From Wikipedia, the free encyclopedia
Charged particle catcher
Faraday cup
Schematic diagram of a Faraday cup
UsesCharged particle detector
Related itemsElectron multiplier
Microchannel plate detector
Daly detector

AFaraday cup is ametal (conductive) cup designed to catchcharged particles. The resulting current can be measured and used to determine the number ofions orelectrons hitting the cup.[1] The Faraday cup was named afterMichael Faraday who first theorized ions around 1830.

Examples of devices which use Faraday cups includespace probes (Voyager 1, &2,Parker Solar Probe, etc.) andmass spectrometers. Faraday cups can also be used to measure chargedaerosol particles.

Principle of operation

[edit]
Faraday cup with an electron-suppressor plate in front

When a beam or packet ofions or electrons (e.g. from anelectron beam) hits the metallic body of the cup, the apparatus gains a small net charge. The cup can then be discharged to measure a small current proportional to the charge carried by the impinging ions or electrons. By measuring theelectric current (the number of electrons flowing through the circuit per second) in the cup, the number of charges can be determined. For a continuous beam of ions (assumed to be singly charged) or electrons, the total number N hitting the cup per unit time (in seconds) is

Nt=Ie{\displaystyle {\frac {N}{t}}={\frac {I}{e}}}

where I is the measured current (inamperes) and e is theelementary charge (1.60 × 10−19C). Thus, a measured current of one nanoamp (10−9 A) corresponds to about 6 billion singly charged particles striking the Faraday cup each second.

Faraday cups are not as sensitive aselectron multiplier detectors, but are highly regarded for accuracy because of the direct relation between the measured current and number of ions.

In plasma diagnostics

[edit]
This sectionmay be too technical for most readers to understand. Pleasehelp improve it tomake it understandable to non-experts, without removing the technical details.(September 2019) (Learn how and when to remove this message)

The Faraday cup uses a physical principle according to which the electrical charges delivered to the inner surface of a hollow conductor are redistributed around its outer surface due to mutual self-repelling of charges of the same sign – a phenomenon discovered byFaraday.[2]

Fig. 1. Faraday cup forplasma diagnostics 1 – cup-receiver, metal (stainless steel). 2 – electron-suppressor lid, metal (stainless steel). 3 – grounded shield, metal (stainless steel). 4 – insulator (teflon, ceramic).

The conventional Faraday cup is applied for measurements of ion (or electron) flows from plasma boundaries and comprises a metallic cylindrical receiver-cup – 1 (Fig. 1) closed with, and insulated from, a washer-type metallic electron-suppressor lid – 2 provided with the round axial through enter-hollow of an aperture with a surface areaSF=πDF2/4{\displaystyle S_{F}=\pi D_{F}^{2}/4}. Both the receiver cup and the electron-suppressor lid are enveloped in, and insulated from, a grounded cylindrical shield – 3 having an axial round hole coinciding with the hole in the electron-suppressor lid – 2. The electron-suppressor lid is connected by 50 Ω RF cable with the sourceBes{\displaystyle B_{es}} of variable DC voltageUes{\displaystyle U_{es}}. The receiver-cup is connected by 50 Ω RF cable through the load resistorRF{\displaystyle R_{F}} with a sweep generator producing saw-type pulsesUg(t){\displaystyle U_{g}(t)}. Electric capacityCF{\displaystyle C_{F}} is formed of the capacity of the receiver-cup – 1 to the grounded shield – 3 and the capacity of the RF cable. The signal fromRF{\displaystyle R_{F}} enables an observer to acquire anI-V characteristic of the Faraday cup by oscilloscope. Proper operating conditions:hDF{\displaystyle h\geq D_{F}} (due to possible potential sag) andhλi{\displaystyle h\ll \lambda _{i}}, whereλi{\displaystyle \lambda _{i}} is the ion free path. Signal fromRF{\displaystyle R_{F}} is the Faraday cupI-V characteristic which can be observed and memorized by oscilloscope

iΣ(Ug)=ii(Ug)CFdUgdt{\displaystyle i_{\Sigma }(U_{g})=i_{i}(U_{g})-C_{F}{\frac {dU_{g}}{dt}}}1

In Fig. 1: 1 – cup-receiver, metal (stainless steel). 2 – electron-suppressor lid, metal (stainless steel). 3 – grounded shield, metal (stainless steel). 4 – insulator (teflon, ceramic).CF{\displaystyle C_{F}} – capacity of Faraday cup.RF{\displaystyle R_{F}} – load resistor.

