, such as the anterior chamber of the eye.[56]
Older technology transducers focused their beam with physical lenses.[citation needed] Contemporary technology transducers usedigital antenna array techniques (piezoelectric elements in the transducer produce echoes at different times) to enable the ultrasound machine to change the direction and depth of focus. Near the transducer, the width of the ultrasound beam almost equals to the width of the transducer, after reaching a distance from the transducer (near zone length orFresnel zone), the beam width narrows to half of the transducer width, and after that the width increases (far zone length orFraunhofer's zone), where the lateral resolution decreases. Therefore, the wider the transducer width and the higher the frequency of ultrasound, the longer the Fresnel zone, and the lateral resolution can be maintained at a greater depth from the transducer.[57] Ultrasound waves travel in pulses. Therefore, a shorter pulse length requires higher bandwidth (greater number of frequencies) to constitute the ultrasound pulse.[7]
As stated, the sound is focused either by the shape of the transducer, a lens in front of the transducer, or a complex set of control pulses from the ultrasound scanner, in thebeamforming or spatial filtering technique. This focusing produces an arc-shaped sound wave from the face of the transducer. The wave travels into the body and comes into focus at a desired depth.
Materials on the face of the transducer enable the sound to be transmitted efficiently into the body (often a rubbery coating, a form ofimpedance matching).[58] In addition, a water-based gel is placed between the patient's skin and the probe to facilitate ultrasound transmission into the body. This is because air causes total reflection of ultrasound; impeding the transmission of ultrasound into the body.[59]
The sound wave is partially reflected from the layers between different tissues or scattered from smaller structures. Specifically, sound is reflected anywhere where there are acoustic impedance changes in the body: e.g.blood cells inblood plasma, small structures in organs, etc. Some of the reflections return to the transducer.[58]
The return of the sound wave to the transducer results in the same process as sending the sound wave, in reverse. The returned sound wave vibrates the transducer and the transducer turns the vibrations into electrical pulses that travel to the ultrasonic scanner where they are processed and transformed into a digital image.[60]
To make an image, the ultrasound scanner must determine two characteristics from each received echo:
How long it took the echo to be received from when the sound was transmitted. (Time and distance are equivalent.)
How strong the echo was.
Once the ultrasonic scanner determines these two, it can locate which pixel in the image to illuminate and with what intensity.
Transforming the received signal into a digital image may be explained by using a blank spreadsheet as an analogy. First picture a long, flat transducer at the top of the sheet. Send pulses down the 'columns' of the spreadsheet (A, B, C, etc.). Listen at each column for any return echoes. When an echo is heard, note how long it took for the echo to return. The longer the wait, the deeper the row (1,2,3, etc.). The strength of the echo determines the brightness setting for that cell (white for a strong echo, black for a weak echo, and varying shades of grey for everything in between.) When all the echoes are recorded on the sheet, agreyscale image has been accomplished.
In modern ultrasound systems, images are derived from the combined reception of echoes by multiple elements, rather than a single one. These elements in the transducer array work together to receive signals, a process essential for optimizing the ultrasonic beam's focus and producing detailed images. One predominant method for this is "delay-and-sum" beamforming. The time delay applied to each element is calculated based on the geometrical relationship between the imaging point, the transducer, and receiver positions. By integrating these time-adjusted signals, the system pinpoints focus onto specific tissue regions, enhancing image resolution and clarity. The utilization of multiple element reception combined with the delay-and-sum principles underpins the high-quality images characteristic of contemporary ultrasound scans.[61]
Images from the ultrasound scanner are transferred and displayed using theDICOM standard. Normally, very little post processing is applied.[citation needed]
Ultrasonography (sonography) uses a probe containing multiple acoustictransducers to send pulses of sound into a material. Whenever a sound wave encounters a material with a different density (acoustical impedance), some of the sound wave is scattered but part is reflected back to the probe and is detected as an echo. The time it takes for theecho to travel back to the probe is measured and used to calculate the depth of the tissue interface causing the echo. The greater the difference between acoustic impedances, the larger the echo is. If the pulse hits gases or solids, the density difference is so great that most of the acoustic energy is reflected and it becomes impossible to progress further.[citation needed]
The frequencies used for medical imaging are generally in the range of 1 to 18 MHz Higher frequencies have a correspondingly smaller wavelength, and can be used to make more detailed sonograms. However, the attenuation of the sound wave is increased at higher frequencies, so penetration of deeper tissues necessitates a lower frequency (3–5 MHz).
Penetrating deep into the body with sonography is difficult. Some acoustic energy is lost each time an echo is formed, but most of it (approximately) is lost from acoustic absorption. (SeeAcoustic attenuation for further details on modeling of acoustic attenuation and absorption.)
