Effective dose is adose quantity in theInternational Commission on Radiological Protection (ICRP) system ofradiological protection.[1]It is the tissue-weighted sum of theequivalent doses in all specified tissues and organs of the human body. It represents thestochastic health risk to the whole body, which is theprobability ofcancer induction and genetic effects, of low levels ofionizing radiation.[2][3] It takes into account the type of radiation and the nature of each organ or tissue being irradiated, and enables summation of organ doses due to varying levels and types of radiation, both internal and external, to produce an overall calculated effective dose.
The SI unit for effective dose is thesievert (Sv) which corresponds to a 5.5% chance of developing cancer.[4] The effective dose is not intended as a measure ofdeterministic health effects, which is theseverity of acute tissue damage that is certain to happen, that is measured by the quantityabsorbed dose.[5]
The concept of effective dose was developed by Wolfgang Jacobi and published in 1975, and was so convincing that the ICRP incorporated it into their 1977 general recommendations (publication 26) as "effective dose equivalent".[6] The name "effective dose" replaced the name "effective dose equivalent" in 1991.[7] Since 1977 it has been the central quantity for dose limitation in the ICRP international system ofradiological protection.[1]

According to the ICRP, the main uses of effective dose are the prospective dose assessment for planning and optimisation in radiological protection, and demonstration of compliance with dose limits for regulatory purposes. The effective dose is thus a central dose quantity for regulatory purposes.[8]
The ICRP also says that effective dose has made a significant contribution to radiological protection as it has enabled doses to be summed from whole and partial body exposure from external radiation of various types and from intakes of radionuclides.[9]
The calculation of effective dose is required for partial or non-uniform irradiation of the human body becauseequivalent dose does not consider the tissue irradiated, but only the radiation type. Various body tissues react to ionising radiation in different ways, so the ICRP has assigned sensitivity factors to specified tissues and organs so that the effect of partial irradiation can be calculated if the irradiated regions are known.[10] A radiation field irradiating only a portion of the body will carry lower risk than if the same field irradiated the whole body. To take this into account, the effective doses to the component parts of the body which have been irradiated are calculated and summed. This becomes the effective dose for the whole body, dose quantityE. It is a "protection" dose quantity which can be calculated, but cannot be measured in practice.
An effective dose will carry the same effective risk to the whole body regardless of where it was applied, and it will carry the same effective risk as the same amount of equivalent dose applied uniformly to the whole body.
Effective dose can be calculated forcommitted dose which is the internal dose resulting from inhaling, ingesting, or injecting radioactive materials.
The dose quantity used is:
Committed effective dose,E(t) is the sum of the products of the committed organ or tissue equivalent doses and the appropriate tissue weighting factorsWT, wheret is the integration time in years following the intake. The commitment period is taken to be 50 years for adults, and to age 70 years for children.[11]


Ionizing radiation deposits energy in the matter being irradiated. The quantity used to express this is theabsorbed dose, a physical dose quantity that is dependent on the level of incident radiation and the absorption properties of the irradiated object. Absorbed dose is a physical quantity, and is not a satisfactory indicator of biological effect, so to allow consideration of the stochastic radiological risk, the dose quantities equivalent dose and effective dose were devised by theInternational Commission on Radiation Units and Measurements (ICRU) and the ICRP to calculate the biological effect of an absorbed dose.
To obtain an effective dose, the calculated absorbed organ doseDT is first corrected for the radiation type using factorWR to give a weighted average of the equivalent dose quantityHT received in irradiated body tissues, and the result is further corrected for the tissues or organs being irradiated using factorWT, to produce the effective dose quantityE.
The sum of effective doses to all organs and tissues of the body represents the effective dose for the whole body. If only part of the body is irradiated, then only those regions are used to calculate the effective dose. The tissue weighting factors summate to 1.0, so that if an entire body is radiated with uniformly penetrating external radiation, the effective dose for the entire body is equal to the equivalent dose for the entire body.
The ICRP tissue weighting factors are given in the accompanying table, and the equations used to calculate from either absorbed dose or equivalent dose are also given.
Some tissues like bone marrow are particularly sensitive to radiation, so they are given a weighting factor that is disproportionately large relative to the fraction of body mass they represent. Other tissues like the hard bone surface are particularly insensitive to radiation and are assigned a disproportionally low weighting factor.
| Organs | Tissue weighting factors | ||
|---|---|---|---|
| ICRP26 1977 | ICRP60 1990[14] | ICRP103 2007[15] | |
| Gonads | 0.25 | 0.20 | 0.08 |
| RedBone Marrow | 0.12 | 0.12 | 0.12 |
| Colon | – | 0.12 | 0.12 |
| Lung | 0.12 | 0.12 | 0.12 |
| Stomach | – | 0.12 | 0.12 |
| Breasts | 0.15 | 0.05 | 0.12 |
| Bladder | – | 0.05 | 0.04 |
| Liver | – | 0.05 | 0.04 |
| Oesophagus | – | 0.05 | 0.04 |
| Thyroid | 0.03 | 0.05 | 0.04 |
| Skin | – | 0.01 | 0.01 |
| Bone surface | 0.03 | 0.01 | 0.01 |
| Salivary glands | – | – | 0.01 |
| Brain | – | – | 0.01 |
| Remainder of body | 0.30 | 0.05 | 0.12 |
| Total | 1.00 | 1.00 | 1.00 |
Calculating from the equivalent dose:
Calculating from the absorbed dose:
Where
The ICRP tissue weighting factors are chosen to represent the fraction of health risk, or biological effect, which is attributable to the specific tissue named. These weighting factors have been revised twice, as shown in the chart above.