Thus we measure the sumiΣ{\displaystyle i_{\Sigma }} of the electric currents through the load resistorRF{\displaystyle R_{F}}:ii{\displaystyle i_{i}} (Faraday cup current) plus the currentic(Ug)=CF(dUg/dt){\displaystyle i_{c}(U_{g})=-C_{F}(dU_{g}/dt)} induced through the capacitorCF{\displaystyle C_{F}} by the saw-type voltageUg{\displaystyle U_{g}}of the sweep-generator: The current componentic(Ug){\displaystyle i_{c}(U_{g})} can be measured at the absence of the ion flow and can be subtracted further from the total currentiΣ(Ug){\displaystyle i_{\Sigma }(U_{g})} measured with plasma to obtain the actual Faraday cupI-V characteristicii(Ug){\displaystyle i_{i}(U_{g})} for processing. All of the Faraday cup elements and their assembly that interact with plasma are fabricated usually of temperature-resistant materials (often these are stainless steel and teflon or ceramic for insulators). For processing of the Faraday cupI-V characteristic, we are going to assume that the Faraday cup is installed far enough away from an investigated plasma source where the flow of ions could be considered as the flow of particles with parallel velocities directed exactly along the Faraday cup axis. In this case, the elementary particle currentdii{\displaystyle di_{i}} corresponding to the ion density differentialdn(v){\displaystyle dn(v)} in the range of velocities betweenv{\displaystyle v} andv+dv{\displaystyle v+dv} of ions flowing in through operating apertureSF{\displaystyle S_{F}} of the electron-suppressor can be written in the form

dii=eZiSFvdn(v){\displaystyle di_{i}=eZ_{i}S_{F}vdn(v)}2

where

dn(v)=nf(v)dv{\displaystyle dn(v)=nf(v)dv}3

e{\displaystyle e} is elementary charge,Zi{\displaystyle Z_{i}} is the ion charge state, andf(v){\displaystyle f(v)} is the one-dimensional ion velocity distribution function. Therefore, the ion current at the ion-decelerating voltageUg{\displaystyle U_{g}} of the Faraday cup can be calculated by integrating Eq. (2) after substituting Eq. (3),

ii(Ug)=eZiniSF2eZiUg/Mif(v)vdv{\displaystyle i_{i}(U_{g})=eZ_{i}n_{i}S_{F}\int \limits _{\sqrt {2eZ_{i}U_{g}/M_{i}}}^{\infty }f(v)vdv}4

where the lower integration limit is defined from the equationMivi,s2/2=eZiUg{\displaystyle M_{i}v_{i,s}^{2}/2=eZ_{i}U_{g}} wherevi,s{\displaystyle v_{i,s}} is the velocity of the ion stopped by the decelerating potentialUg{\displaystyle U_{g}}, andMi{\displaystyle M_{i}} is the ion mass. Thus Eq. (4) represents theI-V characteristic of the Faraday cup. Differentiating Eq. (4) with respect toUg{\displaystyle U_{g}}, one can obtain the relation

dii(Ug)dUg=eniSFeZiMif(2eZiUgMi){\displaystyle {\frac {di_{i}(U_{g})}{dU_{g}}}=-en_{i}S_{F}{\frac {eZ_{i}}{M_{i}}}f\left({\sqrt {\frac {2eZ_{i}U_{g}}{M_{i}}}}\right)}5

where the valueniSF(eZi/Mi)=Ci{\displaystyle -n_{i}S_{F}(eZ_{i}/M_{i})=C_{i}} is an invariable constant for each measurement. Therefore, the average velocityvi{\displaystyle \langle v_{i}\rangle } of ions arriving into the Faraday cup and their average energyEi{\displaystyle \langle {\mathcal {E}}_{i}\rangle } can be calculated (under the assumption that we operate with a single type of ion) by the expressions