The speed of sound varies as it travels through different materials, and is dependent on theacoustical impedance of the material. However, the sonographic instrument assumes that the acoustic velocity is constant at 1540 m/s. An effect of this assumption is that in a real body with non-uniform tissues, the beam becomes somewhat de-focused and image resolution is reduced.
To generate a2-D image, the ultrasonic beam is swept. A transducer may be swept mechanically by rotating or swinging or a 1-Dphased array transducer may be used to sweep the beam electronically. The received data is processed and used to construct the image. The image is then a 2-D representation of the slice into the body.
3-D images can be generated by acquiring a series of adjacent 2-D images. Commonly a specialized probe that mechanically scans a conventional 2-D image transducer is used. However, since the mechanical scanning is slow, it is difficult to make 3D images of moving tissues. Recently, 2-D phased array transducers that can sweep the beam in 3-D have been developed. These can image faster and can even be used to make live 3-D images of a beating heart.
Doppler ultrasonography is used to study blood flow and muscle motion. The different detected speeds are represented in color for ease of interpretation, for example leaky heart valves: the leak shows up as a flash of unique color. Colors may alternatively be used to represent the amplitudes of the received echoes.
An additional expansion of ultrasound isbi-planar ultrasound, in which the probe has two 2D planes perpendicular to each other, providing more efficient localization and detection.[62] Furthermore, anomniplane probe can rotate 180° to obtain multiple images.[62] In3D ultrasound, many 2D planes are digitally added together to create a 3-dimensional image of the object.
Doppler ultrasonography employs theDoppler effect to assess whether structures (usually blood)[60][63] are moving towards or away from the probe, and their relative velocity. By calculating the frequency shift of a particular sample volume, flow in an artery or a jet of blood flow over a heart valve, its speed and direction can be determined and visualized, as an example.Color Doppler is the measurement of velocity by color scale. Color Doppler images are generally combined with gray scale (B-mode) images to displayduplex ultrasonography images.[64] Uses include:
Doppler echocardiography is the use of Doppler ultrasonography to examine theheart.[65] An echocardiogram can, within certain limits, produce accurate assessment of the direction ofblood flow and thevelocity of blood and cardiac tissue at any arbitrary point using the Doppler effect. Velocity measurements allow assessment ofcardiac valve areas and function, abnormal communications between the left and right side of the heart, leaking of blood through the valves (valvular regurgitation), and calculation of thecardiac output andE/A ratio[66] (a measure ofdiastolic dysfunction). Contrast-enhanced ultrasound using gas-filled microbubble contrast media can be used to improve velocity or other flow-related measurements of interest.
Doppler fetal monitors use the Doppler effect to detect thefetal heartbeat duringprenatal care. These are hand-held, and some models also display theheart rate in beats per minute (BPM). Use of this monitor is sometimes known asDopplerauscultation. The Doppler fetal monitor is commonly referred to simply as aDoppler orfetal Doppler and provides information similar to that provided by afetal stethoscope.
Microbubbles-based contrast media is administeredintravenously into the patientblood stream during the ultrasonography examination. Due to their size, the microbubbles remain confined inblood vessels without extravasating towards theinterstitial fluid. Anultrasound contrast media is therefore purely intravascular, making it an ideal agent to imageorgan microvasculature fordiagnostic purposes. A typical clinical use of contrast ultrasonography is detection of ahypervascularmetastatic tumor, which exhibits a contrast uptake (kinetics of microbubbles concentration in blood circulation) faster than healthybiological tissue surrounding thetumor.[70] Other clinical applications using contrast exist, as in echocardiography to improve delineation ofleft ventricle for visualizing contractibility ofheart muscle after amyocardial infarction. Finally, applications in quantitative perfusion[71] (relative measurement ofblood flow[72]) have emerged for identifying early patient response to anti-cancerous drug treatment (methodology and clinical study by Dr. Nathalie Lassau in 2011[73]), enabling the best oncologicaltherapeutic options to be determined.[74]
Parametric imaging of vascular signatures (diagram)
Inoncological practice of medical contrast ultrasonography, clinicians use 'parametric imaging of vascular signatures'[75] invented by Dr. Nicolas Rognin in 2010.[76] This method is conceived as acancer aided diagnostic tool, facilitating characterization of a suspicioustumor (malignant versusbenign) in an organ. This method is based on medicalcomputational science[77][78] to analyze a time sequence of ultrasound contrast images, a digital video recorded in real-time during patient examination. Two consecutivesignal processing steps are applied to eachpixel of the tumor:
calculation of a vascular signature (contrast uptake difference with respect to healthy tissue surrounding the tumor);
automaticclassification of the vascular signature into a uniqueparameter, the latter coded in one of the four followingcolors:
green for continuous hyper-enhancement (contrast uptake higher than healthy tissue one),
blue for continuous hypo-enhancement (contrast uptake lower than healthy tissue one),
red for fast hyper-enhancement (contrast uptake before healthy tissue one) or
yellow for fast hypo-enhancement (contrast uptake after healthy tissue one).
