Enzyme assays arelaboratory methods for measuringenzymatic activity. They are vital for the study ofenzyme kinetics andenzyme inhibition.
The quantity or concentration of anenzyme can be expressed inmolar amounts, as with any other chemical, or in terms of activity inenzyme units.
Enzyme activity is a measure of the quantity of active enzyme present and is thus dependent on various physical conditions,which should be specified.
It is calculated using the following formula:
where
The SI unit is thekatal, 1 katal = 1 mol s−1 (mole per second), but this is an excessively large unit. A more practical and commonly used value isenzyme unit (U) = 1μmol min−1 (micromole per minute). 1 U corresponds to 16.67nanokatals.[1]
Enzyme activity as given in katal generally refers to that of the assumed natural target substrate of the enzyme. Enzyme activity can also be given as that of certain standardized substrates, such asgelatin, then measured ingelatin digesting units (GDU), or milk proteins, then measured inmilk clotting units (MCU). The units GDU and MCU are based on how fast one gram of the enzyme will digest gelatin or milk proteins, respectively. 1 GDU approximately equals 1.5 MCU.[2]
An increased amount of substrate will increase the rate of reaction with enzymes, however once past a certain point, the rate of reaction will level out because the amount of active sites available has stayed constant.
The specific activity of an enzyme is another common unit. This is the activity of an enzyme per milligram of total protein (expressed in μmol min−1 mg−1). Specific activity gives a measurement of enzyme purity in the mixture. It is the micro moles of product formed by anenzyme in a given amount of time (minutes) under given conditions per milligram of total proteins. Specific activity is equal to the rate of reaction multiplied by the volume of reaction divided by the mass of total protein. The SI unit is katal/kg, but a more practical unit is μmol/(mg*min).
Specific activity is a measure ofenzyme processivity (the capability of enzyme to be processed), at a specific (usually saturating)substrate concentration, and is usually constant for a pure enzyme.
An active site titration process can be done for the elimination of errors arising from differences in cultivation batches and/or misfolded enzyme and similar issues. This is a measure of the amount of active enzyme, calculated by e.g. titrating the amount of active sites present by employing an irreversible inhibitor. The specific activity should then be expressed as μmol min−1 mg−1 active enzyme. If the molecular weight of the enzyme is known, theturnover number, or μmol product per second per μmol of active enzyme, can be calculated from the specific activity. The turnover number can be visualized as the number of times each enzyme molecule carries out its catalytic cycle per second.
Therate of a reaction is the concentration of substrate disappearing (or product produced) per unit time (mol L−1 s−1).
The% purity is 100% × (specific activity of enzyme sample / specific activity of pure enzyme). The impure sample has lower specific activity because some of the mass is not actually enzyme. If the specific activity of 100% pure enzyme is known, then an impure sample will have a lower specific activity, allowing purity to be calculated and then getting a clear result.
All enzymeassays measure either the consumption of substrate or production of product over time. A large number of different methods of measuring the concentrations of substrates and products exist and many enzymes can be assayed in several different ways. Biochemists usually study enzyme-catalysed reactions using four types of experiments:[3]
Enzyme assays can be split into two groups according to their sampling method:continuous assays, where the assay gives a continuous reading of activity, anddiscontinuous assays, where samples are taken, the reaction stopped and then the concentration of substrates/products determined.

Continuous assays are most convenient, with one assay giving the rate of reaction with no further work necessary. There are many different types of continuous assays.
Inspectrophotometric assays, you follow the course of the reaction by measuring a change in how much light the assay solution absorbs. If this light is in the visible region you can actually see a change in the color of the assay, and these are calledcolorimetric assays. TheMTT assay, a redox assay using a tetrazolium dye as substrate is an example of a colorimetric assay.
UV light is often used, since the common coenzymesNADH andNADPH absorb UV light in theirreduced forms, but do not in theiroxidized forms. Anoxidoreductase using NADH as a substrate could therefore be assayed by following the decrease in UV absorbance at a wavelength of 340 nm as it consumes the coenzyme.[4]
Direct versus coupled assays

Even when the enzyme reaction does not result in a change in the absorbance of light, it can still be possible to use a spectrophotometric assay for the enzyme by using acoupled assay. Here, the product of one reaction is used as the substrate of another, easily detectable reaction. For example, figure 1 shows the coupled assay for the enzymehexokinase, which can be assayed by coupling its production of glucose-6-phosphate to NADPH production, usingglucose-6-phosphate dehydrogenase.
Fluorescence is when a molecule emits light of onewavelength after absorbing light of a different wavelength. Fluorometric assays use a difference in the fluorescence of substrate from product to measure the enzyme reaction. These assays are in general much more sensitive than spectrophotometric assays, but can suffer from interference caused by impurities and the instability of many fluorescent compounds when exposed to light.
An example of these assays is again the use of the nucleotide coenzymes NADH and NADPH. Here, the reduced forms are fluorescent and the oxidised forms non-fluorescent. Oxidation reactions can therefore be followed by a decrease in fluorescence and reduction reactions by an increase.[5] Synthetic substrates that release a fluorescent dye in an enzyme-catalyzed reaction are also available, such as 4-methylumbelliferyl-β-D-galactoside for assayingβ-galactosidase or 4-methylumbelliferyl-butyrate for assayingCandida rugosalipase.[6]