The United StatesNuclear Regulatory Commission still uses the ICRP's 1977 tissue weighting factors in their regulations, despite the ICRP's later revised recommendations.[16]
| Target organs | Exam type | Effective dose in adults[17] | Equivalent time ofbackground radiation[17] |
|---|---|---|---|
| CT of the head | Single series | 2 mSv | 8 months |
| With + withoutradiocontrast | 4 mSv | 16 months | |
| Chest | CT of the chest | 7 mSv | 2 years |
| CT of the chest, lung cancer screening protocol | 1.5 mSv | 6 months | |
| Chest X-ray | 0.1 mSv | 10 days | |
| Heart | Coronary CT angiography | 12 mSv | 4 years |
| Coronary CT calcium scan | 3 mSv | 1 year | |
| Abdominal | CT of abdomen and pelvis | 10 mSv | 3 years |
| CT of abdomen and pelvis, low dose protocol | 3 mSv[18] | 1 year | |
| CT of abdomen and pelvis, with + withoutradiocontrast | 20 mSv | 7 years | |
| CT Colonography | 6 mSv | 2 years | |
| Intravenous pyelogram | 3 mSv | 1 year | |
| Upper gastrointestinal series | 6 mSv | 2 years | |
| Lower gastrointestinal series | 8 mSv | 3 years | |
| Spine | Spine X-ray | 1.5 mSv | 6 months |
| CT of the spine | 6 mSv | 2 years | |
| Extremities | X-ray of extremity | 0.001 mSv | 3 hours |
| Lower extremityCT angiography | 0.3 - 1.6 mSv[19] | 5 weeks - 6 months | |
| Dental X-ray | 0.005 mSv | 1 day | |
| DEXA (bone density) | 0.001 mSv | 3 hours | |
| PET-CT combination | 25 mSv | 8 years | |
| Mammography | 0.4 mSv | 7 weeks | |
Ionizing radiation is generally harmful and potentially lethal to living things but can have health benefits inradiation therapy for the treatment of cancer andthyrotoxicosis. Its most common impact is theinduction of cancer with alatent period of years or decades after exposure. High doses can cause visually dramaticradiation burns, and/or rapid fatality throughacute radiation syndrome. Controlled doses are used formedical imaging andradiotherapy.
The UKIonising Radiations Regulations 1999 defines its usage of the term effective dose; "Any reference to an effective dose means the sum of the effective dose to the whole body from external radiation and the committed effective dose from internal radiation."[20]
The USNuclear Regulatory Commission has retained in the US regulation system the older termeffective dose equivalent to refer to a similar quantity to the ICRP effective dose. The NRC'stotal effective dose equivalent (TEDE) is the sum of external effective dose with internal committed dose; in other words all sources of dose.
In the US, cumulative equivalent dose due to external whole-body exposure is normally reported to nuclear energy workers in regular dosimetry reports.
The concept of effective dose was introduced in 1975 by Wolfgang Jacobi (1928–2015) in his publication "The concept of an effective dose: a proposal for the combination of organ doses".[6][21] It was quickly included in 1977 as “effective dose equivalent” into Publication 26 by the ICRP. In 1991, ICRP publication 60 shortened the name to "effective dose."[22] This quantity is sometimes incorrectly referred to as the "dose equivalent" because of the earlier name, and that misnomer in turn causes confusion withequivalent dose. The tissue weighting factors were revised in 1990 and 2007 due to new data.
At the ICRP 3rd International Symposium on the System of Radiological Protection in October 2015, ICRP Task Group 79 reported on the "Use of Effective Dose as a Risk-related Radiological Protection Quantity".
This included a proposal to discontinue use of equivalent dose as a separate protection quantity. This would avoid confusion between equivalent dose, effective dose and dose equivalent, and to use absorbed dose in Gy as a more appropriate quantity for limiting deterministic effects to the eye lens, skin, hands & feet.[23]
It was also proposed that effective dose could be used as a rough indicator of possible risk from medical examinations. These proposals will need to go through the following stages:
M.A. Boyd."The Confusing World of Radiation Dosimetry - 9444"(PDF).US Environmental Protection Agency. Archived fromthe original(PDF) on 2016-12-21. Retrieved2014-05-26. – an account of chronological differences between USA and ICRP dosimetry systems