Oncesignal processing in each pixel is completed, a color spatial map of the parameter is displayed on acomputer monitor, summarizing allvascular information of the tumor in a single image called a parametric image (see last figure of press article[79] as clinical examples). This parametric image is interpreted by clinicians based on predominant colorization of the tumor: red indicates a suspicion ofmalignancy (risk of cancer), green or yellow – a high probability of benignity. In the first case (suspicion ofmalignant tumor), the clinician typically prescribes a biopsy to confirm the diagnostic or aCT scan examination as a second opinion. In the second case (quasi-certain ofbenign tumor), only a follow-up is needed with a contrast ultrasonography examination a few months later. The main clinical benefits are to avoid a systemic biopsy (with inherent risks of invasive procedures) of benign tumors or a CT scan examination exposing the patient toX-ray radiation. The parametric imaging of vascular signatures method proved to be effective in humans for characterization of tumors in the liver.[80] In acancer screening context, this method might be potentially applicable to other organs such asbreast[81] orprostate.
The current future of contrast ultrasonography is inmolecular imaging with potential clinical applications expected incancer screening to detectmalignant tumors at their earliest stage of appearance. Molecular ultrasonography (or ultrasound molecular imaging) uses targeted microbubbles originally designed by DrAlexander Klibanov in 1997;[82][83] such targeted microbubbles specifically bind or adhere to tumoral microvessels by targetingbiomolecular cancer expression (overexpression of certain biomolecules that occurs duringneo-angiogenesis[84][85] orinflammation[86] in malignant tumors). As a result, a few minutes after their injection in blood circulation, the targeted microbubbles accumulate in the malignant tumor; facilitating its localization in a unique ultrasound contrast image. In 2013, the very first exploratoryclinical trial in humans forprostate cancer was completed atAmsterdam in theNetherlands by Dr. Hessel Wijkstra.[87]
In molecular ultrasonography, the technique ofacoustic radiation force (also used forshear wave elastography) is applied in order to literally push the targeted microbubbles towards microvessels wall; first demonstrated by Dr. Paul Dayton in 1999.[88] This allows maximization of binding to the malignant tumor; the targeted microbubbles being in more direct contact with cancerous biomolecules expressed at the inner surface of tumoral microvessels. At the stage of scientificpreclinical research, the technique of acoustic radiation force was implemented as a prototype in clinical ultrasound systems and validatedin vivo in 2D[89] and 3D[90][91] imaging modes.
Ultrasound is also used for elastography, which is a relatively new imaging modality that maps the elastic properties of soft tissue.[92][93] This modality emerged in the last two decades. Elastography is useful in medical diagnoses as it can discern healthy from unhealthy tissue for specific organs/growths. For example, cancerous tumors will often be harder than the surrounding tissue, and diseased livers are stiffer than healthy ones.[92][93][94][95]
There are many ultrasound elastography techniques.[93]
Thyroid cysts: High frequency thyroidultrasound (HFUS) can be used to treat several gland conditions. The recurrent thyroid cyst that was usually treated in the past with surgery, can be treated effectively by a new procedure called percutaneous ethanol injection, or PEI.[96][97] With ultrasound guided placement of a 25 gauge needle within the cyst, and after evacuation of the cyst fluid, about 50% of the cyst volume is injected back into the cavity, under strict operator visualization of the needle tip. The procedure is 80% successful in reducing the cyst to minute size.
Metastatic thyroid cancer neck lymph nodes: HFUS may also be used to treat metastatic thyroid cancer neck lymph nodes that occur in patients who either refuse, or are no longer candidates, for surgery. Small amounts of ethanol are injected under ultrasound guided needle placement. A power doppler blood flow study is done prior to injection. The blood flow can be destroyed and the node rendered inactive. Power doppler visualized blood flow can be eradicated, and there may be a drop in the cancer blood marker test,thyroglobulin, TG, as the node become non-functional. Another interventional use for HFUS is to mark a cancer node prior to surgery to help locate the node cluster at the surgery. A minute amount of methylene dye is injected, under careful ultrasound guided placement of the needle on the anterior surface, but not in the node. The dye will be evident to the thyroid surgeon when opening the neck. A similar localization procedure with methylene blue, can be done to locate parathyroid adenomas.[98]
Compression ultrasonography is when the probe is pressed against the skin. This can bring the target structure closer to the probe, increasing spatial resolution of it. Comparison of the shape of the target structure before and after compression can aid in diagnosis.
It is used inultrasonography of deep venous thrombosis, wherein absence of vein compressibility is a strong indicator of thrombosis.[99] Compression ultrasonography has both highsensitivity and specificity for detecting proximal deep vein thrombosis in symptomatic patients. Results are not reliable when the patient is asymptomatic, for example in high risk postoperative orthopedic patients.[100][101]
A normalappendix without and with compression. Absence of compressibility indicatesappendicitis.[102]
Panoramic ultrasonography of a proximalbiceps tendon rupture. Top image shows the contralateral normal side, and lower image shows a retracted muscle, with ahematoma filling out the proximal space.