vi=1.389×106ZiMA0ii(Ug)dUg(0iiUgdUg)1{\displaystyle \langle v_{i}\rangle =1.389\times 10^{6}{\sqrt {\frac {Z_{i}}{M_{A}}}}\int \limits _{0}^{\infty }i_{i}^{\prime }(U_{g})dU_{g}\left(\int \limits _{0}^{\infty }{\frac {i_{i}^{\prime }}{\sqrt {U_{g}}}}dU_{g}\right)^{-1}} [cm/s]6
Ei=0ii(Ug)UgdUg(0iiUgdUg)1{\displaystyle \langle {\mathcal {E}}_{i}\rangle =\int \limits _{0}^{\infty }i_{i}^{\prime }(U_{g}){\sqrt {U_{g}}}dU_{g}\left(\int \limits _{0}^{\infty }{\frac {i_{i}^{\prime }}{\sqrt {U_{g}}}}dU_{g}\right)^{-1}} [eV]7

whereMA{\displaystyle M_{A}} is the ion mass in atomic units. The ion concentrationni{\displaystyle n_{i}} in the ion flow at the Faraday cup vicinity can be calculated by the formula

ni=ii(0)eZiviSF{\displaystyle n_{i}={\frac {i_{i}(0)}{eZ_{i}\langle v_{i}\rangle S_{F}}}}8

which follows from Eq. (4) atUg=0{\displaystyle U_{g}=0},

0f(v)vdv=v{\displaystyle \int \limits _{0}^{\infty }f(v)vdv=\langle v\rangle }9
Fig, 2. Faraday cup I-V characteristic

and from the conventional condition for distribution function normalizing

0f(v)dv=1{\displaystyle \int \limits _{0}^{\infty }f(v)dv=1}10

Fig. 2 illustrates theI-V characteristicii(V){\displaystyle i_{i}(V)} and its first derivativeii(V){\displaystyle i_{i}^{\prime }(V)} of the Faraday cup withSF=0.5cm2{\displaystyle S_{F}=0.5cm^{2}} installed at output of theInductively coupled plasma source powered with RF13.56 MHz and operating at 6 mTorr of H2. The value of the electron-suppressor voltage (accelerating the ions) was set experimentally atUes=170V{\displaystyle U_{es}=-170V}, near the point of suppression of thesecondary electron emission from the inner surface of the Faraday cup.[3]

Error sources

[edit]

The counting of charges collected per unit time is impacted by two error sources: 1) the emission of low-energysecondary electrons from the surface struck by the incident charge and 2)backscattering (~180 degree scattering) of the incident particle, which causes it to leave the collecting surface, at least temporarily. Especially with electrons, it is fundamentally impossible to distinguish between a fresh new incident electron and one that has been backscattered or even a fast secondary electron.

See also

[edit]

References

[edit]
  1. ^Brown, K. L.; G. W. Tautfest (September 1956)."Faraday-Cup Monitors for High-Energy Electron Beams"(PDF).Review of Scientific Instruments.27 (9):696–702.Bibcode:1956RScI...27..696B.doi:10.1063/1.1715674. Retrieved2007-09-13.
  2. ^Frank A. J. L. James (2004). "Faraday, Michael (1791–1867)".Oxford Dictionary of National Biography. Vol. 1 (online ed.). Oxford University Press.doi:10.1093/ref:odnb/9153. (Subscription orUK public library membership required.)
  3. ^E. V. Shun'ko. (2009).Langmuir Probe in Theory and Practice. Universal Publishers, Boca Raton, Fl. 2008. p. 249.ISBN 978-1-59942-935-9.

External links

[edit]
Ion source
Mass analyzer
Detector
MS combination
Fragmentation
Physics
Lectures
Related
Retrieved from "https://en.wikipedia.org/w/index.php?title=Faraday_cup&oldid=1242714529"