Calorimetry is the measurement of the heat released or absorbed by chemical reactions. These assays are very general, since many reactions involve some change in heat and with use of a microcalorimeter, not much enzyme or substrate is required. These assays can be used to measure reactions that are impossible to assay in any other way.[7]
Chemiluminescence is the emission of light by a chemical reaction. Some enzyme reactions produce light and this can be measured to detect product formation. These types of assay can be extremely sensitive, since the light produced can be captured by photographic film over days or weeks, but can be hard to quantify, because not all the light released by a reaction will be detected.
The detection ofhorseradish peroxidase by enzymatic chemiluminescence (ECL) is a common method of detecting antibodies inwestern blotting. Another example is the enzymeluciferase, this is found in fireflies and naturally produces light from its substrate luciferin.
Static light scattering measures the product of weight-averaged molar mass and concentration of macromolecules in solution. Given a fixed total concentration of one or more species over the measurement time, the scattering signal is a direct measure of the weight-averaged molar mass of the solution, which will vary as complexes form or dissociate.Hence the measurement quantifies the stoichiometry of the complexes as well as kinetics. Light scattering assays of protein kinetics is a very general technique that does not require an enzyme.
Microscale thermophoresis (MST)[8] measures the size, charge and hydration entropy of molecules/substrates at equilibrium.[9] The thermophoretic movement of a fluorescently labeled substrate changes significantly as it is modified by an enzyme. This enzymatic activity can be measured with high time resolution in real time.[10] The material consumption of the all optical MST method is very low, only 5 μl sample volume and 10nM enzyme concentration are needed to measure the enzymatic rate constants for activity and inhibition. MST allows analysts to measure the modification of two different substrates at once (multiplexing) if both substrates are labeled with different fluorophores. Thus substrate competition experiments can be performed.
Discontinuous assays are when samples are taken from an enzyme reaction at intervals and the amount of product production or substrate consumption is measured in these samples.
Radiometric assays measure the incorporation ofradioactivity into substrates or its release from substrates. Theradioactive isotopes most frequently used in these assays are14C,32P,35S and125I. Since radioactive isotopes can allow the specific labelling of a single atom of a substrate, these assays are both extremely sensitive and specific. They are frequently used in biochemistry and are often the only way of measuring a specific reaction in crude extracts (the complex mixtures of enzymes produced when you lyse cells).
Radioactivity is usually measured in these procedures using ascintillation counter.
Chromatographic assays measure product formation by separating the reaction mixture into its components bychromatography. This is usually done byhigh-performance liquid chromatography (HPLC), but can also use the simpler technique ofthin layer chromatography. Although this approach can need a lot of material, its sensitivity can be increased by labelling the substrates/products with a radioactive or fluorescent tag. Assay sensitivity has also been increased by switching protocols to improved chromatographic instruments (e.g. ultra-high pressure liquid chromatography) that operate at pump pressure a few-fold higher than HPLC instruments (seeHigh-performance liquid chromatography#Pump pressure).[11]

Several factors effect the assay outcome and a recent review summarizes the various parameters that needs to be monitored to keep an assay up and running.
Most enzymes cannot tolerate extremely high salt concentrations. The ions interfere with the weakionic bonds ofproteins. Typical enzymes are active in salt concentrations of 1-500 mM. As usual there are exceptions such as thehalophilicalgae and halophilicbacteria.
All enzymes work within a range of temperature specific to the organism. Increases in temperature generally lead to increases in reaction rates. There is a limit to the increase because higher temperatures lead to a sharp decrease in reaction rates. This is due to the denaturating (alteration) ofprotein structure resulting from the breakdown of the weakionic andhydrogen bonding that stabilize the three-dimensional structure of the enzyme active site.[12] The "optimum" temperature for human enzymes is usually between 35 and 40 °C. The average temperature for humans is 37 °C. Human enzymes start to denature quickly at temperatures above 40 °C. Enzymes fromthermophilicarchaea found in the hot springs are stable up to 100 °C.[13] However, the idea of an "optimum" rate of an enzyme reaction is misleading, as the rate observed at any temperature is the product of two rates, the reaction rate and the denaturation rate. If you were to use an assay measuring activity for one second, it would give high activity at high temperatures, however if you were to use an assay measuring product formation over an hour, it would give you low activity at these temperatures.
Most enzymes are sensitive topH and have specific ranges of activity. All have an optimum pH. The pH can stop enzyme activity by denaturating (altering) the three-dimensional shape of the enzyme by breakingionic, andhydrogen bonds. Most enzymes function between a pH of 6 and 8; however pepsin in the stomach works best at a pH of 2 and trypsin at a pH of 8.
Increasing thesubstrate concentration increases the rate of reaction (enzyme activity). However, enzyme saturation limits reaction rates. An enzyme is saturated when the active sites of all the molecules are occupied most of the time. At the saturation point, the reaction will not speed up, no matter how much additional substrate is added. The graph of the reaction rate will plateau.
Large amounts ofmacromolecules in a solution will alter therates andequilibrium constants of enzyme reactions, through an effect calledmacromolecular crowding.[14]