Panoramic ultrasonography is the digitalstitching of multiple ultrasound images into a broader one.[103] It can display an entire abnormality and show its relationship to nearby structures on a single image.[103]
Multiparametric ultrasonography (mpUSS) combines multiple ultrasound techniques to produce a composite result. For example, one study combined B-mode, colour Doppler, real-time elastography, and contrast-enhanced ultrasound, achieving an accuracy similar to that ofmultiparametric MRI.[104]
Speed-of-sound (SoS) imaging aims to find the spatial distribution of the SoS within the tissue. The idea is to find relative delay measurements for different transmission events and solve the limited-angle tomographic reconstruction problem using delay measurements and transmission geometry. Compared to shear-wave elastography, SoS imaging has better ex-vivo tissue differentiation[105] for benign and malignant tumors.[106][107][108]
Muscle,soft tissue, and bone surfaces are imaged very well including the delineation of interfaces between solid and fluid-filled spaces.
"Live" images can be dynamically selected, permitting diagnosis and documentation often rapidly. Live images also permit ultrasound-guided biopsies or injections, which can be cumbersome with other imaging modalities.
Organ structure can be demonstrated.
There are no known long-term side effects when used according to guidelines, and discomfort is minimal.
Ability to image local variations in the mechanical properties of soft tissue.[109]
Equipment is widely available and comparatively flexible.
Small, easily carried scanners are available which permit bedside examinations.
Spatial resolution is better in high frequency ultrasound transducers than most other imaging modalities.
Use of anultrasound research interface can offer a relatively inexpensive, real-time, and flexible method for capturing data required for specific research purposes of tissue characterization and development of new image processing techniques.
Double aorta artifact in sonography due to difference in velocity of sound waves in muscle and fat
Sonographic devices have trouble penetratingbone. For example, sonography of the adult brain is currently very limited.
Sonography performs very poorly when there is gas between the transducer and the organ of interest, due to the extreme differences inacoustic impedance. For example, overlying gas in the gastrointestinal tract often makes ultrasound scanning of thepancreas difficult. This is also the case forsubcutaneous emphysema when attempting to examine the lungs.[110] Lung imaging however can be useful in demarcating pleural effusions, detectingheart failure and pneumonia.[111]
Even in the absence of bone or air, the depth penetration of ultrasound may be limited depending on the frequency of imaging. Consequently, there might be difficulties imaging structures deep in the body, especially in obese patients.
Image quality and accuracy of diagnosis is limited with obese patients and overlying subcutaneous fat attenuates the sound beam. A lower frequency transducer is required with subsequent lower resolution.
The method is operator-dependent. Skill and experience is needed to acquire good-quality images and make accurate diagnoses.
There is no scout image as there is with CT and MRI. Once an image has been acquired there is no exact way to tell which part of the body was imaged.
80% of sonographers experience Repetitive Strain Injuries (RSI) or so-called Work-Related Musculoskeletal Disorders (WMSD) because of bad ergonomic positions.
Ultrasonography is generally considered safe imaging,[112] with the World Health Organization stating:[113]
"Diagnostic ultrasound is recognized as a safe, effective, and highly flexible imaging modality capable of providing clinically relevant information about most parts of the body in a rapid and cost-effective fashion".
Diagnostic ultrasound studies of the fetus are generally considered to be safe during pregnancy. However, this diagnostic procedure should be performed only when there is a valid medical indication, and the lowest possible ultrasonic exposure setting should be used to gain the necessary diagnostic information under the "as low as reasonably practicable" orALARP principle.[114]
Although there is no evidence that ultrasound could be harmful to the fetus, medical authorities typically strongly discourage the promotion, selling, or leasing of ultrasound equipment for making "keepsake fetal videos".[23][115]
A meta-analysis of several ultrasonography studies published in 2000 found no statistically significant harmful effects from ultrasonography. It was noted that there is a lack of data on long-term substantive outcomes such as neurodevelopment.[116]
A study at theYale School of Medicine published in 2006 found a small but significant correlation between prolonged and frequent use of ultrasound and abnormal neuronal migration in mice.[117]
A study performed in Sweden in 2001[118] has shown that subtle effects of neurological damage linked to ultrasound were implicated by an increased incidence in left-handedness in boys (a marker for brain problems when not hereditary) and speech delays.[119][120]
The above findings, however, were not confirmed in a follow-up study.[121]
A later study, however, performed on a larger sample of 8865 children, has established a statistically significant, albeit weak association of ultrasonography exposure and being non-right handed later in life.[122]
Diagnostic andtherapeutic ultrasound equipment is regulated in the US by theFood and Drug Administration, and worldwide by other national regulatory agencies. The FDA limits acoustic output using several metrics; generally, other agencies accept the FDA-established guidelines.
The primary regulated metrics areMechanical Index (MI), a metric associated with the cavitation bio-effect, and Thermal Index (TI) a metric associated with the tissue heating bio-effect. The FDA requires that the machine not exceed established limits, which are reasonably conservative in an effort to maintain diagnostic ultrasound as a safe imaging modality. This requiresself-regulation on the part of the manufacturer in terms of machine calibration.[125]
Ultrasound-based pre-natal care and sex screening technologies were launched in India in the 1980s. With concerns about its misuse forsex-selective abortion, the Government of India passed thePre-natal Diagnostic Techniques Act (PNDT) in 1994 to distinguish and regulate legal and illegal uses of ultrasound equipment.[126] The law was further amended as the Pre-Conception and Pre-natal Diagnostic Techniques (Regulation and Prevention of Misuse) (PCPNDT) Act in 2004 to deter and punish prenatal sex screening and sex selective abortion.[127] It is currently illegal and a punishable crime in India to determine or disclose the sex of a fetus using ultrasound equipment.[128]
Ultrasound is also a valuable tool inveterinary medicine, offering the same non-invasive imaging that helps in the diagnosis and monitoring of conditions in animals.
After theFrenchphysicistPierre Curie's discovery ofpiezoelectricity in 1880, ultrasonic waves could be deliberately generated for industry. In 1940, theAmerican acoustical physicistFloyd Firestone devised the first ultrasonic echo imaging device, the Supersonic Reflectoscope, to detect internal flaws in metal castings. In 1941,AustrianneurologistKarl Theo Dussik, in collaboration with his brother, Friedrich, a physicist, was likely the first person to image the human body ultrasonically, outlining the ventricles of a human brain.[129][130] Ultrasonic energy was first applied to the human body for medical purposes by Dr George Ludwig at the Naval Medical Research Institute,Bethesda, Maryland, in the late 1940s.[131][132]English-born physicistJohn Wild (1914–2009) first used ultrasound to assess the thickness ofbowel tissue as early as 1949; he has been described as the "father of medical ultrasound".[133] Subsequent advances took place concurrently in several countries but it was not until 1961 that David Robinson and George Kossoff's work at theAustralianDepartment of Health resulted in the first commercially practical water bath ultrasonic scanner.[134] In 1963 Meyerdirk & Wright launched production of the first commercial, hand-held, articulated arm, compound contact B-mode scanner, which made ultrasound generally available for medical use.
Léandre Pourcelot, a researcher and teacher at INSA (Institut National des Sciences Appliquées), Lyon, co-published a report in 1965 at the Académie des sciences, "Effet Doppler et mesure du débit sanguin" ("Doppler effect and measure of the blood flow"), the basis of his design of a Doppler flow meter in 1967.
Parallel developments inGlasgow, Scotland by ProfessorIan Donald and colleagues at theGlasgow Royal Maternity Hospital (GRMH) led to the first diagnostic applications of the technique.[135] Donald was anobstetrician with a self-confessed "childish interest in machines, electronic and otherwise", who, having treated the wife of one of the company's directors, was invited to visit the Research Department of boilermakersBabcock & Wilcox atRenfrew. He adapted their industrial ultrasound equipment to conduct experiments on various anatomical specimens and assess their ultrasonic characteristics. Together with the medical physicistTom Brown.[136] and fellow obstetricianJohn MacVicar, Donald refined the equipment to enable differentiation of pathology in live volunteer patients. These findings were reported inThe Lancet on 7 June 1958[137] as "Investigation of Abdominal Masses by Pulsed Ultrasound" – possibly one of the most important papers published in the field of diagnosticmedical imaging.
At GRMH, Professor Donald and James Willocks then refined their techniques to obstetric applications includingfetal head measurement to assess the size and growth of the fetus. With the opening of the new Queen Mother's Hospital inYorkhill in 1964, it became possible to improve these methods even further.Stuart Campbell's pioneering work on fetalcephalometry led to it acquiring long-term status as the definitive method of study of foetal growth. As the technical quality of the scans was further developed, it soon became possible to study pregnancy from start to finish and diagnose its many complications such as multiple pregnancy, fetal abnormality andplacenta praevia. Diagnostic ultrasound has since been imported into practically every other area of medicine.
Edler had asked Hertz if it was possible to useradar to look into the body, but Hertz said this wasimpossible. However, he said, it might be possible to use ultrasonography. Hertz was familiar with using ultrasonic reflectoscopes of the American acoustical physicistFloyd Firestone's invention fornondestructive materials testing, and together Edler and Hertz developed the idea of applying this methodology in medicine.
The first successful measurement of heart activity was made on October 29, 1953, using a device borrowed from the ship construction companyKockums inMalmö. On December 16 the same year, the method was applied to generate an echo-encephalogram (ultrasonic probe of thebrain). Edler and Hertz published their findings in 1954.[138]
In 1962, after about two years of work, Joseph Holmes, William Wright, and Ralph Meyerdirk developed the first compound contact B-mode scanner. Their work had been supported byU.S. Public Health Services and theUniversity of Colorado. Wright and Meyerdirk left the university to form Physionic Engineering Inc., which launched the first commercial hand-held articulated arm compound contact B-mode scanner in 1963. This was the start of the most popular design in the history of ultrasound scanners.[139]
In the late 1960sGene Strandness and the bio-engineering group at theUniversity of Washington conducted research on Doppler ultrasound as a diagnostic tool for vascular disease. Eventually, they developed technologies to use duplex imaging, or Doppler in conjunction with B-mode scanning, to view vascular structures in real time while also providing hemodynamic information.[140]
The first demonstration of color Doppler was by Geoff Stevenson, who was involved in the early developments and medical use of Doppler shifted ultrasonic energy.[141]
^It is for this reason that the person subjected to ultrasound of organs that can contain quantities of air or gas, such as the stomach, intestine and bladder, must follow a food preparation designed to reduce their quantity: specific diet and supplements for the intestine and intake of non-carbonated water to fill the bladder; sometimes, during the examination, it may be required to fill the stomach with non-carbonated water.
^Postema M, Kotopoulis S, Jenderka KV (2019). "Physical principles of medical ultrasound". In Dietrich CF (ed.).EFSUMB Course Book(PDF) (2nd ed.). London: EFSUMB. pp. 1–23.doi:10.37713/ECB01.S2CID216415694.
^Wachsberg RH (June 2007). "B-Flow Imaging of the Hepatic Vasculature: Correlation with Color Doppler Sonography".American Journal of Roentgenology.188 (6):W522–W533.doi:10.2214/AJR.06.1161.PMID17515342.
^Van Woudenberg N, Liao Z, Abdi AH, Girgis H, Luong C, Vaseli H, Behnami D, Zhang H, Gin K, Rohling R, Tsang T, Abolmaesumi P (2018). "Quantitative Echocardiography: Real-Time Quality Estimation and View Classification Implemented on a Mobile Android Device".Simulation, Image Processing, and Ultrasound Systems for Assisted Diagnosis and Navigation. Lecture Notes in Computer Science. Vol. 11042. pp. 74–81.doi:10.1007/978-3-030-01045-4_9.ISBN978-3-030-01044-7.
^Miles DA, Levi CS, Uhanova J, Cuvelier S, Hawkins K, Minuk GY (2020). "Pocket-Sized Versus Conventional Ultrasound for Detecting Fatty Infiltration of the Liver".Digestive Diseases and Sciences.65 (1):82–85.doi:10.1007/s10620-019-05752-x.PMID31376083.
^Costantino A, Piagnani A, Caccia R, Sorge A, Maggioni M, Perbellini R, Donato F, d'Ambrosio R, Sed NP, Valenti L, Prati D, Vecchi M, Lampertico P, Fraquelli M (2024). "Reproducibility and accuracy of a pocket-size ultrasound device in assessing liver steatosis".Digestive and Liver Disease.56 (6):1032–1038.doi:10.1016/j.dld.2023.11.014.PMID38016894.
^Hellman L, Duffus G, Donald I, Sundén B (May 1970). "Safety of Diagnostic Ultrasound in Obstetrics".The Lancet.295 (7657):1133–1135.doi:10.1016/s0140-6736(70)91212-2.PMID4192094.
^Brat R, Yousef N, Klifa R, Reynaud S, Shankar Aguilera S, De Luca D (August 2015). "Lung Ultrasonography Score to Evaluate Oxygenation and Surfactant Need in Neonates Treated With Continuous Positive Airway Pressure".JAMA Pediatrics.169 (8): e151797.doi:10.1001/jamapediatrics.2015.1797.ISSN2168-6211.PMID26237465.
^Kelner J, Moote D, Shah R, Anuar A, Golioto A (August 9, 2024). "Lung Ultrasound Score for Prediction of Surfactant Administration in Preterm Infants with Respiratory Failure".Journal of Perinatology.44 (9):1258–1263.doi:10.1038/s41372-024-02090-3.ISSN1476-5543.PMID39122885.
^Blanco PA, Cianciulli TF (2016). "Pulmonary Edema Assessed by Ultrasound: Impact in Cardiology and Intensive Care Practice".Echocardiography.33 (5):778–787.doi:10.1111/echo.13182.PMID26841270.S2CID37476194.
^Volpicelli G, Elbarbary M, Blaivas M, Lichtenstein DA, Mathis G, Kirkpatrick AW, Melniker L, Gargani L, Noble VE, Via G, Dean A, Tsung JW, Soldati G, Copetti R, Bouhemad B, Reissig A, Agricola E, Rouby JJ, Arbelot C, Liteplo A, Sargsyan A, Silva F, Hoppmann R, Breitkreutz R, Seibel A, Neri L, Storti E, Petrovic T (April 2012)."International evidence-based recommendations for point-of-care lung ultrasound".Intensive Care Medicine.38 (4):577–591.doi:10.1007/s00134-012-2513-4.PMID22392031.
^Arend CF. Ultrasound of the Shoulder. Porto Alegre: Master Medical Books; 2013. (Free access atShoulderUS.com)[page needed]
^Hübner U, Schlicht W, Outzen S, Barthel M, Halsband H (November 2000). "Ultrasound in the diagnosis of fractures in children".The Journal of Bone and Joint Surgery. British Volume.82-B (8):1170–1173.doi:10.1302/0301-620x.82b8.10087.PMID11132281.
^Abe T, Loene JP, Young KC, Thiebaud RS, Nahar VK, Hollaway KM, Stover CD, Ford MA, Bass MA (February 1, 2015). "Validity of ultrasound prediction equations for total and regional muscularity in middle-aged and older men and women".Ultrasound in Medicine & Biology.41 (2):557–564.doi:10.1016/j.ultrasmedbio.2014.09.007.PMID25444689.
^Ostensen H (2005).Basic physics of ultrasound imaging(PDF). Geneva: Diagnostic Imaging and Laboratory Technology - World Health Organisation. pp. 25–26. RetrievedOctober 2, 2021.
^[1] Abdul Latif Mohamed, Jun Yong, Jamil Masiyati, Lee Lim, Sze Chec Tee.The Prevalence Of Diastolic Dysfunction In Patients With Hypertension Referred For Echocardiographic Assessment of Left Ventricular Function. Malaysian Journal of Medical Sciences, Vol. 11, No. 1, January 2004, pp. 66–74
^Claudon M, Dietrich CF, Choi BI, Cosgrove DO, Kudo M, Nolsøe CP, Piscaglia F, Wilson SR, Barr RG, Chammas MC, Chaubal NG, Chen MH, Clevert DA, Correas JM, Ding H, Forsberg F, Fowlkes JB, Gibson RN, Goldberg BB, Lassau N, Leen EL, Mattrey RF, Moriyasu F, Solbiati L, Weskott HP, Xu HX, World Federation for Ultrasound in Medicine, European Federation of Societies for Ultrasound (2013). "Guidelines and Good Clinical Practice Recommendations for Contrast Enhanced Ultrasound (CEUS) in the Liver – Update 2012".Ultrasound in Medicine & Biology.39 (2):187–210.doi:10.1016/j.ultrasmedbio.2012.09.002.hdl:11585/144895.PMID23137926.S2CID2224370.
^Lassau N, Koscielny S, Chami L, Chebil M, Benatsou B, Roche A, Ducreux M, Malka D, Boige V (2010). "Advanced Hepatocellular Carcinoma: Early Evaluation of Response to Bevacizumab Therapy at Dynamic Contrast-enhanced US with Quantification—Preliminary Results".Radiology.258 (1):291–300.doi:10.1148/radiol.10091870.PMID20980447.
^Sugimoto K, Moriyasu F, Saito K, Rognin N, Kamiyama N, Furuichi Y, Imai Y (2013). "Hepatocellular carcinoma treated with sorafenib: Early detection of treatment response and major adverse events by contrast-enhanced US".Liver International.33 (4):605–15.doi:10.1111/liv.12098.PMID23305331.S2CID19338115.
^Rognin NG, Arditi M, Mercier L, Frinking PJ, Schneider M, Perrenoud G, Anaye A, Meuwly J, Tranquart F (2010). "Parametric imaging for characterizing focal liver lesions in contrast-enhanced ultrasound".IEEE Transactions on Ultrasonics, Ferroelectrics and Frequency Control.57 (11):2503–11.Bibcode:2010ITUFF..57.2503R.doi:10.1109/TUFFC.2010.1716.PMID21041137.S2CID19339331.
^Tranquart F, Mercier L, Frinking P, Gaud E, Arditi M (2012). "Perfusion Quantification in Contrast-Enhanced Ultrasound (CEUS) – Ready for Research Projects and Routine Clinical Use".Ultraschall in der Medizin.33: S31–8.doi:10.1055/s-0032-1312894.PMID22723027.S2CID8513304.
^Angelelli P, Nylund K, Gilja OH, Hauser H (2011). "Interactive visual analysis of contrast-enhanced ultrasound data based on small neighborhood statistics".Computers & Graphics.35 (2):218–226.doi:10.1016/j.cag.2010.12.005.
^Klibanov AL, Hughes MS, Marsh JN, Hall CS, Miller JG, Wilble JH, Brandenburger GH (1997). "Targeting of ultrasound contrast material. An in vitro feasibility study".Acta Radiologica Supplementum.412:113–120.PMID9240089.
^Klibanov A (1999). "Targeted delivery of gas-filled microspheres, contrast agents for ultrasound imaging".Advanced Drug Delivery Reviews.37 (1–3):139–157.doi:10.1016/S0169-409X(98)00104-5.PMID10837732.
^Pochon S, Tardy I, Bussat P, Bettinger T, Brochot J, Von Wronski M, Passantino L, Schneider M (2010). "BR55: A lipopeptide-based VEGFR2-targeted ultrasound contrast agent for molecular imaging of angiogenesis".Investigative Radiology.45 (2):89–95.doi:10.1097/RLI.0b013e3181c5927c.PMID20027118.S2CID24089981.
^Dayton P, Klibanov A, Brandenburger G,Ferrara, Kathy (1999). "Acoustic radiation force in vivo: A mechanism to assist targeting of microbubbles".Ultrasound in Medicine & Biology.25 (8):1195–1201.doi:10.1016/S0301-5629(99)00062-9.PMID10576262.
^Frinking PJ, Tardy I, Théraulaz M, Arditi M, Powers J, Pochon S, Tranquart F (2012). "Effects of Acoustic Radiation Force on the Binding Efficiency of BR55, a VEGFR2-Specific Ultrasound Contrast Agent".Ultrasound in Medicine & Biology.38 (8):1460–9.doi:10.1016/j.ultrasmedbio.2012.03.018.PMID22579540.
^Ophir, J., Céspides, I., Ponnekanti, H., Li, X. (1991). "Elastography: A quantitative method for imaging the elasticity of biological tissues".Ultrasonic Imaging.13 (2):111–34.doi:10.1016/0161-7346(91)90079-W.PMID1858217.
^Ozderya A, Aydin K, Gokkaya N, Temizkan S (June 2018). "Percutaneous Ethanol Injection for Benign Cystic and Mixed Thyroid Nodules".Endocrine Practice.24 (6):548–555.doi:10.4158/EP-2018-0013.PMID29624094.S2CID4665114.
^Jongbloets L, Koopman M, Büller H, Ten Cate J, Lensing A (1994). "Limitations of compression ultrasound for the detection of symptomless postoperative deep vein thrombosis".The Lancet.343 (8906):1142–4.doi:10.1016/S0140-6736(94)90240-2.PMID7910237.S2CID23576444.
^abKumar S."Panoramic Ultrasound".Conference: Proceedings of the Second National Conference on Signal & Image Processing, at S.M.K. Fomra Institute of Technology Chennai, India. April 2010
^Glozman T, Azhari H (March 2010). "A Method for Characterization of Tissue Elastic Properties Combining Ultrasonic Computed Tomography With Elastography".Journal of Ultrasound in Medicine.29 (3):387–398.doi:10.7863/jum.2010.29.3.387.PMID20194935.S2CID14869006.
^Goss SA, Johnston RL, Dunn F (July 1980). "Compilation of empirical ultrasonic properties of mammalian tissues. II".The Journal of the Acoustical Society of America.68 (1):93–108.Bibcode:1980ASAJ...68R..93G.doi:10.1121/1.384509.PMID11683186.
^Goss SA, Johnston RL, Dunn F (August 1978). "Comprehensive compilation of empirical ultrasonic properties of mammalian tissues".The Journal of the Acoustical Society of America.64 (2):423–457.Bibcode:1978ASAJ...64..423G.doi:10.1121/1.382016.PMID361793.
^Nightingale KR, Soo MS, Nightingale R, Trahey GE (2002). "Acoustic radiation force impulse imaging: in vivo demonstration of clinical feasibility".Ultrasound in Medicine & Biology.28 (2):227–235.doi:10.1016/s0301-5629(01)00499-9.PMID11937286.{{cite journal}}: CS1 maint: multiple names: authors list (link)
^Llamas-Álvarez AM, Tenza-Lozano EM, Latour-Pérez J (February 2017). "Accuracy of Lung Ultrasonography in the Diagnosis of Pneumonia in Adults".Chest.151 (2):374–382.doi:10.1016/j.chest.2016.10.039.PMID27818332.S2CID24399240.
^Training in diagnostic ultrasound: essentials, principles and standards: report of a WHO study group. World Health Organization. 1998. p. 2.hdl:10665/42093.ISBN978-92-4-120875-8.
^Levine, H., III (2010).Medical Imaging. Santa Barbara, California: ABC-CLIO, p. 62, describing earlier not completely successful attempt by the brothers to image a brain in 1937, which may be the same experiment.
^Edler I, Hertz CH (2004). "The Use of Ultrasonic Reflectoscope for the Continuous Recording of the Movements of Heart Walls".Clinical Physiology and Functional Imaging.24 (3):118–36.doi:10.1111/j.1475-097X.2004.00539.x.PMID15165281.S2CID46092067.